Spectroscopic separation of Čerenkov radiation in high-resolution radiation fiber dosimeters

Abstract. We have investigated Čerenkov radiation generated in phosphor-based optical fiber dosimeters irradiated with clinical electron beams. We fabricated two high-spatial resolution fiber-optic probes, with 200 and 400  μm core diameters, composed of terbium-based phosphor tips. A generalizable spectroscopic method was used to separate Čerenkov radiation from the transmitted signal by the fiber based on the assumption that the recorded signal is a linear superposition of two basis spectra: characteristic luminescence of the phosphor medium and Čerenkov radiation. We performed Monte Carlo simulations of the Čerenkov radiation generated in the fiber and found a strong dependence of the recorded Čerenkov radiation on the numerical aperture of the fiber at shallow phantom depths; however, beyond the depth of maximum dose that dependency is minimal. The simulation results agree with the experimental results for Čerenkov radiation generated in fibers. The spectroscopic technique used in this work can be used for development of high-spatial resolution fiber micro dosimeters and for optical characterization of various scintillating materials, such as phosphor nanoparticles, in ionizing radiation fields of high energy.

[1]  Brian W. Pogue,et al.  Separation of Čerenkov radiation in irradiated optical fibers by optical spectroscopy , 2015, Photonics West - Biomedical Optics.

[2]  Erin Jackson,et al.  Cerenkov Radiation Energy Transfer (CRET) Imaging: A Novel Method for Optical Imaging of PET Isotopes in Biological Systems , 2010, PloS one.

[3]  Hanli Liu,et al.  SU-E-T-610: Phosphor-Based Fiber Optic Probes for Proton Beam Characterization , 2015 .

[4]  Brian W. Pogue,et al.  Čerenkov radiation emission and excited luminescence (CREL) sensitivity during external beam radiation therapy: Monte Carlo and tissue oxygenation phantom studies , 2012, Biomedical optics express.

[5]  N Suchowerska,et al.  Cerenkov-free scintillation dosimetry in external beam radiotherapy with an air core light guide , 2008, Physics in medicine and biology.

[6]  W. C. Kvaraceus,et al.  Principles and Practices , 2006 .

[7]  James A. Harrington,et al.  A Review of IR Transmitting, Hollow Waveguides , 2000 .

[8]  Scott C Davis,et al.  Projection imaging of photon beams using Čerenkov-excited fluorescence , 2013, Physics in medicine and biology.

[9]  Louis Archambault,et al.  A new water-equivalent 2D plastic scintillation detectors array for the dosimetry of megavoltage energy photon beams in radiation therapy. , 2011, Medical physics.

[10]  Scott C Davis,et al.  Čerenkov excited fluorescence tomography using external beam radiation. , 2013, Optics letters.

[11]  Wook Jae Yoo,et al.  Feasibility of fiber-optic radiation sensor using Cerenkov effect for detecting thermal neutrons. , 2013, Optics express.

[12]  L. Beaulieu,et al.  On the nature of the light produced within PMMA optical light guides in scintillation fiber-optic dosimetry , 2013, Physics in medicine and biology.

[13]  Brian W Pogue,et al.  Optical dosimetry of radiotherapy beams using Cherenkov radiation: the relationship between light emission and dose , 2014, Physics in medicine and biology.

[14]  Brian W Pogue,et al.  Superficial dosimetry imaging of Čerenkov emission in electron beam radiotherapy of phantoms , 2013, Physics in medicine and biology.

[15]  Arash Darafsheh,et al.  Characterization of rare-earth-doped nanophosphors for photodynamic therapy excited by clinical ionizing radiation beams , 2015, Photonics West - Biomedical Optics.

[16]  Timothy C. Zhu,et al.  Phosphor-based fiber optic microprobes for ionizing beam radiation dosimetry , 2015, Photonics West - Biomedical Optics.

[17]  E. Hull,et al.  Porphyrin Bleaching and PDT-induced Spectral Changes are Irradiance Dependent in ALA-sensitized Normal Rat Skin In Vivo¶ , 2001, Photochemistry and photobiology.

[18]  Louis Archambault,et al.  Spectral method for the correction of the Cerenkov light effect in plastic scintillation detectors: a comparison study of calibration procedures and validation in Cerenkov light-dominated situations. , 2011, Medical physics.

[19]  Louis Archambault,et al.  Dosimetric performance and array assessment of plastic scintillation detectors for stereotactic radiosurgery quality assurance. , 2011, Medical physics.

[20]  S. A. Pozzi,et al.  Measurement of neutron yields from UF4 , 2008, 2008 IEEE Nuclear Science Symposium Conference Record.

[21]  Scott C Davis,et al.  Three-dimensional Čerenkov tomography of energy deposition from ionizing radiation beams. , 2013, Optics letters.

[22]  J. M. Fontbonne,et al.  Scintillating fiber dosimeter for radiation therapy accelerator , 2001 .

[23]  L. Beaulieu,et al.  A phantom study of an in vivo dosimetry system using plastic scintillation detectors for real-time verification of 192Ir HDR brachytherapy. , 2011, Medical physics.

[24]  Francis Berghmans,et al.  Fibre-optic gamma-flux monitoring in a fission reactor by means of Cerenkov radiation , 2007 .

