Effect of Signals on the Encapsulation Performance of Parylene Coated Platinum Tracks for Active Medical Implants

Platinum is widely used as the electrode material for implantable devices. Owing to its high biostability and corrosion resistivity, platinum could also be used as the main metallization for tracks in active implants. Towards this goal, in this work we investigate the stability of parylene-coated Pt tracks using passive and active tests. The test samples in this study are Pt-on-SiO2 interdigitated comb structures. During testing all samples were immersed in saline for 150 days; for passive testing, the samples were left unbiased, whilst for active testing, samples were exposed to two different stress signals: a 5 V DC and a 5 Vp 500 pulses per second biphasic signal. All samples were monitored over time using impedance spectroscopy combined with optical inspection. After the first two weeks of immersion, delamination spots were observed on the Pt tracks for both passive and actively tested samples. Despite the delamination spots, the unbiased samples maintained high impedances until the end of the study. For the actively stressed samples, two different failure mechanisms were observed which were signal related. DC stressed samples showed severe parylene cracking mainly due to the electrolysis of the condensed water. Biphasically stressed samples showed gradual Pt dissolution and migration. These results contribute to a better understanding of the failure mechanisms of Pt tracks in active implants and suggest that new testing paradigms may be necessary to fully assess the long-term reliability of these devices.

[1]  Christopher E. Larson,et al.  Techniques and Considerations in the Microfabrication of Parylene C Microelectromechanical Systems , 2018, Micromachines.

[2]  Zhongqi Wang,et al.  In Situ Analysis of Chloride Effect on Platinum Dissolution by a Channel-Flow Multi-Electrode System , 2014 .

[3]  Gerald E Loeb,et al.  Accelerated life-test methods and results for implantable electronic devices with adhesive encapsulation , 2017, Biomedical microdevices.

[4]  K. Mayrhofer,et al.  Dissolution of Platinum in Presence of Chloride Traces , 2015 .

[5]  N. Donaldson,et al.  Apparatus to investigate the insulation impedance and accelerated life-testing of neural interfaces , 2018, Journal of neural engineering.

[6]  G J Suaning,et al.  Fabrication of implantable microelectrode arrays by laser cutting of silicone rubber and platinum foil , 2005, Journal of neural engineering.

[7]  Andreas Demosthenous,et al.  Flexible active electrode arrays with ASICs that fit inside the rat’s spinal canal , 2015, Biomedical microdevices.

[8]  A Vanhoestenberghe,et al.  Corrosion of silicon integrated circuits and lifetime predictions in implantable electronic devices , 2013, Journal of neural engineering.

[9]  Florian Solzbacher,et al.  Long-Term Bilayer Encapsulation Performance of Atomic Layer Deposited Al $_{\bf 2}$O$_{\bf 3}$ and Parylene C for Biomedical Implantable Devices , 2013, IEEE Transactions on Biomedical Engineering.

[10]  Wouter A Serdijn,et al.  Realizing flexible bioelectronic medicines for accessing the peripheral nerves – technology considerations , 2018, Bioelectronic Medicine.