Normal and transplanted rat kidneys: diffusion MR imaging at 7 T.

PURPOSE To investigate the feasibility of obtaining reproducible apparent diffusion coefficient (ADC) maps of normal rat kidneys by using respiratory-triggered spin-echo diffusion-weighted magnetic resonance (MR) imaging, to investigate the sensitivity of ADC maps in the evaluation of renal blood flow, and to use this technique to monitor acute graft rejection in transplanted rat kidneys. MATERIALS AND METHODS Spin-echo diffusion-weighted MR imaging measurements were performed in 20 normal rats and nine rats that had undergone transplantation (six rats had received allografts; three had received isografts) at 7 T. To evaluate the effect of alteration in blood flow and water transport function, angiotensin II was infused in six normal rats and a series of spin-echo diffusion-weighted MR images was obtained at five time points. Transplanted kidneys were monitored by obtaining spin-echo diffusion-weighted MR images and gradient-echo MR images every 2 hours for 8 hours on postoperative day 4. Statistical analysis was performed with repeated-measures multivariate analysis of variance and the paired t test. RESULTS No significant differences in ADC values were observed between right and left kidneys in all three orthogonal directions; however, a small difference was observed between the cortex and medulla. ADC values in the cephalocaudal and mediolateral directions were higher than those in the anteroposterior direction (P <.01 for all). ADC values in the cortex and medulla decreased significantly (by >35%, P <.01) during angiotensin II-induced reduction in renal blood flow. No significant signal intensity change was observed between native and transplanted kidneys on gradient-echo MR images. Allografts exhibited decreased ADC values (P <.01) and isografts exhibited similar ADC values compared with native kidneys. CONCLUSION These findings suggest that reproducible renal ADC maps can be obtained in rats by using spin-echo diffusion-weighted MR imaging at 7 T. Spin-echo diffusion-weighted MR imaging may have potential as a noninvasive tool for monitoring early graft rejection after kidney transplantation.

[1]  K. Hendrich,et al.  Perfusion quantitation in transplanted rat kidney by MRI with arterial spin labeling. , 1998, Kidney international.

[2]  C. Lorenz,et al.  Magnetic resonance diffusion/perfusion phantom experiments , 1991, Magnetic resonance in medicine.

[3]  T E Conturo,et al.  Diffusion MRI: Precision, accuracy and flow effects , 1995, NMR in biomedicine.

[4]  P. Grenier,et al.  MR imaging of intravoxel incoherent motions: application to diffusion and perfusion in neurologic disorders. , 1986, Radiology.

[5]  Qing Ye,et al.  USPIO‐enhanced dynamic MRI: Evaluation of normal and transplanted rat kidneys , 2001, Magnetic resonance in medicine.

[6]  B. Siewert,et al.  Abdominal diffusion mapping with use of a whole-body echo-planar system. , 1994, Radiology.

[7]  E. Dumont,et al.  Hemodynamic Effect of Iodinated High-Viscosity Contrast Medium in the Rat Kidney: A Diffusion-Weighted MRI Feasibility Study , 2000, Investigative radiology.

[8]  F. Ebner,et al.  Functional magnetic resonance imaging of human renal allografts during the post-transplant period: preliminary observations. , 1997, Magnetic resonance imaging.

[9]  T. Ichikawa,et al.  Diffusion-weighted MR imaging with single-shot echo-planar imaging in the upper abdomen: preliminary clinical experience in 61 patients , 1999, Abdominal Imaging.

[10]  Y Zhang,et al.  Magnetic resonance imaging detection of rat renal transplant rejection by monitoring macrophage infiltration. , 2000, Kidney international.

[11]  R. Ordidge,et al.  Correction of motional artifacts in diffusion-weighted MR images using navigator echoes. , 1994, Magnetic resonance imaging.

[12]  R R Edelman,et al.  Functional imaging of the kidney by means of measurement of the apparent diffusion coefficient. , 1994, Radiology.

[13]  M. Reiser,et al.  Contrast-enhanced MR angiography in patients after kidney transplantation , 2001, European Radiology.

[14]  M. Rudin,et al.  Magnetic resonance imaging for the evaluation of rejection of a kindney allograft in the rat , 1996, Transplant international : official journal of the European Society for Organ Transplantation.

[15]  Y. Yamashita,et al.  Measurement of the apparent diffusion coefficient in diffuse renal disease by diffusion‐weighted echo‐planar MR imaging , 1999, Journal of magnetic resonance imaging : JMRI.

[16]  D. S. Williams,et al.  Perfusion imaging of the rat kidney with MR. , 1994, Radiology.

[17]  I. Ohashi,et al.  Anisotropic diffusion in kidney: Apparent diffusion coefficient measurements for clinical use , 2000, Journal of magnetic resonance imaging : JMRI.

[18]  M. Reiser,et al.  Diffusion-weighted imaging of tumor recurrencies and posttherapeutical soft-tissue changes in humans , 2001, European Radiology.

[19]  S. Riederer,et al.  Analysis of T2 limitations and off‐resonance effects on spatial resolution and artifacts in echo‐planar imaging , 1990, Magnetic resonance in medicine.

[20]  C T Moonen,et al.  Diffusion tensor MRI of the human kidney , 2001, Journal of magnetic resonance imaging : JMRI.

[21]  H. Hricak,et al.  Complex posttransplantation abnormalities of renal allografts: evaluation with MR imaging. , 1999, Radiology.

[22]  Donald S. Williams,et al.  In vivo detection of acute rat renal allograft rejection by MRI with USPIO particles. , 2002, Kidney international.

[23]  A. Aisen,et al.  Feasibility of MR Diffusion Studies in the Kidney , 1995, Journal of magnetic resonance imaging : JMRI.

[24]  I. Yamada,et al.  Diffusion coefficients in abdominal organs and hepatic lesions: evaluation with intravoxel incoherent motion echo-planar MR imaging. , 1999, Radiology.

[25]  D. Le Bihan,et al.  Separation of diffusion and perfusion in intravoxel incoherent motion MR imaging. , 1988, Radiology.