Development and dosimetric assessment of a patient-specific elastic skin applicator for high-dose-rate brachytherapy.

PURPOSE The purpose of this study was to develop a patient-specific elastic skin applicator and to evaluate its dosimetric characteristics for high-dose-rate (HDR) brachytherapy. METHODS AND MATERIALS We simulated the treatment of a nonmelanoma skin cancer on the nose. An elastic skin applicator was manufactured by pouring the Dragon Skin (Smooth-On Inc., Easton, PA) with a shore hardness of 10A into an applicator mold. The rigid skin applicator was printed using high-impact polystyrene with a shore hardness of 73D. HDR plans were generated using a Freiburg Flap (FF) applicator and patient-specific rigid and elastic applicators. For dosimetric assessment, dose-volumetric parameters for target volume and normal organs were evaluated. Global gamma evaluations were performed, comparing film measurements and treatment planning system calculations with various gamma criteria. The 10% low-dose threshold was applied. RESULTS The V120% values of the target volume were 56.9%, 70.3%, and 70.2% for HDR plans using FF, rigid, and elastic applicators, respectively. The maximum doses of the right eyeball were 21.7 Gy, 20.5 Gy, and 20.5 Gy for the HDR plans using FF, rigid, and elastic applicators, respectively. The average gamma passing rates were 82.5% ± 1.5%, 91.6% ± 0.8%, and 94.8% ± 0.2% for FF, rigid, and elastic applicators, respectively, with 3%/3 mm criterion. CONCLUSIONS Patient-specific elastic skin applicator showed better adhesion to irregular or curved body surfaces, resulting in better agreement between planned and delivered dose distributions. The applicator suggested in this study can be effectively implemented clinically.

[1]  D. Peiffert,et al.  High dose rate brachytherapy with customized applicators for malignant facial skin lesions. , 2016, Cancer radiotherapie : journal de la Societe francaise de radiotherapie oncologique.

[2]  J F Williamson,et al.  Code of practice for brachytherapy physics: report of the AAPM Radiation Therapy Committee Task Group No. 56. American Association of Physicists in Medicine. , 1997, Medical physics.

[3]  Benjamin D. Harris,et al.  A feasibility study for using ABS plastic and a low-cost 3D printer for patient-specific brachytherapy mould design , 2015, Australasian Physical & Engineering Sciences in Medicine.

[4]  J. Carlson,et al.  New conformity indices based on the calculation of distances between the target volume and the volume of reference isodose. , 2014, The British journal of radiology.

[5]  R. L. Macdonald,et al.  Intrapatient study comparing 3D printed bolus versus standard vinyl gel sheet bolus for postmastectomy chest wall radiation therapy. , 2017, Practical radiation oncology.

[6]  Ravinder Nath,et al.  Guidelines by the AAPM and GEC-ESTRO on the use of innovative brachytherapy devices and applications: Report of Task Group 167. , 2016, Medical physics.

[7]  Mammo Yewondwossen,et al.  Clinical applications of 3-dimensional printing in radiation therapy. , 2017, Medical dosimetry : official journal of the American Association of Medical Dosimetrists.

[8]  Janusz Skowronek,et al.  Brachytherapy in the treatment of skin cancer: an overview , 2015, Postepy dermatologii i alergologii.

[9]  J. Fijuth HDR endobronchial brachytherapy in palliative and combined radical treatment of lung cancer , 2009, Journal of contemporary brachytherapy.

[10]  Sang‐wook Lee,et al.  Clinical Implementation of 3D Printing in the Construction of Patient Specific Bolus for Photon Beam Radiotherapy for Mycosis Fungoides , 2017 .

[11]  Tom Burrows,et al.  Introduction of novel 3D-printed superficial applicators for high-dose-rate skin brachytherapy. , 2017, Brachytherapy.

[12]  G. Baroni,et al.  3D-printed applicators for high dose rate brachytherapy: Dosimetric assessment at different infill percentage. , 2016, Physica medica : PM : an international journal devoted to the applications of physics to medicine and biology : official journal of the Italian Association of Biomedical Physics.

