A Near-Infrared Optomechanical Intracranial Pressure Microsensor

We present a wireless and power-free, optomechanical, implantable microsensor that can potentially be used to accurately monitor intracranial pressure (ICP) over long periods of time. The developed microsensor vertically integrates a glass mini-lens with a two-wavelength quantum dot (QD) micropillar that is photolithographically patterned on an ICP-exposed silicon nitride membrane. The operation principle is based on a novel optomechanical transduction scheme that converts ICP changes into changes in the intensity ratio of the two-wavelength, near-infrared fluorescent light emitted from the QDs. The microsensor is microfabricated using silicon bulk micromachining, and it operates at an ICP clinically relevant pressure dynamic range (0-40 mmHg). The microsensor has a maximum error of less than 15% throughout its dynamic range, and it is extremely photostable. We believe that the proposed microsensor will open up a new direction not only in ICP monitoring but in other pressure-related biomedical applications.

[1]  M. Poca,et al.  Fiberoptic intraparenchymal brain pressure monitoring with the Camino V420 monitor: reflections on our experience in 163 severely head-injured patients. , 2002, Journal of neurotrauma.

[2]  M. Poca,et al.  Is intracranial pressure monitoring in the epidural space reliable? Fact and fiction. , 2007, Journal of neurosurgery.

[3]  G. Clifton,et al.  Fluid thresholds and outcome from severe brain injury* , 2002, Critical care medicine.

[4]  C. Hierold,et al.  Implantable low power integrated pressure sensor system for minimal invasive telemetric patient monitoring , 1998, Proceedings MEMS 98. IEEE. Eleventh Annual International Workshop on Micro Electro Mechanical Systems. An Investigation of Micro Structures, Sensors, Actuators, Machines and Systems (Cat. No.98CH36176.

[5]  Pressure-controlled drainage of cerebrospinal fluid: clinical experience with a new type of ventricular catheter (Ventcontrol MTC)and an integrated Piezo-resistive sensor at its tip: technical note. , 1996, Neurosurgery.

[6]  M. Allen,et al.  Micromachined endovascularly-implantable wireless aneurysm pressure sensors: from concept to clinic , 2005, The 13th International Conference on Solid-State Sensors, Actuators and Microsystems, 2005. Digest of Technical Papers. TRANSDUCERS '05..

[7]  Rainer Laur,et al.  Advanced hybrid integrated low-power telemetric pressure monitoring system for biomedical applications , 2000, Proceedings IEEE Thirteenth Annual International Conference on Micro Electro Mechanical Systems (Cat. No.00CH36308).

[8]  T. Fukuda,et al.  Micro force sensor for intravascular neurosurgery and in vivo experiment , 1998, Proceedings MEMS 98. IEEE. Eleventh Annual International Workshop on Micro Electro Mechanical Systems. An Investigation of Micro Structures, Sensors, Actuators, Machines and Systems (Cat. No.98CH36176.

[9]  M A Poca,et al.  Interhemispheric supratentorial intracranial pressure gradients in head-injured patients: are they clinically important? , 1999, Journal of neurosurgery.

[10]  J. Maibach,et al.  A new analytical solution for the load-deflection of square membranes , 1995 .

[11]  W. Ko,et al.  Development of a miniature pressure transducer for biomedical applications , 1979, IEEE Transactions on Electron Devices.

[12]  John L. Hutchison,et al.  Uniform GaAs quantum dots in a polymer matrix , 1994 .

[13]  J. Pickard,et al.  Monitoring and interpretation of intracranial pressure , 2004, Journal of Neurology, Neurosurgery & Psychiatry.

[14]  R. Penn,et al.  Long-term testing of an intracranial pressure monitoring device. , 2000, Journal of neurosurgery.

[15]  J Meixensberger,et al.  Clinical experience with 118 brain tissue oxygen partial pressure catheter probes. , 1998, Neurosurgery.

[16]  M. Kohl,et al.  Near-infrared optical properties of ex vivo human skin and subcutaneous tissues measured using the Monte Carlo inversion technique. , 1998, Physics in medicine and biology.

[17]  M. Dujovny,et al.  Advances in ICP monitoring techniques , 2003, Neurological research.

[18]  I. Piper,et al.  Bench test assessment of the new Raumedic Neurovent-P ICP sensor: a technical report by the BrainIT group , 2004, Acta Neurochirurgica.

[19]  Kyihwan Park,et al.  Design optimization of a piezoresistive pressure sensor considering the output signal-to-noise ratio , 2004 .

[20]  A. Brawanski,et al.  Clinical evaluation of the InnerSpace fibreoptic intracranial pressure monitoring device. , 1998, Brain injury.

[21]  M. Gelabert-González,et al.  The Camino intracranial pressure device in clinical practice. Assessment in a 1000 cases , 2006, Acta Neurochirurgica.

[22]  A. N. Bashkatov,et al.  Optical properties of human skin, subcutaneous and mucous tissues in the wavelength range from 400 to 2000 nm , 2005 .

