A temperature-dependent gain control system for improving the stability of Si-PM-based PET systems

The silicon-photomultiplier (Si-PM) is a promising photodetector for the development of new PET systems due to its small size, high gain and relatively low sensitivity to the static magnetic field. One drawback of the Si-PM is that it has significant temperature-dependent gain that poses a problem for the stability of the Si-PM-based PET system. To reduce this problem, we developed and tested a temperature-dependent gain control system for the Si-PM-based PET system. The system consists of a thermometer, analog-to-digital converter, personal computer, digital-to-analog converter and variable gain amplifiers in the weight summing board of the PET system. Temperature characteristics of the Si-PM array are measured and the calculated correction factor is sent to the variable gain amplifier. Without this correction, the temperature-dependent peak channel shifts of the block detector were -55% from 20 °C to 35 °C. With the correction, the peak channel variations were corrected within ±8%. The coincidence count rate of the Si-PM-based PET system was measured using a Na-22 point source while monitoring the room temperature. Without the correction, the count rate inversely changed with the room temperature by 10% for 1.5° C temperature changes. With the correction, the count rate variation was reduced to within 3.7%. These results indicate that the developed temperature-dependent gain control system can contribute to improving the stability of Si-PM-based PET systems.

[1]  T. Utsu,et al.  Growth and applications of Gd2SiO5: Ce scintillators , 1991 .

[2]  R. Mirzoyan,et al.  A test of silicon photomultipliers as readout for PET , 2005 .

[3]  M. Senda,et al.  A dual layer GSO PET system for small animal: K-PET II , 2007 .

[4]  In Chan Song,et al.  An Investigation Into the Use of Geiger-Mode Solid-State Photomultipliers for Simultaneous PET and MRI Acquisition , 2008, IEEE Transactions on Nuclear Science.

[5]  First results in the application of silicon photomultiplier matrices to small animal PET , 2009 .

[6]  Active control of the gain of a Silicon PhotoMultiplier , 2009 .

[7]  M. G. Bisogni,et al.  Energy, Timing and Position Resolution Studies With 16-Pixel Silicon Photomultiplier Matrices for Small Animal PET , 2009, IEEE Transactions on Nuclear Science.

[8]  Nan Zhang,et al.  SiPM performance in PET applications: An experimental and theoretical analysis , 2009, 2009 IEEE Nuclear Science Symposium Conference Record (NSS/MIC).

[9]  F. Beekman,et al.  A novel, SiPM-array-based, monolithic scintillator detector for PET , 2009, Physics in medicine and biology.

[10]  L. Cosentino,et al.  Features of Silicon Photo Multipliers: Precision Measurements of Noise, Cross-Talk, Afterpulsing, Detection Efficiency , 2009, IEEE Transactions on Nuclear Science.

[11]  M. G. Bagliesi,et al.  Scintillation and Cherenkov light detection with a 3 mm × 3 mm Silicon PhotoMultiplier , 2009 .

[12]  Tadashi Watabe,et al.  Development of a Si-PM-based high-resolution PET system for small animals , 2010, Physics in medicine and biology.

[13]  F. Beekman,et al.  LaBr3:Ce and SiPMs for time-of-flight PET: achieving 100 ps coincidence resolving time , 2010, Physics in medicine and biology.

[14]  Dieter Renker,et al.  Evaluation of Geiger-mode APDs for PET block detector designs , 2010, Physics in medicine and biology.

[15]  D. Renker,et al.  A G-APD based Camera for Imaging Atmospheric Cherenkov Telescopes , 2011 .