Pushing the limits: latest developments in angle metrology for the inspection of ultra-precise synchrotron optics

The requirements on the quality of ultra-precise X-ray optical components for application in the Synchrotron Radiation (SR) community are increasing continually and strongly depend on the quality of the metrology devices available to measure such optics. To meet the upcoming accuracy goal of 50 nrad rms for slope measuring profilers, a dedicated project, SIB58 Angles, consisting of 16 worldwide partners and supported by the European Metrology Research Programme (EMRP) was started in Sep 2013. The project covers investigations on autocollimators under extremely challenging measuring conditions, ray-tracing models, 2D autocollimator calibration (for the first time worldwide), determination of error sources in angle encoders providing traceability by ‘sub-division of 2π rad’ with nrad uncertainty, angle generation by 'ratio of two lengths' in nrad level, and on the development of portable precise Small Angle Generators (SAGs) for regular in-situ checks of autocollimators’ performance. Highlights from the project will be reported in the paper and the community of metrology for X-Ray and EUV Optics will be informed about its progress and the latest work in angle metrology.

[1]  S Asli Akgoz,et al.  Calibration of high-resolution electronic autocollimators with demanded low uncertainties using single reading head angle encoders , 2013 .

[2]  Satoshi Kiyono,et al.  Self-calibration of precision angle sensor and polygon mirror , 1997 .

[3]  Gerald Futterer Simulation of the detectors response of an autocollimator , 2007 .

[4]  Clemens Elster,et al.  A New Flatness Reference Measurement System Based on Deflectometry and Difference Deflectometry , 2009 .

[5]  Th. Tschentscher,et al.  Technical Report: The European X-ray Free-Electron Laser Facility: A New Infrastructure for Research Using Ultrashort, Coherent X-ray Pulses of Extreme Brightness , 2006 .

[6]  Frank Siewert,et al.  The Diamond-NOM: A non-contact profiler capable of characterizing optical figure error with sub-nanometre repeatability , 2010 .

[7]  P Cloetens,et al.  Comparative study of multilayers used in monochromators for synchrotron-based coherent hard X-ray imaging. , 2010, Journal of synchrotron radiation.

[8]  Jana Buchheim,et al.  Characterization and calibration of 2nd generation slope measuring profiler , 2010 .

[9]  Zoran Ninkov,et al.  Influence of nonuniform charge-coupled device pixel response on aperture photometry , 2001 .

[10]  François Polack,et al.  Latest metrology results with the SOLEIL synchrotron LTP , 2005, SPIE Optics + Photonics.

[11]  Michael Schulz,et al.  Optical measurement of absolute flatness with the deflectometric measurement systems at PTB , 2013 .

[12]  Gregory Y. Morrison,et al.  Performance of the upgraded LTP-II at the ALS Optical Metrology Laboratory , 2008, Optical Engineering + Applications.

[13]  Gerald Futterer Enhancement of high-resolution electronic autocollimators by application of phase grating technology , 2005, SPIE Optical Metrology.

[14]  Ralf D. Geckeler,et al.  Use and treaceable calibration of autocollimators for ultra-precise measurement of slope and topography , 2001, Lasers in Metrology and Art Conservation.

[15]  Peter Z. Takacs,et al.  Calibration of a vertical-scan long trace profiler at MSFC , 2001, Optics + Photonics.

[16]  Zoran Ninkov,et al.  Subpixel sensitivity maps for a back-illuminated charge-coupled device and the effects of nonuniform response on measurement accuracy , 2002 .

[17]  Frank Siewert,et al.  Fully automated, fixed exit, in vacuum double-multilayer monochromator for synchrotron-based hard X-ray micro-imaging applications , 2010 .

[18]  Gianluca Geloni,et al.  The European X-ray Free-Electron Laser , 2015 .

[19]  K. Kaznatcheev,et al.  Current status of the NSLS-II optical metrology laboratory , 2013 .

[20]  Makoto Kajitani,et al.  Automatic high-precision calibration system for angle encoder (II) , 2003, SPIE Optics + Photonics.

[21]  Valeriy V. Yashchuk Positioning errors of pencil-beam interferometers for long trace profilers , 2006, SPIE Optics + Photonics.

[22]  Ralf D. Geckeler,et al.  Optimized use and calibration of autocollimators in deflectometry , 2007, SPIE Optical Engineering + Applications.

[23]  Ralf D. Geckeler Error minimization in high-accuracy scanning deflectometry , 2006, SPIE Optics + Photonics.

[24]  Makoto Kajitani,et al.  High Accuracy Calibration System for Angular Encoders , 1993, J. Robotics Mechatronics.

[25]  Frank Siewert,et al.  Upgrade of long trace profiler for characterization of high-precision X-ray mirrors at SPring-8 , 2010 .

[26]  Garth J. Williams,et al.  Ultra-precise characterization of LCLS hard X-ray focusing mirrors by high resolution slope measuring deflectometry. , 2012, Optics express.

[27]  T. Ishikawa,et al.  Breaking the 10 nm barrier in hard-X-ray focusing , 2010 .

[28]  T. Masuda,et al.  An automatic calibration system for angular encoders , 1989 .