[25]  Brian W Pogue,et al.  Real-time in vivo Cherenkoscopy imaging during external beam radiation therapy , 2013, Journal of biomedical optics.

[26]  Jan Grimm,et al.  Quantitative imaging of disease signatures through radioactive decay signal conversion , 2013, Nature Medicine.

[27]  Vikram D Kodibagkar,et al.  On the potential for molecular imaging with Cerenkov luminescence. , 2010, Optics letters.

[28]  Thomas R. Mackie,et al.  Cerenkov light generated in optical fibres and other light pipes irradiated by electron beams , 1992 .

[29]  Wook Jae Yoo,et al.  Application of Cerenkov radiation generated in plastic optical fibers for therapeutic photon beam dosimetry , 2013, Journal of biomedical optics.

[30]  A Isambert,et al.  Spectral discrimination of Čerenkov radiation in scintillating dosimeters. , 2005, Medical physics.

[31]  Brian W Pogue,et al.  Cherenkov-excited luminescence scanned imaging. , 2015, Optics letters.

[32]  Louis Archambault,et al.  Validating plastic scintillation detectors for photon dosimetry in the radiologic energy range. , 2012, Medical physics.

[33]  Zhen Cheng,et al.  Harnessing the Power of Radionuclides for Optical Imaging: Cerenkov Luminescence Imaging , 2011, The Journal of Nuclear Medicine.

[34]  Arash Darafsheh,et al.  Optical super-resolution and periodical focusing effects by dielectric microspheres , 2013 .

[35]  Scott C Davis,et al.  Video-rate optical dosimetry and dynamic visualization of IMRT and VMAT treatment plans in water using Cherenkov radiation. , 2014, Medical physics.

[36]  Arion F. Chatziioannou,et al.  Cerenkov radiation imaging as a method for quantitative measurements of beta particles in a microfluidic chip , 2008, 2008 IEEE Nuclear Science Symposium Conference Record.

[37]  Brian W Pogue,et al.  Superficial dosimetry imaging based on Čerenkov emission for external beam radiotherapy with megavoltage x-ray beam. , 2013, Medical physics.

[38]  Brian W Pogue,et al.  Cherenkoscopy based patient positioning validation and movement tracking during post-lumpectomy whole breast radiation therapy , 2015, Physics in medicine and biology.

[39]  Giovanni Lucignani,et al.  Čerenkov radioactive optical imaging: a promising new strategy , 2011, European Journal of Nuclear Medicine and Molecular Imaging.

[40]  A Darafsheh,et al.  MO-G-BRF-07: Optical Characterization of Novel Terbium-Doped Nanophosphors Excited by Clinical Electron and Photon Beams for Potential Use in Molecular Imaging Or Photodynamic Therapy. , 2014, Medical physics.

[41]  S. Shin,et al.  Development of a fiber-optic dosimeter based on modified direct measurement for real-time dosimetry during radiation diagnosis , 2013 .

[42]  N Suchowerska,et al.  A plastic scintillation dosimeter for high dose rate brachytherapy , 2006, Physics in medicine and biology.

[43]  Scott C Davis,et al.  Oxygen tomography by Čerenkov-excited phosphorescence during external beam irradiation , 2013, Journal of biomedical optics.

[44]  Robert Speller,et al.  Imaging Cerenkov emission as a quality assurance tool in electron radiotherapy , 2014, Physics in medicine and biology.

[45]  A S Beddar,et al.  Optical filtering and spectral measurements of radiation-induced light in plastic scintillation dosimetry , 1993 .

[46]  P. A. Čerenkov Visible radiation produced by electrons moving in a medium with velocities exceeding that of light , 1937 .

[47]  Brian W Pogue,et al.  Cherenkov video imaging allows for the first visualization of radiation therapy in real time. , 2014, International journal of radiation oncology, biology, physics.

[48]  S. Brown,et al.  Cerenkov radiation and its applications , 1955 .

[49]  Jan Grimm,et al.  Cerenkov Luminescence Imaging of Medical Isotopes , 2010, Journal of Nuclear Medicine.

[50]  M Saiful Huq,et al.  Recommendations for clinical electron beam dosimetry: supplement to the recommendations of Task Group 25. , 2009, Medical physics.

[51]  P N Johnston,et al.  A temporal method of avoiding the Cerenkov radiation generated in organic scintillator dosimeters by pulsed mega-voltage electron and photon beams. , 2002, Physics in medicine and biology.

[52]  Wook Jae Yoo,et al.  Development of a Cerenkov radiation sensor to detect low-energy beta-particles. , 2013, Applied radiation and isotopes : including data, instrumentation and methods for use in agriculture, industry and medicine.

[53]  Jan Grimm,et al.  Cerenkov luminescence imaging , 2015, Imaging Modalities for Biological and Preclinical Research: A Compendium, Volume 2.

[54]  Yunlin Xu,et al.  Measurement of Neutron Yields From ${\rm UF}_{4}$ , 2010, IEEE Transactions on Nuclear Science.

[55]  Arash Darafsheh,et al.  Light Sources, Drugs, and Dosimetry , 2016 .

[56]  F. Khan The physics of radiation therapy , 1985 .

[57]  Brian W. Pogue,et al.  Projection imaging of photon beams by the Čerenkov effect. , 2012, Medical physics.

[58]  I. Tamm,et al.  Coherent visible radiation of fast electrons passing through matter , 1937 .