Coded Spectral Doppler Imaging: From Simulation to Real-Time Processing

Transmission of coded pulses and matched receive filtering can improve the ultrasound imaging penetration depth while preserving the axial resolution. This paper shows that the pulse compression technique may be integrated in a low-cost scanner to be profitably used also in spectral Doppler investigations. By operating on beamformed, demodulated, and down-sampled data in the frequency domain, a single digital signal processor is proved sufficient to perform both pulse compression and multigate spectral Doppler algorithms in real time. Simulations, phantom, and in vivo experiments demonstrate that the transmission of (2.5 or 5 μs long) linear frequency-modulated chirps with bandwidths over the range 1.6-5.4 MHz, rather than of corresponding sine-burst pulses, provides signal-to-noise ratio (SNR) improvements very close to theory. Even in the presence of selective tissue attenuation, SNR gains up to 11 and 13.3 dB have been obtained for the short and the longer chirp, respectively. This may be important in clinical Doppler applications where the needed penetration depth is not achieved with sufficient SNR unless very long bursts are transmitted.

[1]  Dong C. Liu,et al.  Chirp-Coded Pulse Excitation for Ultrasound Elasticity Imaging , 2010, 2010 4th International Conference on Bioinformatics and Biomedical Engineering.

[2]  Piero Tortoli,et al.  High-frame-rate 2-D vector blood flow imaging in the frequency domain , 2014, IEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control.

[3]  Sevan Harput,et al.  Superharmonic imaging with chirp coded excitation: filtering spectrally overlapped harmonics , 2014, IEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control.

[4]  A. Ramalli,et al.  Real-time pulse compression in multigate spectral Doppler imaging , 2015, 2015 IEEE International Ultrasonics Symposium (IUS).

[5]  M. O'Donnell,et al.  Coded excitation system for improving the penetration of real-time phased-array imaging systems , 1992, IEEE Transactions on Ultrasonics, Ferroelectrics and Frequency Control.

[6]  J. Klauder,et al.  The theory and design of chirp radars , 1960 .

[7]  Tai-Kyong Song,et al.  Coded excitation for ultrasound tissue harmonic imaging. , 2010, Ultrasonics.

[8]  C. Atzeni,et al.  Spectral velocity profiles for detailed ultrasound flow analysis , 1996, IEEE Transactions on Ultrasonics, Ferroelectrics and Frequency Control.

[9]  B. Lamboul Analysis of the potential for coded excitation to improve the detection of tissue and blood motion in medical ultrasound , 2010 .

[10]  P Tortoli,et al.  A real-time chirp-coded imaging system with tissue attenuation compensation. , 2015, Ultrasonics.

[11]  Yangmo Yoo,et al.  An efficient pulse compression method of chirp-coded excitation in medical ultrasound imaging , 2013, IEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control.

[12]  J.A. Jensen,et al.  Use of modulated excitation signals in medical ultrasound. Part I: basic concepts and expected benefits , 2005, IEEE Transactions on Ultrasonics, Ferroelectrics and Frequency Control.

[13]  M. Lewandowski,et al.  High frequency coded imaging system with RF , 2008, IEEE Transactions on Ultrasonics, Ferroelectrics and Frequency Control.

[14]  Piero Tortoli,et al.  Interaction between secondary velocities, flow pulsation and vessel morphology in the common carotid artery. , 2003, Ultrasound in medicine & biology.

[15]  Jørgen A Jensen,et al.  Clinical evaluation of chirp-coded excitation in medical ultrasound. , 2003, Ultrasound in medicine & biology.

[16]  R. Finkelhor,et al.  Characteristics and impact of obesity on the outpatient echocardiography laboratory. , 2006, The American journal of cardiology.

[17]  P. Tortoli,et al.  Real-time base-band pulse compression imaging , 2013, 2013 IEEE International Ultrasonics Symposium (IUS).

[18]  K V Ramnarine,et al.  Validation of a new blood-mimicking fluid for use in Doppler flow test objects. , 1998, Ultrasound in medicine & biology.

[19]  Armando Manduca,et al.  Coded excitation plane wave imaging for shear wave motion detection , 2015, IEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control.

[20]  A. Bouakaz,et al.  Harmonic chirp imaging method for ultrasound contrast agent , 2005, IEEE Transactions on Ultrasonics, Ferroelectrics and Frequency Control.

[21]  P Tortoli,et al.  Effect of vessel curvature on Doppler derived velocity profiles and fluid flow. , 2005, Ultrasound in medicine & biology.

[22]  M. Oelze,et al.  An ultrasonic imaging speckle-suppression and contrast-enhancement technique by means of frequency compounding and coded excitation , 2009, IEEE Transactions on Ultrasonics, Ferroelectrics and Frequency Control.

[23]  R. Y. Chiao,et al.  Coded excitation for diagnostic ultrasound: a system developer's perspective , 2003, IEEE Transactions on Ultrasonics, Ferroelectrics and Frequency Control.

[24]  M. Oelze Bandwidth and resolution enhancement through pulse compression , 2007, IEEE Transactions on Ultrasonics, Ferroelectrics and Frequency Control.

[25]  A. Dallai,et al.  A reconfigurable and programmable FPGA-based system for nonstandard ultrasound methods , 2012, IEEE Transactions on Ultrasonics, Ferroelectrics and Frequency Control.

[26]  R. Y. Chiao,et al.  B-mode blood flow (B-flow) imaging , 2000, 2000 IEEE Ultrasonics Symposium. Proceedings. An International Symposium (Cat. No.00CH37121).

[27]  P. Banelli,et al.  Design and performance of Huffman sequences in medical ultrasound coded excitation , 2012, IEEE Transactions on Ultrasonics, Ferroelectrics and Frequency Control.

[28]  M. Insana,et al.  Modeling and phantom studies of ultrasonic wall shear rate measurements using coded pulse excitation , 2006, IEEE Transactions on Ultrasonics, Ferroelectrics and Frequency Control.

[29]  Vera Behar,et al.  Parameter optimization of pulse compression in ultrasound imaging systems with coded excitation. , 2004, Ultrasonics.

[30]  T. Misaridis,et al.  Use of modulated excitation signals in medical ultrasound. Part II: design and performance for medical imaging applications , 2005, IEEE Transactions on Ultrasonics, Ferroelectrics and Frequency Control.

[31]  Sevan Harput,et al.  Use of chirps in medical ultrasound imaging , 2012 .

[32]  Joanne Cowe,et al.  Improving performance of pulse compression in a Doppler ultrasound system using amplitude modulated chirps and Wiener filtering. , 2008, Ultrasound in medicine & biology.