Non-Contact Simultaneous Measurements of Electrocardiogram and Respiratory Movements Using Capacitive Sheet Electrodes

In this study, we improved upon a high-input impedance monitoring device comprising capacitive sheet electrodes placed under a bed sheet, aiming to detect all ECG waves including T waves. To accomplish this, we widened the passbands for ECG and respiratory movement (RM) signals. However, because this widening may lower noise tolerance, we substituted conventional front-end buffers with buffers able to discharge DC potential. Further, the backside of the modi ed electrode was doubly shielded by using the electric potential of the buffer’s output and ground. We also introduced a virtual midpoint circuit to realize stable ECG and independent RM measurements of the chest and abdomen with the goal to detect obstructive sleep apnea in future applications. We tested our device on seven healthy adults during a six-hour sleep experiment. In all subjects, we detected T waves and observed ECG and RM signals in supine and lateral positions. In experiments in which respiratory rate was changed at regular intervals, the RM signal was synchronized with the simultaneously measured reference signal. The sensitivity and accuracy of R waves averaged over all subjects were 93.2% and 92.9%, respectively. Both averaged sensitivity and accuracy for T waves were 94.5%, while averaged accuracy of chest and abdominal RM was 70.9% and 74.0%, respectively. The results for ECG signals indicate that our improved system is applicable to unconstrained monitoring of vital signs in adults. Although there is still room for improvement in RM measurements, the results suggest that our system may also be used to detect variations in breathing.

[1]  Hitoshi Maruyama Risk Management-Vital Signs- , 2005 .

[2]  J. Mor,et al.  Prevalence and risk factors for prolonged QTc in a multiethnic cohort in rural Hawaii. , 2005, Clinical biochemistry.

[3]  Ueno Akinori,et al.  A study on breath detection algorithm for capacitive in-bed cardiorespiratory monitor , 2016 .

[4]  Yeon-Sik Noh,et al.  An improved algorithm for respiration signal extraction from electrocardiogram measured by conductive textile electrodes using instantaneous frequency estimation , 2008, Medical & Biological Engineering & Computing.

[5]  Akinori Ueno,et al.  Multi-layered fabric electrode for movement artifact reduction in capacitive ECG measurement , 2013, 2013 35th Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC).

[6]  Janko Drnovsek,et al.  Novel Methods for Noncontact Heart Rate Measurement: A Feasibility Study , 2014, IEEE Transactions on Instrumentation and Measurement.

[7]  Akinori Ueno,et al.  A Primary Study of Indirect ECG Monitor Embedded in a Bed for Home Health Care , 2007 .

[8]  Mikko Peltokangas,et al.  Night-Time EKG and HRV Monitoring With Bed Sheet Integrated Textile Electrodes , 2012, IEEE Transactions on Information Technology in Biomedicine.

[9]  M. Ishijima,et al.  Cardiopulmonary monitoring by textile electrodes without subject-awareness of being monitored , 1997, Medical and Biological Engineering and Computing.

[10]  Tohru Kiryu,et al.  Development of Unaware Detection System for Respiration and Heart Beats Using a Magnetic Sensor , 2000 .

[11]  Juan Antonio Gómez Galán,et al.  Design and Implementation of a Smart Sensor for Respiratory Rate Monitoring , 2014, Sensors.

[12]  H. Bazett,et al.  AN ANALYSIS OF THE TIME‐RELATIONS OF ELECTROCARDIOGRAMS. , 1997 .

[13]  Akira Murata Diagnosis and Treatment of Sleep Apnea Syndrome: Sleep Apnea Syndrome Triggers Life-Threatening Disorders During Sleep , 2007 .

[14]  Nitaro Shibata,et al.  Development of a QT measurement method by applying wavelet transform analysis and its correlation with QaT measurement , 2006 .

[15]  W. Flemons,et al.  Home diagnosis of sleep apnea: a systematic review of the literature. An evidence review cosponsored by the American Academy of Sleep Medicine, the American College of Chest Physicians, and the American Thoracic Society. , 2003, Chest.

[16]  Yong Gyu Lim,et al.  Flexible Capacitive Electrodes for Minimizing Motion Artifacts in Ambulatory Electrocardiograms , 2014, Sensors.

[17]  Akinori Ueno,et al.  Unconstrained monitoring of ECG and respiratory variation in infants with underwear during sleep using a bed-sheet electrode unit , 2008, 2008 30th Annual International Conference of the IEEE Engineering in Medicine and Biology Society.

[18]  Kwang Suk Park,et al.  Heart Rate Variability Monitoring during Sleep Based on Capacitively Coupled Textile Electrodes on a Bed , 2015, Sensors.