[13]  J. Seuntjens,et al.  Dose comparison between TG-43-based calculations and radiochromic film measurements of the Freiburg flap applicator used for high-dose-rate brachytherapy treatments of skin lesions. , 2017, Brachytherapy.

[14]  Yunzhi Ma,et al.  Validation of the Oncentra Brachy Advanced Collapsed cone Engine for a commercial (192)Ir source using heterogeneous geometries. , 2015, Brachytherapy.

[15]  Jong Min Park,et al.  The dosimetric effect of mixed‐energy IMRT plans for prostate cancer , 2011, Journal of applied clinical medical physics.

[16]  Zhigang Xu,et al.  Use of 3D printers to create a patient‐specific 3D bolus for external beam therapy , 2015, Journal of applied clinical medical physics.

[17]  Yonggang Huang,et al.  Conformal piezoelectric systems for clinical and experimental characterization of soft tissue biomechanics. , 2015, Nature materials.

[18]  N Hodapp,et al.  [The ICRU Report 83: prescribing, recording and reporting photon-beam intensity-modulated radiation therapy (IMRT)]. , 2012, Strahlentherapie und Onkologie : Organ der Deutschen Rontgengesellschaft ... [et al].

[19]  J. Williamson,et al.  Report of the Task Group 186 on model-based dose calculation methods in brachytherapy beyond the TG-43 formalism: current status and recommendations for clinical implementation. , 2012, Medical physics.

[20]  P Schaffer,et al.  3D printed plastics for beam modulation in proton therapy. , 2015, Physics in medicine and biology.

[21]  Peter Guion,et al.  Inverse treatment planning based on MRI for HDR prostate brachytherapy. , 2005, International journal of radiation oncology, biology, physics.

[22]  S. Gauden,et al.  HDR Brachytherapy for Superficial Non‐Melanoma Skin Cancers , 2013, Journal of medical imaging and radiation oncology.

[23]  Zoubir Ouhib,et al.  Aspects of dosimetry and clinical practice of skin brachytherapy: The American Brachytherapy Society working group report. , 2015, Brachytherapy.

[24]  T. Brüning,et al.  Improved method for cannula fixation for long-term intracerebral brain infusion , 2017, Journal of Neuroscience Methods.

[25]  J. Chard,et al.  Pediatric peri-operative fractionated high-dose-rate brachytherapy for recurrent Wilms’ tumor using a reconstructed Freiburg flap , 2016, Journal of contemporary brachytherapy.

[26]  Jung In Kim,et al.  Evaluation of Dosimetric Effect and Treatment Time by Plan Parameters for Endobronchial Brachytherapy , 2017 .

[27]  H. Jeon,et al.  Adjuvant external beam radiation and brachytherapy for vaginal resection margin positive cervical cancer , 2018, Radiation oncology journal.

[28]  Hong Kai Yap,et al.  Characterisation and evaluation of soft elastomeric actuators for hand assistive and rehabilitation applications , 2016, Journal of medical engineering & technology.

[29]  Kyubo Kim,et al.  Evaluation of treatment plans using various treatment techniques for the radiotherapy of cutaneous Kaposi's sarcoma developed on the skin of feet , 2014, Journal of applied clinical medical physics.

[30]  Chang Heon Choi,et al.  A Patient-Specific Polylactic Acid Bolus Made by a 3D Printer for Breast Cancer Radiation Therapy , 2016, PloS one.

[31]  T. Mate,et al.  Long-term outcome by risk factors using conformal high-dose-rate brachytherapy (HDR-BT) boost with or without neoadjuvant androgen suppression for localized prostate cancer. , 2004, International journal of radiation oncology, biology, physics.

[32]  K. Cho,et al.  Multimodal therapy for locally advanced prostate cancer: the roles of radiotherapy, androgen deprivation therapy, and their combination , 2017, Radiation oncology journal.