The Accuracy of Portable Ultrasonography to Diagnose Fractures in an Austere Environment

Background. Previous literature has shown the ability of ultrasonography technology to accurately assess orthopedic pathology. Over the past two decades, the use of ultrasound in the prehospital setting has become an important tool for triage, diagnosis, and treatment, especially in austere and remote environments that lack appropriate radiography capability and logistical support. The purpose of our study was to assess the accuracy of ultrasound in the austere, combat environment for diagnosis of orthopedic fracture. Methods. We conducted a longitudinal, prospective, observational study on patients presenting with suspected closed fractures using a digital handheld ultrasound device. All patients presenting with suspected fracture underwent an ultrasound examination by a board-certified emergency medicine physician credentialed in emergency ultrasonography. Patients were then categorized into ultrasound-positive and ultrasound-negative groups. Results. A total of 44 subjects underwent ultrasound examination for suspected fractures. There were initially 12 (27%) positive and 32 (73%) negative scans. Of the initial 12 positive scans, ten had a true fracture verified by plain radiography. Ultrasonography yielded an overall sensitivity of 100% and a specificity of 94%. Only four patients with an initial negative ultrasound scan continued to have clinical symptoms for more than three days and were found to have no evidence of fracture by radiograph. Conclusions. Our data show that use of ultrasound by an experienced clinician in the austere environment can be performed accurately and can possibly prevent unnecessary evacuations for suspected fractures requiring radiographic verification. The purpose of our study was to assess the accuracy of ultrasound examination in the austere, combat environment for diagnosis of orthopedic fracture.

[1]  Y. Itzchak,et al.  Sonographic detection of occult bone fractures , 1988, Pediatric Radiology.

[2]  W. Grechenig,et al.  Scope and limitations of ultrasonography in the documentation of fractures – an experimental study , 1998, Archives of Orthopaedic and Trauma Surgery.

[3]  M. Stuart,et al.  Ultrasound assessment and conservative management of inversion injuries of the ankle in children: plaster of Paris versus Tubigrip. , 1996, The Journal of bone and joint surgery. British volume.

[4]  E. McEnery Fracture of the clavicle in the newborn. , 1945, American journal of diseases of children.

[5]  G. Steiner,et al.  The value of ultrasound in the assessment of bone. , 1992, The British journal of radiology.

[6]  J. Mcmanus,et al.  Use of Ultrasonography to Avoid an Unnecessary Procedure in the Prehospital Combat Environment: A Case Report , 2006, Prehospital emergency care : official journal of the National Association of EMS Physicians and the National Association of State EMS Directors.

[7]  D. Lichtenstein,et al.  Ultrasound diagnosis of occult pneumothorax* , 2005, Critical care medicine.

[8]  M. Blaivas,et al.  Ultrasound-guided drainage of peritonsillar abscess by the emergency physician. , 2003, The American journal of emergency medicine.

[9]  A G Randolph,et al.  Ultrasound guidance for placement of central venous catheters: a meta-analysis of the literature. , 1996, Critical care medicine.

[10]  J. Shieh,et al.  Sonographic detection of occult fractures in the foot and ankle , 1999, Journal of clinical ultrasound : JCU.

[11]  J. Jacobson,et al.  The Use of Ultrasound in Evaluating Orthopaedic Trauma Patients , 2005, The Journal of the American Academy of Orthopaedic Surgeons.

[12]  J. Bar-ziv,et al.  Fracture of the clavicle in the newborn. An ultrasound diagnosis. , 1988, Journal of ultrasound in medicine : official journal of the American Institute of Ultrasound in Medicine.

[13]  A. Hargens,et al.  Intracranial pressure dynamics assessed by noninvasive ultrasound during 30 days of bed rest. , 2005, Aviation, space, and environmental medicine.

[14]  D. Hamilton,et al.  Advanced ultrasonic diagnosis of extremity trauma: the FASTER examination. , 2002, Journal of Trauma.