How to measure CT image quality: variations in CT-numbers, uniformity and low contrast resolution for a CT quality assurance phantom.

PURPOSE Quality assurance (QA) phantoms for testing different image quality parameters in computed tomography (CT) are commercially available. Such phantoms are also used as reference for acceptance in the specifications of CT-scanners. The aim of this study was to analyze the characteristics of the most commonly used QA phantom in CT: Catphan 500/504/600. METHODS Nine different phantoms were scanned on the same day, on one CT-scanner with the same parameter settings. Interphantom variations in CT-number values, image uniformity and low contrast resolution were evaluated for the phantoms. Comparisons between manual image analysis and results obtained from the automatic evaluation software QAlite were performed. RESULTS Some interphantom variations were observed in the low contrast resolution and the CT-number modules of the phantoms. Depending on the chosen regulatory framework, the variations in CT-numbers can be interpreted as substantial. The homogenous modules were found more invariable. However, the automatic image analysis software QAlite measures image uniformity differently than recommended in international standards, and will not necessarily give results in agreement with these standards. CONCLUSIONS It is important to consider the interphantom variations in relation to ones framework, and to be aware of which phantom is used to study CT-numbers and low contrast resolution for a specific scanner. Comparisons with predicted values from manual and acceptance values should be performed with care and consideration. If automatic software-based evaluations are to be used, users should be aware that large differences can exist for the image uniformity testing.

[1]  Jonathan Ophir,et al.  Signal-to-noise ratio, contrast-to-noise ratio and their trade-offs with resolution in axial-shear strain elastography. , 2007, Physics in medicine and biology.

[2]  Sasa Mutic,et al.  Performance evaluation of an 85-cm-bore X-ray computed tomography scanner designed for radiation oncology and comparison with current diagnostic CT scanners. , 2002, International journal of radiation oncology, biology, physics.

[3]  R. Reznek,et al.  The optimal imaging of adrenal tumours: a comparison of different methods. , 2007, Endocrine-related cancer.

[4]  A. C. Martinsen,et al.  Interphantom and interscanner variations for Hounsfield units—establishment of reference values for HU in a commercial QA phantom , 2010, Physics in medicine and biology.

[5]  Francis R Verdun,et al.  Iterative reconstruction methods in two different MDCT scanners: physical metrics and 4-alternative forced-choice detectability experiments--a phantom approach. , 2013, Physica medica : PM : an international journal devoted to the applications of physics to medicine and biology : official journal of the Italian Association of Biomedical Physics.

[6]  James S Babb,et al.  Multi-detector row CT attenuation measurements: assessment of intra- and interscanner variability with an anthropomorphic body CT phantom. , 2007, Radiology.

[7]  E C McCullough,et al.  The unreliability of CT numbers as absolute values. , 1982, AJR. American journal of roentgenology.

[8]  Allan Siperstein,et al.  Clinical utility of noncontrast computed tomography attenuation value (hounsfield units) to differentiate adrenal adenomas/hyperplasias from nonadenomas: Cleveland Clinic experience. , 2005, The Journal of clinical endocrinology and metabolism.

[9]  M. Bosniak,et al.  The current radiological approach to renal cysts. , 1986, Radiology.

[10]  E A Zerhouni,et al.  A standard phantom for quantitative CT analysis of pulmonary nodules. , 1983, Radiology.

[11]  David O Wilson,et al.  Quantitative Computed Tomography Analysis, Airflow Obstruction, and Lung Cancer in the Pittsburgh Lung Screening Study , 2011, Journal of thoracic oncology : official publication of the International Association for the Study of Lung Cancer.

[12]  C. Boffano,et al.  Performance Assessment of Four 64-Slice Computed Tomographic Devices for a Typical Clinical Protocol , 2011, Journal of computer assisted tomography.

[13]  Jacob Cohen,et al.  The Equivalence of Weighted Kappa and the Intraclass Correlation Coefficient as Measures of Reliability , 1973 .