Optimizing and validating an electromagnetic tracker in a human performance laboratory

Measurement errors have previously been observed using electromagnetic motion trackers in applied laboratories. The aims of this study were to optimize the layout of a human performance laboratory for assessing ergometer rowing technique, and to assess the precision and repeatability of measured rotations and trajectories using the Flock of Birds electromagnetic tracker. Four experiments investigated system performance over a large experimental volume: optimization of laboratory space, repeatability of laboratory layout, precision of measured rotations, and repeatability of measured displacements. Measurement accuracy was influenced by varying the global position of the system transmitter; results suggested a correlation with increasing distance between the electromagnetic source and equivalent sensors. Bringing the transmitter or sensors into closer proximity of metallic items may be another source of measurement error. An optimal location for the transmitter was identified, into which the transmitter was repositioned with good repeatability. Measurements were not negativelyaffected by the presence of a rowing ergometer in the experimental volume. Induced sensor rotations were reconstructed with high precision, and the system calculated small changes in sensor displacement with good repeatability. The system is a suitable technology for measuring the trajectory and rotation of moving body segments in applied human movement laboratories.

[1]  G A Dumas,et al.  Calibration of position and angular data from a magnetic tracking device. , 2000, Journal of biomechanics.

[2]  A M Bull,et al.  Measuring spinal motion in rowers: the use of an electromagnetic device. , 2000, Clinical biomechanics.

[3]  A A Amis,et al.  Accuracy of an electromagnetic measurement device and application to the measurement and description of knee joint motion , 1998, Proceedings of the Institution of Mechanical Engineers. Part H, Journal of engineering in medicine.

[4]  W. Richard Fright,et al.  The Effects of Metals and Interfering Fields on Electromagnetic Trackers , 1998, Presence.

[5]  A J Tate,et al.  Evaluation and calibration of an electromagnetic tracking device for biomechanical analysis of lifting tasks. , 2002, Journal of biomechanics.

[6]  Anthony M J Bull,et al.  Modelling multivariate biomechanical measurements of the spine during a rowing exercise. , 2003, Clinical biomechanics.

[7]  S. Jaberzadeh,et al.  Accuracy of an electromagnetic tracking device for measuring hip joint kinematics during gait: effects of metallic total hip replacement prosthesis, source-sensor distance and sensor orientation , 2005, Australasian Physics & Engineering Sciences in Medicine.

[8]  Suzanne LaScalza,et al.  Effect of metal and sampling rate on accuracy of Flock of Birds electromagnetic tracking system. , 2003, Journal of biomechanics.

[9]  F. V. D. van der Helm,et al.  Calibration of the "Flock of Birds" electromagnetic tracking device and its application in shoulder motion studies. , 1999, Journal of biomechanics.

[10]  James A. Johnson,et al.  Accuracy of an electromagnetic tracking device: a study of the optimal range and metal interference. , 1996, Journal of biomechanics.

[11]  P M M Cashman,et al.  Kinematics of spinal motion during prolonged rowing. , 2003, International journal of sports medicine.

[12]  T D Brown,et al.  Factors influencing accuracy of screw displacement axis detection with a D.C.-based electromagnetic tracking system. , 1998, Journal of biomechanical engineering.

[13]  J Cholewicki,et al.  Methodological considerations for using inductive sensors (3SPACE ISOTRAK) to monitor 3-D orthopaedic joint motion. , 1997, Clinical biomechanics.