Digital radiology using active matrix readout of amorphous selenium: theoretical analysis of detective quantum efficiency.

A flat-panel x-ray imaging detector using a layer of amorphous selenium (a-Se) for direct conversion of x rays (to charge) and an active matrix for self-scanned readout is being investigated for digital radiology. A theoretical analysis of the spatial frequency dependent detective quantum efficiency (DQE(f)) of the self-scanned a-Se detector is performed based on a model of signal and noise propagation in a cascaded imaging system. Because of the high intrinsic resolution of a-Se and the pixelated active matrix readout method, such detectors are inherently undersampled and aliasing is present. The presampling modulation transfer function (MTF) and aliased noise power spectrum (NPS) of the detector were used in the analysis of DQE(f). It is proven that the aliased NPS for the self-scanned a-Se detectors is white. Since the shape of DQE(f) is determined by the ratio of MTF squared and the NPS, the shape of DQE(f) follows the square of the presampling MTF of the detector as a result of the white NPS. The analysis also shows that DQE(0) is proportional to the pixel fill factor, i.e., the fraction of each pixel area used for image charge collection. The DQE analysis is applied to detector parameters for three x-ray imaging applications: mammography, chest radiography, and fluoroscopy. The effects of pixel fill factor, imaging geometry (i.e., incident angle of x rays), and various sources of electronic noise on the detector DQE(f) are discussed. Strategies for maximizing detector DQE for each x-ray imaging application are proposed.

[1]  Martin J. Yaffe,et al.  Noise in hybrid photodiode array: CCD x-ray image detectors for digital mammography , 1996, Medical Imaging.

[2]  John A. Rowlands,et al.  Digital radiology using self-scanned readout of amorphous selenium: design considerations for mammography , 1995, Medical Imaging.

[3]  J A Rowlands,et al.  X-ray imaging using amorphous selenium: inherent spatial resolution. , 1995, Medical physics.

[4]  J T Dobbins,et al.  Effects of undersampling on the proper interpretation of modulation transfer function, noise power spectra, and noise equivalent quanta of digital imaging systems. , 1995, Medical physics.

[5]  A Fenster,et al.  A spatial-frequency dependent quantum accounting diagram and detective quantum efficiency model of signal and noise propagation in cascaded imaging systems. , 1994, Medical physics.

[6]  Denny L. Y. Lee,et al.  Discussion on resolution and dynamic range of Se-TFT direct digital radiographic detector , 1996, Medical Imaging.

[7]  Gerhard Spekowius,et al.  Simulation of the imaging performance of x-ray image intensifier/TV camera chains , 1995, Medical Imaging.

[8]  M. Rabbani,et al.  Detective quantum efficiency of imaging systems with amplifying and scattering mechanisms. , 1987, Journal of the Optical Society of America. A, Optics and image science.

[9]  Aaron Fenster,et al.  Effect of finite detector-element width on the spatial-frequency-dependent detective quantum efficiency , 1995, Medical Imaging.

[10]  M J Yaffe,et al.  Effect of various noise sources on the detective quantum efficiency of phosphor screens. , 1990, Medical physics.

[11]  Larry E. Antonuk,et al.  Theoretical performance of amorphous silicon imagers in diagnostic radiology , 1996, Medical Imaging.

[12]  Youming Li,et al.  High-resolution amorphous silicon image sensor , 1996, Medical Imaging.

[13]  M. Giger,et al.  Investigation of basic imaging properties in digital radiography. 2. Noise Wiener spectrum. , 1984, Medical physics.

[14]  J A Rowlands,et al.  X-ray imaging using amorphous selenium: determination of x-ray sensitivity by pulse height spectroscopy. , 1992, Medical physics.

[15]  R. K. Swank Absorption and noise in x‐ray phosphors , 1973 .

[16]  J A Rowlands,et al.  X-ray imaging with amorphous selenium: detective quantum efficiency of photoconductive receptors for digital mammography. , 1995, Medical physics.

[17]  R. K. Swank Measurement of absorption and noise in an x‐ray image intensifier , 1974 .

[18]  P C Bunch,et al.  Analysis of the detective quantum efficiency of a radiographic screen-film combination. , 1987, Journal of the Optical Society of America. A, Optics and image science.

[19]  P. Munro,et al.  A quantum accounting and detective quantum efficiency analysis for video-based portal imaging. , 1997, Medical physics.

[20]  M J Yaffe,et al.  Model of the spatial-frequency-dependent detective quantum efficiency of phosphor screens. , 1990, Medical physics.

[21]  Norbert Conrads,et al.  Fluoroscopic x-ray imaging with amorphous silicon thin-film arrays , 1994, Medical Imaging.

[22]  Thierry Ducourant,et al.  Amorphous silicon x-ray image sensor , 1996, Medical Imaging.

[23]  J A Rowlands,et al.  X-ray imaging using amorphous selenium: photoinduced discharge (PID) readout for digital general radiography. , 1995, Medical physics.

[24]  J A Rowlands,et al.  X-ray imaging using amorphous selenium: feasibility of a flat panel self-scanned detector for digital radiology. , 1995, Medical physics.

[25]  J Yorkston,et al.  Demonstration of megavoltage and diagnostic x-ray imaging with hydrogenated amorphous silicon arrays. , 1992, Medical physics.

[26]  J A Rowlands,et al.  Videofluorography: the role of temporal averaging. , 1984, Medical physics.

[27]  R. F. Wagner,et al.  Absolute measures of physical image quality: measurement and application to radiographic magnification. , 1982, Medical physics.

[28]  Robert G. Gould,et al.  Specification, acceptance testing and quality control of diagnostic x-ray imaging equipment , 1994 .

[29]  K Doi,et al.  Investigation of basic imaging properties in digital radiography. I. Modulation transfer function. , 1984, Medical physics.

[30]  Ian A. Cunningham Degradation of the detective quantum efficiency due to a nonunity detector fill factor , 1997, Medical Imaging.

[31]  Majid Rabbani,et al.  Analysis of signal and noise propagation for several imaging mechanisms , 1989 .

[32]  Walter Dr. Hillen,et al.  Signal-to-noise performance in cesium iodide x-ray fluorescent screens , 1991, Medical Imaging.

[33]  Denny L. Y. Lee,et al.  New digital detector for projection radiography , 1995, Medical Imaging.

[34]  John A. Rowlands,et al.  Large-area solid state detector for radiology using amorphous selenium , 1992, Medical Imaging.

[35]  H. Saunders,et al.  Signal Recovery from Noise in Electronic Instrumentation , 1985 .