Prediction-Based Compensation for Gate On/Off Latency during Respiratory-Gated Radiotherapy

During respiratory-gated radiotherapy (RGRT), gate on and off latencies cause deviations of gating windows, possibly leading to delivery of low- and high-dose radiations to tumors and normal tissues, respectively. Currently, there are no RGRT systems that have definite tools to compensate for the delays. To address the problem, we propose a framework consisting of two steps: (1) multistep-ahead prediction and (2) prediction-based gating. For each step, we have devised a specific algorithm to accomplish the task. Numerical experiments were performed using respiratory signals of a phantom and ten volunteers, and our prediction-based RGRT system exhibited superior performance in more than a few signal samples. In some, however, signal prediction and prediction-based gating did not work well, maybe due to signal irregularity and/or baseline drift. The proposed approach has potential applicability in RGRT, and further studies are needed to verify and refine the constituent algorithms.

[1]  Claus Belka,et al.  Characteristics of gated treatment using an optical surface imaging and gating system on an Elekta linac , 2015, Radiation oncology.

[2]  Antti Sorjamaa,et al.  Multiple-output modeling for multi-step-ahead time series forecasting , 2010, Neurocomputing.

[3]  T. Araki,et al.  A simple respiratory indicator for irradiation during voluntary breath holding: a one-touch device without electronic materials. , 2010, Radiology.

[4]  Peter J. Nicholson,et al.  Algebraic Theory of Finite Fourier Transforms , 1971, Journal of computer and system sciences (Print).

[5]  R. Wiersma,et al.  Technical Note: High temporal resolution characterization of gating response time. , 2016, Medical physics.

[6]  Aaas News,et al.  Book Reviews , 1893, Buffalo Medical and Surgical Journal.

[7]  G. G. Stokes "J." , 1890, The New Yale Book of Quotations.

[8]  Steve B. Jiang,et al.  The management of respiratory motion in radiation oncology report of AAPM Task Group 76. , 2006, Medical physics.

[9]  R Mohan,et al.  Determining parameters for respiration-gated radiotherapy. , 2001, Medical physics.

[10]  H. Onishi,et al.  Technical Note: Evaluation of the latency and the beam characteristics of a respiratory gating system using an Elekta linear accelerator and a respiratory indicator device, Abches , 2018, Medical physics.

[11]  Petre Stoica,et al.  Spectral Analysis of Signals , 2009 .

[12]  H. Kubo,et al.  Respiration gated radiotherapy treatment: a technical study. , 1996, Physics in medicine and biology.

[13]  Y. Motai,et al.  Review: Prediction of Respiratory Motion , 2014 .