Analysis of resting noise characteristics of three EIT systems in order to compare suitability for time difference imaging with scalp electrodes during epileptic seizures

Electrical impedance tomography measurements in clinical applications are limited by an undesired noise component. We have investigated the noise in three systems suitable for imaging epileptic seizures, the UCH Mark1b, UCH Mark2.5 and KHU Mark1 16 channel, at applied frequencies in three steps from 1 to 100 kHz, by varying load impedance, single terminal or multiplexed measurements, and in test objects of increasing complexity from a resistor to a saline filled tank and human volunteer. The noise was white, and increased from about 0.03% rms on the resistor to 0.08% on the human; it increased with load but was independent of use of the multiplexer. The KHU Mark1 delivered the best performance with noise spectra of about 0.02%, which could be further reduced by averaging to a level where reliable imaging of changes of about 0.1% estimated during epileptic seizures appears plausible.

[1]  Dr D. Jennings,et al.  Design of an electrical impedance tomography phantom using active elements , 2000, Medical and Biological Engineering and Computing.

[2]  Marom Bikson,et al.  Tissue resistance changes and the profile of synchronized neuronal activity during ictal events in the low-calcium model of epilepsy. , 2004, Journal of neurophysiology.

[3]  R Bayford,et al.  Design and calibration of a compact multi-frequency EIT system for acute stroke imaging. , 2006, Physiological measurement.

[4]  H Griffiths,et al.  A Cole phantom for EIT. , 1995, Physiological measurement.

[5]  C D Binnie,et al.  Factors limiting the application of electrical impedance tomography for identification of regional conductivity changes using scalp electrodes during epileptic seizures in humans , 2006, Physiological measurement.

[6]  R H Smallwood,et al.  Mk3.5: a modular, multi-frequency successor to the Mk3a EIS/EIT system. , 2001, Physiological measurement.

[7]  T Schröder,et al.  Quantitative Evaluation of the Performance of Different Electrical Tomography Devices - Quantitative Evaluierung der Systemeigenschaften verschiedener Elektrotomographie-Geräte , 2001, Biomedizinische Technik. Biomedical engineering.

[8]  D S Holder,et al.  Identification of a suitable current waveform for acute stroke imaging , 2006, Physiological measurement.

[9]  P. Hua,et al.  Measuring lung resistivity using electrical impedance tomography , 1992, IEEE Transactions on Biomedical Engineering.

[10]  H Griffiths A phantom for electrical impedance tomography. , 1988, Clinical physics and physiological measurement : an official journal of the Hospital Physicists' Association, Deutsche Gesellschaft fur Medizinische Physik and the European Federation of Organisations for Medical Physics.

[11]  T Schröder,et al.  A simple method to check the dynamic performance of electrical impedance tomography systems. , 2000, Physiological measurement.

[12]  B H Blott,et al.  EIT data noise evaluation in the clinical environment. , 1996, Physiological measurement.

[13]  A Tizzard,et al.  Generating accurate finite element meshes for the forward model of the human head in EIT , 2005, Physiological measurement.

[14]  B H Brown,et al.  Noise equalization within EIT images. , 1994, Physiological measurement.

[15]  W. R. Bennett,et al.  Spectra of quantized signals , 1948, Bell Syst. Tech. J..

[16]  Lior Horesh,et al.  Specification and calibration of a multi-frequency MEIT system for stroke , 2005 .

[17]  Alejandro F. Frangi,et al.  Propagation of measurement noise through backprojection reconstruction in electrical impedance tomography , 2002, IEEE Transactions on Medical Imaging.

[18]  C. Grimbergen,et al.  Investigation into the origin of the noise of surface electrodes , 2002, Medical and Biological Engineering and Computing.

[19]  J. Schadé,et al.  Changes in the electrical conductivity of cerebral cortex during seizure activity. , 1962, Experimental neurology.

[20]  R H Bayford,et al.  Multi-frequency electrical impedance tomography (EIT) of the adult human head: initial findings in brain tumours, arteriovenous malformations and chronic stroke, development of an analysis method and calibration , 2006, Physiological measurement.

[21]  B H Brown,et al.  The Sheffield data collection system. , 1987, Clinical physics and physiological measurement : an official journal of the Hospital Physicists' Association, Deutsche Gesellschaft fur Medizinische Physik and the European Federation of Organisations for Medical Physics.

[22]  Lior Horesh,et al.  A feasibility study for imaging of epileptic seizures by EIT using a realistic FEM of the head 12th Conf. on Biomedical Application of EIT (Seoul, Korea) , 2006 .

[23]  W. R. Adey,et al.  Impedance Changes during Epileptic Seizures , 1966, Epilepsia.

[24]  D S Holder,et al.  Some practical biological phantoms for calibrating multifrequency electrical impedance tomography. , 1996, Physiological measurement.

[25]  I Frerichs,et al.  Electrical impedance tomography (EIT) in applications related to lung and ventilation: a review of experimental and clinical activities. , 2000, Physiological measurement.

[26]  R H Bayford,et al.  Design and performance of the UCLH mark 1b 64 channel electrical impedance tomography (EIT) system, optimized for imaging brain function. , 2002, Physiological measurement.