Fluoroscopic Radiation Exposure during Percutaneous Kyphoplasty.

OBJECTIVE The author measured levels of fluoroscopic radiation exposure to the surgeon's body based on the different beam directions during kyphoplasty. METHODS This is an observational study. A series of 84 patients (96 vertebral bodies) were treated with kyphoplasty over one year. The patients were divided into four groups based on the horizontal and vertical directions of the X-Ray beams. We measured radiation exposure with the seven dosimetry badges which were worn by the surgeon in each group (total of 28 badges). Twenty-four procedures were measured in each group. Cumulative dose and dose rates were compared between groups. RESULTS Fluoroscopic radiation is received by the operator in real-time for approximately 50% (half) of the operation time. Thyroid protectors and lead aprons can block radiation almost completely. The largest dose was received in the chest irrespective of beam directions. The lowest level of radiation were received when X-ray tube was away from the surgeon and beneath the bed (dose rate of head, neck, chest, abdomen and knee : 0.2986, 0.2828, 0.9711, 0.8977, 0.8168 mSv, respectively). The radiation differences between each group were approximately 2.7-10 folds. CONCLUSION When fluoroscopic guided-KP is performed, the X-Ray tube should be positioned on the opposite side of the operator and below the table, otherwise the received radiation to the surgeon's body would be 2.7-10 times higher than such condition.

[1]  W. Davros,et al.  Radiation Exposure to the Surgeon and the Patient During Kyphoplasty , 2008, Journal of spinal disorders & techniques.

[2]  Mahadevappa Mahesh,et al.  Fluoroscopy: patient radiation exposure issues. , 2001, Radiographics : a review publication of the Radiological Society of North America, Inc.

[3]  P M Rommens,et al.  Radiation exposure to the hands and the thyroid of the surgeon during intramedullary nailing. , 1998, Injury.

[4]  Y. Rampersaud,et al.  Radiation Exposure to the Spine Surgeon During Fluoroscopically Assisted Pedicle Screw Insertion , 2000, Spine.

[5]  J Anthony Seibert,et al.  Vertebroplasty and kyphoplasty: do fluoroscopy operators know about radiation dose, and should they want to know? , 2004, Radiology.

[6]  Masayuki Zuguchi,et al.  Relationship between fluoroscopic time, dose-area product, body weight, and maximum radiation skin dose in cardiac interventional procedures. , 2006, AJR. American journal of roentgenology.

[7]  David F Kallmes,et al.  Radiation dose to the operator during vertebroplasty: prospective comparison of the use of 1-cc syringes versus an injection device. , 2003, AJNR. American journal of neuroradiology.

[8]  S. Glaze,et al.  Surgeon and patient radiation exposure in minimally invasive transforaminal lumbar interbody fusion. , 2008, Journal of neurosurgery. Spine.

[9]  S. G. Wetzel,et al.  Radiation dose in vertebroplasty , 2004, Neuroradiology.

[10]  P. Quaretti,et al.  A study on maximum skin dose in cerebral embolization procedures. , 2007, AJNR. American journal of neuroradiology.

[11]  M. Synowitz,et al.  Surgeon's radiation exposure during percutaneous vertebroplasty. , 2006, Journal of neurosurgery. Spine.

[12]  Michael Oberst,et al.  Computer Navigation in Balloon Kyphoplasty Reduces the Intraoperative Radiation Exposure , 2009, Spine.

[13]  John Damilakis,et al.  Patient exposure and associated radiation risks from fluoroscopically guided vertebroplasty or kyphoplasty. , 2004, Radiology.

[14]  Choll W. Kim,et al.  Use of navigation-assisted fluoroscopy to decrease radiation exposure during minimally invasive spine surgery. , 2008, The spine journal : official journal of the North American Spine Society.

[15]  C. Cousins,et al.  Medical interventional procedures-reducing the radiation risks. , 2004, Clinical radiology.

[16]  Radiation Dose Reduction to Medical Staff During Vertebroplasty: A Review of Techniques and Methods to Mitigate Occupational Dose , 2003 .