Fabrication and preliminary assessment of a fiber optic-based relative humidity sensor for application in mechanical ventilation

During mechanical ventilation, specific devices are used to heat and to humidify gases delivered by mechanical ventilators. Basically, they must replace the function of the upper airways that are bypassed by tracheostomy or endotracheal tubes. Among others heated wire humidifiers (HWHs) are the most popular to accomplish this task. Their control is not based on the measure of relative humidity (RH) but on the gas temperature in a single point of the breathing circuit. Owing to the mentioned control, HWHs' performance shows some weaknesses in particular when they work under demanding ventilatory settings. Moreover, also environmental conditions (e.g., temperature) can affect HWHs' performance. The direct control of the air humidity may be beneficial to overcome these hurdles. The aim of this work is to develop a fiber optic sensors, based on fiber Bragg grating (FBG) technology for RH measurements in mechanical ventilation. It consists of a coated FBG, whose output changes with RH due to the volume expansion of the coating material. The proposed sensor has been fabricated with a simple coating procedure and it is coated with a cost effective polymer (i.e., agarose), which can be used in a wide range of RH values. An experimental set-up allowing the delivering of air with different amounts of vapour content (relative humidity from about 0% to saturation) has been prepared to perform a preliminary assessment of the sensor in a wide range of RH values. The proposed sensor shows promising results in terms of both capability to follow RH changes, as well as sensitivity and range of measurement. Future testing will assess in depth the metrological properties of the proposed sensor and will evaluate the potential of the sensor for RH monitoring during mechanical ventilation. The implementation of a feedback based on RH may be useful both to improve the performance of HWHs, to minimize the influence of external factors on the humidification process, as well as to pave the way to new strategy of control in this field.

[1]  Emiliano Schena,et al.  Flow measurement in mechanical ventilation: a review. , 2015, Medical engineering & physics.

[2]  T. Erdogan Fiber grating spectra , 1997 .

[3]  L. Brochard,et al.  Water content of delivered gases during non-invasive ventilation in healthy subjects , 2009, Intensive Care Medicine.

[4]  P. Rastogi,et al.  Relative humidity sensor with optical fiber Bragg gratings. , 2002, Optics letters.

[5]  S. Korposh,et al.  Simultaneous temperature and humidity measurements in a mechanical ventilator using an optical fibre sensor , 2016, European Workshop on Optical Fibre Sensors.

[6]  R. Branson The effects of inadequate humidity. , 1998, Respiratory care clinics of North America.

[7]  Y. Gong,et al.  Polyimide-coated fiber Bragg grating for relative humidity sensing , 2015 .

[8]  Emiliano Schena,et al.  Optical Fiber-Based MR-Compatible Sensors for Medical Applications: An Overview , 2013, Sensors.

[9]  Emiliano Schena,et al.  Design and Feasibility Assessment of a Magnetic Resonance-Compatible Smart Textile Based on Fiber Bragg Grating Sensors for Respiratory Monitoring , 2016, IEEE Sensors Journal.

[10]  Sergio Silvestri,et al.  Mathematical model and minimal measurement system for optimal control of heated humidifiers in neonatal ventilation. , 2010, Medical engineering & physics.

[11]  Emiliano Schena,et al.  Medical Smart Textiles Based on Fiber Optic Technology: An Overview , 2015, Journal of functional biomaterials.

[12]  Ruben D Restrepo,et al.  Humidification During Invasive and Noninvasive Mechanical Ventilation: 2012 , 2012, Respiratory Care.

[13]  Mark Hochmann,et al.  The effect of inlet gas temperatures on heated humidifier performance. , 2002, Journal of aerosol medicine : the official journal of the International Society for Aerosols in Medicine.

[14]  Sergio Silvestri,et al.  A high sensitivity fiber optic macro-bend based gas flow rate transducer for low flow rates: theory, working principle, and static calibration. , 2013, The Review of scientific instruments.

[15]  Yuliya Semenova,et al.  A fiber bend based humidity sensor with a wide linear range and fast measurement speed , 2012 .

[16]  T Mashimo,et al.  Performance of heated humidifiers with a heated wire according to ventilatory settings. , 2001, Journal of aerosol medicine : the official journal of the International Society for Aerosols in Medicine.

[17]  E Schena,et al.  A novel control strategy to improve the performances of heated wire humidifiers in artificial neonatal ventilation. , 2012, Physiological measurement.

[18]  J. Pillow,et al.  Oxygen, temperature and humidity of inspired gases and their influences on airway and lung tissue in near-term lambs , 2009, Intensive Care Medicine.

[19]  Arden L. Buck,et al.  New Equations for Computing Vapor Pressure and Enhancement Factor , 1981 .

[20]  P. Pelosi,et al.  Performance of heated wire humidifiers: an in vitro study. , 2007, Journal of critical care.

[21]  D. Formica,et al.  Smart Textile Based on Fiber Bragg Grating Sensors for Respiratory Monitoring: Design and Preliminary Trials , 2015, Biosensors.

[22]  Emiliano Schena,et al.  Performances of heated wire humidifiers during adult mechanical ventilation: Estimation of the amount of condensation , 2016, 2016 IEEE International Symposium on Medical Measurements and Applications (MeMeA).

[23]  E Schena,et al.  Mechanical ventilation with heated humidifiers: measurements of condensed water mass within the breathing circuit according to ventilatory settings. , 2013, Physiological measurement.

[24]  T. L. Yeo,et al.  Characterisation of a polymer-coated fibre Bragg grating sensor for relative humidity sensing , 2005 .

[25]  François Lellouche,et al.  Influence of ambient and ventilator output temperatures on performance of heated-wire humidifiers. , 2004, American journal of respiratory and critical care medicine.

[26]  G. Smaldone,et al.  Y-piece temperature and humidification during mechanical ventilation. , 2009, Respiratory care.

[27]  Umesh Tiwari,et al.  Fiber grating sensors in medicine: Current and emerging applications , 2011 .