Non-contact left ventricular endocardial mapping in cardiac resynchronisation therapy

Background: Up to 30% of patients with heart failure do not respond to cardiac resynchronisation therapy (CRT). This may reflect placement of the coronary sinus lead in regions of slow conduction despite optimal positioning on current criteria. Objectives: To characterise the effect of CRT on left ventricular activation using non-contact mapping and to examine the electrophysiological factors influencing optimal left ventricular lead placement. Methods and results: 10 patients implanted with biventricular pacemakers were studied. In six, the coronary sinus lead was found to be positioned in a region of slow conduction with an average conduction velocity of 0.4 m/s, v 1.8 m/s in normal regions (p < 0.02). Biventricular pacing with the left ventricle paced 32 ms before the right induced the optimal mean velocity time integral and timing for fusion of depolarisation wavefronts from the right and left ventricular pacing sites. Pacing outside regions of slow conduction decreased left ventricular activation time and increased cardiac output and dP/dtmax significantly. Conclusions: In patients undergoing CRT for heart failure, non-contact mapping can identify regions of slow conduction. Significant haemodynamic improvements can occur when the site of left ventricular pacing is outside these slow conduction areas. Failure of CRT to produce clinical benefits may reflect left ventricular lead placement in regions of slow conduction which can be overcome by pacing in more normally activating regions.

[1]  AngeloAuricchio,et al.  Effect of Pacing Chamber and Atrioventricular Delay on Acute Systolic Function of Paced Patients With Congestive Heart Failure , 1999 .

[2]  C. Grines,et al.  Functional abnormalities in isolated left bundle branch block. The effect of interventricular asynchrony. , 1989, Circulation.

[3]  Optimization of ventricular pacing: where should we implant the leads? , 1999, Journal of the American College of Cardiology.

[4]  D. Kass,et al.  Mechanical dyssynchrony in dilated cardiomyopathy with intraventricular conduction delay as depicted by 3D tagged magnetic resonance imaging. , 2000, Circulation.

[5]  Functional Abnormalities in Isolated Left Bundle Branch Block , 1989 .

[6]  Ruben Coronel,et al.  Activation Delay After Premature Stimulation in Chronically Diseased Human Myocardium Relates to the Architecture of Interstitial Fibrosis , 2001, Circulation.

[7]  Arma,et al.  EFFECTS OF MULTISITE BIVENTRICULAR PACING IN PATIENTS WITH HEART FAILURE AND INTRAVENTRICULAR CONDUCTION DELAY , 2001 .

[8]  A. Weyland,et al.  Comparison of cardiac output assessed by pulse-contour analysis and thermodilution in patients undergoing minimally invasive direct coronary artery bypass grafting. , 1999, Journal of cardiothoracic and vascular anesthesia.

[9]  N. Bayley,et al.  Failure , 1890, The Hospital.

[10]  J. Deharo,et al.  Multisite Pacing as a Supplemental Treatment of Congestive Heart Failure: Preliminary Results of the Medtronic Inc. InSync Study , 1998, Pacing and clinical electrophysiology : PACE.

[11]  A. Lazarus,et al.  Multisite Pacing for End‐Stage Heart Failure: Early Experience , 1996, Pacing and clinical electrophysiology : PACE.

[12]  N. Peters,et al.  Simultaneous endocardial mapping in the human left ventricle using a noncontact catheter: comparison of contact and reconstructed electrograms during sinus rhythm. , 1998, Circulation.

[13]  D. Gibson,et al.  Effect of left bundle branch block on diastolic function in dilated cardiomyopathy. , 1991, British heart journal.

[14]  A. Kadish,et al.  Mapping of atrial activation with a noncontact, multielectrode catheter in dogs. , 1999, Circulation.

[15]  B. Reichart,et al.  Continuous cardiac output by femoral arterial thermodilution calibrated pulse contour analysis: comparison with pulmonary arterial thermodilution. , 1999, Critical care medicine.

[16]  Gregory S. Nelson,et al.  Left Ventricular or Biventricular Pacing Improves Cardiac Function at Diminished Energy Cost in Patients With Dilated Cardiomyopathy and Left Bundle-Branch Block , 2000, Circulation.

[17]  N S Peters,et al.  Characterization of the anatomy and conduction velocities of the human right atrial flutter circuit determined by noncontact mapping. , 2001, Journal of the American College of Cardiology.

[18]  C. H. Chen,et al.  Improved left ventricular mechanics from acute VDD pacing in patients with dilated cardiomyopathy and ventricular conduction delay. , 1999, Circulation.

[19]  J. Daubert,et al.  Effects of multisite biventricular pacing in patients with heart failure and intraventricular conduction delay. , 2001, The New England journal of medicine.

[20]  Andrew P. Kramer,et al.  Predictors of systolic augmentation from left ventricular preexcitation in patients with dilated cardiomyopathy and intraventricular conduction delay. , 2000, Circulation.

[21]  J. Daubert,et al.  Electrocardiographic predictive factors of long-term clinical improvement with multisite biventricular pacing in advanced heart failure. , 1999, The American journal of cardiology.

[22]  D. Durrer,et al.  Total Excitation of the Isolated Human Heart , 1970, Circulation.

[23]  J. Spinelli,et al.  Cardiac pacing in heart failure patients with left bundle branch block: impact of pacing site for optimizing left ventricular resynchronization. , 2000, Italian heart journal : official journal of the Italian Federation of Cardiology.

[24]  W. Braham,et al.  Cardiac Resynchronization in Chronic Heart Failure , 2002 .

[25]  J. Blanc,et al.  Evaluation of different ventricular pacing sites in patients with severe heart failure: results of an acute hemodynamic study. , 1997, Circulation.

[26]  N. Peters,et al.  Mapping and ablation of ventricular tachycardia with the aid of a non-contact mapping system , 1999, Heart.

[27]  S. Anker,et al.  Intraventricular conduction delay: a prognostic marker in chronic heart failure. , 1999, International journal of cardiology.

[28]  G. Newton,et al.  Electrical and hemodynamic correlates of the maximal rate of pressure increase in the human left ventricle. , 1999, Journal of cardiac failure.

[29]  E. Botvinick,et al.  Ventricular contraction abnormalities in dilated cardiomyopathy: effect of biventricular pacing to correct interventricular dyssynchrony. , 2000, Journal of the American College of Cardiology.

[30]  A. Liebold,et al.  Continuous cardiac output measurement: pulse contour analysis vs thermodilution technique in cardiac surgical patients. , 1999, British journal of anaesthesia.

[31]  E Fleck,et al.  Effect of Resynchronization Therapy Stimulation Site on the Systolic Function of Heart Failure Patients , 2001, Circulation.

[32]  Milton Packer,et al.  Cardiac resynchronization in chronic heart failure. , 2002, The New England journal of medicine.

[33]  W. Kim,et al.  Three dimensional echocardiography documents haemodynamic improvement by biventricular pacing in patients with severe heart failure , 2001, Heart.