[29]  P. Křen,et al.  Optical mirror referenced capacitive flatness measurement and straightness evaluation of translation stages , 2014 .

[30]  Andreas Just,et al.  Calibration of high-resolution electronic autocollimators against an angle comparator , 2003 .

[31]  T. Zeschke,et al.  The Nanometer Optical Component Measuring Machine: a new Sub-nm Topography Measuring Device for X-ray Optics at BESSY , 2004 .

[32]  Clemens Elster,et al.  Alignment methods for ultraprecise deflectometric flatness metrology , 2011, Optical Metrology.

[33]  Clemens Elster,et al.  Deflectometric systems for absolute flatness measurements at PTB , 2012 .

[34]  Shinan Qian,et al.  The penta‐prism LTP: A long‐trace‐profiler with stationary optical head and moving penta prism (abstract)a) , 1995 .

[35]  Ralf D. Geckeler,et al.  A novel approach for extending autocollimator calibration from plane to spatial angles , 2012 .

[36]  Ralf D. Geckeler,et al.  Capabilities and limitations of the self-calibration of angle encoders , 2014 .

[37]  Hidekazu Mimura,et al.  Requirements on hard x-ray grazing incidence optics for European XFEL: analysis and simulation of wavefront transformations , 2009, Optics + Optoelectronics.

[38]  Petr Křen A simple interferometric method for determining the flatness of large optical flats with 1 nm repeatability , 2008 .

[39]  Frank Siewert Slope Error and Surface Roughness , 2008 .

[40]  Peter Z. Takacs,et al.  Design Of A Long Trace Surface Profiler , 1987, Photonics West - Lasers and Applications in Science and Engineering.

[41]  Makoto Kajitani,et al.  Automatic high-precision calibration system for angle encoder , 2001, Lasers in Metrology and Art Conservation.

[42]  H Han Haitjema,et al.  A novel technique for calibration of polygon angles with non-integer subdivision of indexing table , 2002 .

[43]  S. C. Irick,et al.  Determining surface profile from sequential interference patterns from a long trace profiler , 1991 .

[44]  Ralf D. Geckeler,et al.  Distance-dependent influences on angle metrology with autocollimators in deflectometry , 2008, Optical Engineering + Applications.

[45]  T. Yandayana,et al.  A novel technique for calibration of polygon angles with non-integer subdivision of indexing table , 2002 .

[46]  Valeriy V. Yashchuk,et al.  Proposal for a universal test mirror for characterization of slope measuring instruments , 2007, SPIE Optical Engineering + Applications.

[47]  A. Vivo,et al.  Optical metrology facility at the ESRF , 1995 .

[48]  Clemens Elster,et al.  Concept, design and capability analysis of the new Deflectometric Flatness Reference at PTB , 2010 .

[49]  Gregory Y. Morrison,et al.  Flat-field calibration of CCD detector for long trace profiler , 2007, SPIE Optical Engineering + Applications.

[50]  Valeriy V. Yashchuk,et al.  Development of a high-performance gantry system for a new generation of optical slope measuring profilers , 2012 .

[51]  S. Irick Improved measurement accuracy in a long trace profiler: Compensation for laser pointing instability , 1993 .

[52]  D. Kavaldjiev,et al.  SUBPIXEL SENSITIVITY MAP FOR A CHARGE-COUPLED DEVICE SENSOR , 1998 .

[53]  Josep Nicolas,et al.  Characterization of the error budget of Alba-NOM , 2013 .

[54]  Thomas Weiland,et al.  XFEL: The European X-Ray Free-Electron Laser - Technical Design Report , 2006 .

[55]  K. Bieren,et al.  Pencil Beam Interferometer For Aspherical Optical Surfaces , 1982 .

[56]  Valeriy V. Yashchuk,et al.  Autocollimators for Deflectometry: Current Status and Future Progress , 2009 .

[57]  M. Krause,et al.  The new PTB angle comparator , 1998 .

[58]  Riccardo Signorato,et al.  Structured slope errors on real x-ray mirrors: ray tracing versus experiment , 1997, Optics & Photonics.

[59]  M. Pisani,et al.  The new INRiM nanoangle generator , 2009 .

[60]  W. Tyler Estler,et al.  Uncertainty Analysis for Angle Calibrations Using Circle Closure , 1998, Journal of research of the National Institute of Standards and Technology.

[61]  Ralf D. Geckeler,et al.  A shearing-based method for the simultaneous calibration of angle measuring devices , 2014 .

[62]  Ali M. Khounsary,et al.  Performance of the APS optical slope measuring system , 2013 .

[63]  R. Hamid,et al.  Picometre displacement measurements using a differential Fabry–Perot optical interferometer and an x-ray interferometer , 2012 .

[64]  William T. Estler,et al.  An Advanced Angle Metrology System , 1993 .

[65]  Samuel K. Barber,et al.  Sub-microradian surface slope metrology with the ALS Developmental Long Trace Profiler , 2009 .

[66]  Ralf D. Geckeler,et al.  Modelling PTB's spatial angle autocollimator calibrator , 2013, Optical Metrology.

[67]  T. Yandayan,et al.  High precision small angle generator for realization of the SI unit of plane angle and calibration of high precision autocollimators , 2012 .