Performance of noninvasive ventilation modes on ICU ventilators during pressure support: a bench model study

ObjectiveNoninvasive ventilation (NIV) is often applied with ICU ventilators. However, leaks at the patient-ventilator interface interfere with several key ventilator functions. Many ICU ventilators feature an NIV-specific mode dedicated to preventing these problems. The present bench model study aimed to evaluate the performance of these modes.Design and settingBench model study in an intensive care research laboratory of a university hospital.MethodsEight ICU ventilators, widely available in Europe and featuring an NIV mode, were connected by an NIV mask to a lung model featuring a plastic head to mimic NIV conditions, driven by an ICU ventilator imitating patient effort. Tests were conducted in the absence and presence of leaks, the latter condition with and without activation of the NIV mode. Trigger delay, trigger-associated inspiratory workload, and pressurization were tested in conditions of normal respiratory mechanics, and cycling was also assessed in obstructive and restrictive conditions.ResultsOn most ventilators leaks led to an increase in trigger delay and workload, a decrease in pressurization, and delayed cycling. On most ventilators the NIV mode partly or totally corrected these problems, but with large variations between machines. Furthermore, on some ventilators the NIV mode worsened the leak-induced dysfunction.ConclusionsThe results of this bench-model NIV study confirm that leaks interfere with several key functions of ICU ventilators. Overall, NIV modes can correct part or all of this interference, but with wide variations between machines in terms of efficiency. Clinicians should be aware of these differences when applying NIV with an ICU ventilator.

[1]  L. Brochard,et al.  Influence of respiratory rate on gas trapping during low volume ventilation of patients with acute lung injury , 2002, Intensive Care Medicine.

[2]  S. Mehta,et al.  Leak compensation in positive pressure ventilators: a lung model study. , 2001, The European respiratory journal.

[3]  A. Torres,et al.  Noninvasive ventilation in severe hypoxemic respiratory failure: a randomized clinical trial. , 2003, American journal of respiratory and critical care medicine.

[4]  M. Antonelli,et al.  Predictors of failure of noninvasive positive pressure ventilation in patients with acute hypoxemic respiratory failure: a multi-center study , 2001, Intensive Care Medicine.

[5]  L. Brochard,et al.  Effects of flow triggering on breathing effort during partial ventilatory support. , 1998, American journal of respiratory and critical care medicine.

[6]  N. Hill,et al.  Acute applications of noninvasive positive pressure ventilation. , 2003, Chest.

[7]  R. Natarajan,et al.  Air leaking through the mouth during nocturnal nasal ventilation: effect on sleep quality. , 1997, Sleep.

[8]  Marc Gainnier,et al.  Expiratory trigger setting in pressure support ventilation: From mathematical model to bedside , 2004, Critical care medicine.

[9]  D. Tassaux,et al.  Comparative bench study of triggering, pressurization, and cycling between the home ventilator VPAP II and three ICU ventilators , 2002, Intensive Care Medicine.

[10]  G. Heldt,et al.  Airway leak size in neonates and autocycling of three flow-triggered ventilators. , 1995, Critical care medicine.

[11]  M. Nishimura,et al.  Effects of gas leak on triggering function, humidification, and inspiratory oxygen fraction during noninvasive positive airway pressure ventilation. , 2005, Chest.

[12]  D. Georgopoulos Patient–Ventilator Interaction , 2005 .

[13]  S. Nava,et al.  Effect of varying the pressurisation rate during noninvasive pressure support ventilation , 2004, European Respiratory Journal.

[14]  S. Nava,et al.  Respiratory response and inspiratory effort during pressure support ventilation in COPD patients , 1995, Intensive Care Medicine.

[15]  A. Davies,et al.  Skin complications in the intensive care unit. , 1999, Clinics in chest medicine.

[16]  C. Carvalho,et al.  Mask mechanics and leak dynamics during noninvasive pressure support ventilation: a bench study , 2001, Intensive Care Medicine.

[17]  N. Ambrosino Noninvasive mechanical ventilation in acute respiratory failure. , 1996, The European respiratory journal.

[18]  M. Wysocki,et al.  Noninvasive versus conventional mechanical ventilation. An epidemiologic survey. , 2001, American journal of respiratory and critical care medicine.

[19]  G. Schmidt,et al.  Non-invasive Ventilation: Why Does It Fail? , 2000 .

[20]  J. Moran,et al.  Noninvasive ventilation in acute respiratory failure--a meta-analysis update. , 2002, Critical care medicine.

[21]  Marc Gainnier,et al.  Impact of expiratory trigger setting on delayed cycling and inspiratory muscle workload. , 2005, American journal of respiratory and critical care medicine.

[22]  C. Gregoretti,et al.  Patient-ventilator asynchrony during noninvasive ventilation: the role of expiratory trigger , 1999, Intensive Care Medicine.

[23]  S. Jaber,et al.  Performance characteristics of 10 home mechanical ventilators in pressure-support mode: a comparative bench study. , 2005, Chest.

[24]  L. Breton,et al.  Bench testing of pressure support ventilation with three different generations of ventilators , 2002, Intensive Care Medicine.

[25]  M. Tobin,et al.  Patient-ventilator interaction. , 2001, American journal of respiratory and critical care medicine.