Post-NCS Performance of the HST NICMOS

We describe the on-orbit performance of the Near Infrared Camera and Multi-Object Spectrometer (NICMOS) aboard the Hubble Space Telescope (HST) following the installation of the NICMOS Cooling System (NCS). NICMOS is operated at a higher temperature (~77 K) than in the previous observing 1997-1998 period (~62 K). Due to the higher operating temperature, the detector QE is higher, while the well depth is less. The spatial structure of the flat field response remained essentially unchanged. We will show the effects of operating at the higher temperature and present current NICMOS calibration images. In addition, we present an overview of on-orbit testing and report on the re-enabling of NICMOS.