[23]  J. M. de Campos,et al.  Camino® intracranial pressure monitor: prospective study of accuracy and complications , 2000, Journal of neurology, neurosurgery, and psychiatry.

[24]  Kjell Hansson Mild,et al.  Tumour risk associated with use of cellular telephones or cordless desktop telephones , 2006, World journal of surgical oncology.

[25]  E. Cibula,et al.  Miniature fiber-optic pressure sensor with a polymer diaphragm. , 2005, Applied optics.

[26]  V. Bulović,et al.  Large‐Area Ordered Quantum‐Dot Monolayers via Phase Separation During Spin‐Casting , 2005 .

[27]  W. Gray,et al.  A clinical study of parenchymal and subdural miniature strain-gauge transducers for monitoring intracranial pressure. , 1996, Neurosurgery.

[28]  H. Coakham,et al.  Telemetric Implantable Pressure Sensor for Short- and Long-Term Monitoring of Intracranial Pressure , 2007, 2007 29th Annual International Conference of the IEEE Engineering in Medicine and Biology Society.

[29]  Guidelines for the management of severe head injury: are emergency physicians following them? , 2002, Academic emergency medicine : official journal of the Society for Academic Emergency Medicine.

[30]  Philippe Renaud,et al.  A Telemetric Pressure Sensor System for Biomedical Applications , 2008, IEEE Transactions on Biomedical Engineering.

[31]  E. Schmidt,et al.  Long-term implants of Parylene-C coated microelectrodes , 2006, Medical and Biological Engineering and Computing.

[32]  Vasilis Ntziachristos,et al.  Looking and listening to light: the evolution of whole-body photonic imaging , 2005, Nature Biotechnology.

[33]  A. Lacaita,et al.  Avalanche photodiodes and quenching circuits for single-photon detection. , 1996, Applied optics.

[34]  Martin Smith Monitoring Intracranial Pressure in Traumatic Brain Injury , 2008, Anesthesia and analgesia.

[35]  S. Nie,et al.  In vivo cancer targeting and imaging with semiconductor quantum dots , 2004, Nature Biotechnology.

[36]  T. Ducker,et al.  Effect of intracranial pressure monitoring and aggressive treatment on mortality in severe head injury. , 1982, Journal of neurosurgery.

[37]  L. Koskinen,et al.  Clinical Experience with the Intraparenchymal Intracranial Pressure Monitoring Codman MicroSensor System , 2005, Neurosurgery.

[38]  D. Abeysinghe,et al.  A novel MEMS pressure sensor fabricated on an optical fiber , 2001, IEEE Photonics Technology Letters.

[39]  Fabrizio Martelli,et al.  Light Propagation Through Biological Tissue and Other Diffusive Media: Theory, Solutions, and Software , 2009 .

[40]  M. Zornow,et al.  Continuous monitoring of intracranial pressure with a miniaturized fiberoptic device. , 1987, Journal of neurosurgery.

[41]  R. Jensen,et al.  Risk factors of intracranial pressure monitoring in children with fiberoptic devices: a critical review. , 1997, Surgical neurology.

[42]  Kwang-Su Lee,et al.  The measurement of the dielectric and optical properties of nano thin films by THz differential time-domain spectroscopy , 2003, Microelectron. J..

[43]  Thomas F Budinger,et al.  Biomonitoring with wireless communications. , 2003, Annual review of biomedical engineering.

[44]  R Ravi,et al.  Intracranial pressure monitoring , 2003 .

[45]  Marija Drndic,et al.  Efficient polymer-nanocrystal quantum-dot photodetectors , 2005 .

[46]  U. Kawoos,et al.  A permanently implantable intracranial pressure monitor , 2005, Proceedings of the IEEE 31st Annual Northeast Bioengineering Conference, 2005..

[47]  James G. Yusko Medical Device Safety: The Regulation of Medical Devices for Public Health and Safety , 2002 .

[48]  U. Kawoos,et al.  In-Vitro and In-Vivo Trans-Scalp Evaluation of an Intracranial Pressure Implant at 2.4 GHz , 2008, IEEE Transactions on Microwave Theory and Techniques.

[49]  M. Scheffler,et al.  Shape and stability of quantum dots , 1997 .

[50]  D S Yoo,et al.  Ventricular pressure monitoring during bilateral decompression with dural expansion. , 1999, Journal of neurosurgery.

[51]  Sung C. Choi,et al.  Impact of ICP instability and hypotension on outcome in patients with severe head trauma , 1991 .

[52]  Stella W. Pang,et al.  Three-dimensional SU-8 structures by reversal UV imprint , 2006 .

[53]  Nader Najafi,et al.  Initial Animal Studies of a Wireless, Batteryless, MEMS Implant for Cardiovascular Applications , 2004, Biomedical microdevices.

[54]  P. Choyke,et al.  Optimizing quantitative in vivo fluorescence imaging with near-infrared quantum dots. , 2011, Contrast media & molecular imaging.