The impact of rotor configurations on hemodynamic features, hemocompatibility and dynamic balance of the centrifugal blood pump: A numerical study

To investigate the effect of rotor design configuration on hemodynamic features, hemocompatibility and dynamic balance of blood pumps. Computational fluid dynamics was employed to investigate the effects of rotor type (closed impeller, semi‐open impeller), clearance height and back vanes on blood pump performance. In particular, the Eulerian hemolysis model based on a power‐law function and the Lagrangian thrombus model with integrated stress accumulation and residence time were applied to evaluate the hemocompatibility of the blood pump. This study shows that compared to the closed impeller, the semi‐open impeller can improve hemolysis at a slight sacrifice in head pressure, but increase the risk of thrombogenic potential and disrupt rotor dynamic balance. For the semi‐open impeller, the pressure head, hemolysis, and axial thrust of the blood pump decrease with increasing front clearance, and the risk of thrombosis increases first and then decreases with increasing front clearance. Variations in back clearance have little effect on pressure head, but larger on back clearance, worsens hemolysis, thrombogenic potential and rotor dynamic balance. The employment of back vanes has little effect on the pressure head. All back vanes configurations have an increased risk of hemolysis in the blood pump but are beneficial for the improvement of the rotor dynamic balance of the blood pump. Reasonable back vanes configuration (higher height, wider width, longer length and more number) decreases the flow separation, increases the velocity of blood in the back clearance, and reduces the risk of blood pooling and thrombosis. It was also found that hemolysis index (HI) was highly negatively correlated with pressure difference between the top and back clearances (r = −.87), and thrombogenic potential was positively correlated with pressure difference between the top and back clearances (r = .71). This study found that rotor type, clearance height, and back vanes significantly affect the hydraulic performance, hemocompatibility and rotor dynamic balance of centrifugal blood pumps through secondary flow. These parameters should be carefully selected when designing and optimizing centrifugal blood pumps for improving the blood pump clinical outcomes.

[1]  Xiaoyan Deng,et al.  Impact of volute design features on hemodynamic performance and hemocompatibility of centrifugal blood pumps used in ECMO. , 2022, Artificial organs.

[2]  Xiaoyan Deng,et al.  A New Mathematical Numerical Model to Evaluate the Risk of Thrombosis in Three Clinical Ventricular Assist Devices , 2022, Bioengineering.

[3]  Amparo Gil,et al.  CFD analysis of the HVAD’s hemodynamic performance and blood damage with insight into gap clearance , 2022, Biomechanics and Modeling in Mechanobiology.

[4]  Xiaoyan Deng,et al.  Investigation of the influence of blade configuration on the hemodynamic performance and blood damage of the centrifugal blood pump. , 2022, Artificial organs.

[5]  Peng Wu,et al.  The Design and Evaluation of a Portable Extracorporeal Centrifugal Blood Pump , 2021, Frontiers in Physiology.

[6]  Wei‐Tao Wu,et al.  On the Optimization of a Centrifugal Maglev Blood Pump Through Design Variations , 2021, Frontiers in Physiology.

[7]  D. Papavassiliou,et al.  Hemolysis estimation in turbulent flow for the FDA critical path initiative centrifugal blood pump , 2021, Biomechanics and Modeling in Mechanobiology.

[8]  Barry N. Gellman,et al.  Design and Computational Evaluation of a Pediatric MagLev Rotary Blood Pump , 2020, ASAIO journal.

[9]  Guang-Mao Liu,et al.  A two-phase flow approach for modeling blood stasis and estimating the thrombosis potential of a ventricular assist device , 2020, The International journal of artificial organs.

[10]  A. Redaelli,et al.  Thrombotic Risk of Rotor Speed Modulation Regimes of Contemporary Centrifugal Continuous-flow Left Ventricular Assist Devices , 2020, ASAIO journal.

[11]  Jianjun Du,et al.  Impeller (straight blade) design variations and their influence on the performance of a centrifugal blood pump , 2020, The International journal of artificial organs.

[12]  Jiafeng Zhang,et al.  Computational characterization of flow and blood damage potential of the new maglev CH-VAD pump versus the HVAD and HeartMate II pumps , 2020, The International journal of artificial organs.

[13]  Jiafeng Zhang,et al.  Impact of High Mechanical Shear Stress and Oxygenator Membrane Surface on Blood Damage Relevant to Thrombosis and Bleeding in a Pediatric ECMO Circuit. , 2020, Artificial organs.

[14]  T. Maruyama,et al.  Detection of thrombosis in a magnetically levitated blood pump by vibrational excitation of the impeller. , 2020, Artificial organs.

[15]  F. Fernández‐Avilés,et al.  Blood Stasis Imaging Predicts Cerebral Microembolism during Acute Myocardial Infarction. , 2019, Journal of the American Society of Echocardiography : official publication of the American Society of Echocardiography.

[16]  Jiafeng Zhang,et al.  The impact of shear stress on device-induced platelet hemostatic dysfunction relevant to thrombosis and bleeding in mechanically assisted circulation. , 2019, Artificial organs.

[17]  J. Kirkpatrick,et al.  Left Ventricular Assist Device Inflow Cannula Insertion Depth Influences Thrombosis Risk , 2019, ASAIO journal.

[18]  R. Cheng,et al.  Small Left Ventricular Size Is an Independent Risk Factor for Ventricular Assist Device Thrombosis , 2019, ASAIO journal.

[19]  Po-Lin Hsu,et al.  On the representation of effective stress for computing hemolysis , 2019, Biomechanics and modeling in mechanobiology.

[20]  Amir Nejat,et al.  Shape optimization of a centrifugal blood pump by coupling CFD with metamodel-assisted genetic algorithm , 2018, Journal of Artificial Organs.

[21]  Yu Chang,et al.  Computational Fluid Dynamics Analysis of Shroud Design on Hemodynamic Performance and Blood Damage in a Centrifugal Blood Pump , 2018, Computer Modeling in Engineering & Sciences.

[22]  I. Lazoglu,et al.  Effect of blade curvature on the hemolytic and hydraulic characteristics of a centrifugal blood pump , 2018, The International journal of artificial organs.

[23]  M. Meboldt,et al.  Blood Pump Design Variations and Their Influence on Hydraulic Performance and Indicators of Hemocompatibility , 2018, Annals of Biomedical Engineering.

[24]  Bartley P Griffith,et al.  Flow features and device‐induced blood trauma in CF‐VADs under a pulsatile blood flow condition: A CFD comparative study , 2018, International journal for numerical methods in biomedical engineering.

[25]  R. Kaufmann,et al.  Investigation of the Axial Gap Clearance in a Hydrodynamic‐Passive Magnetically Levitated Rotary Blood Pump Using X‐Ray Radiography , 2018, Artificial organs.

[26]  Alireza Riasi,et al.  Numerical Investigation of Back Vane Design and Its Impact on Pump Performance , 2017 .

[27]  A. Redaelli,et al.  High Frequency Components of Hemodynamic Shear Stress Profiles are a Major Determinant of Shear-Mediated Platelet Activation in Therapeutic Blood Recirculating Devices , 2017, Scientific Reports.

[28]  Ryo Kosaka,et al.  Properties of a monopivot centrifugal blood pump manufactured by 3D printing , 2016, Journal of Artificial Organs.

[29]  Zhongjun J. Wu,et al.  Paradoxical Effect of Nonphysiological Shear Stress on Platelets and von Willebrand Factor. , 2016, Artificial organs.

[30]  X. Xu,et al.  Mathematical modeling of thrombus formation in idealized models of aortic dissection: initial findings and potential applications , 2016, Journal of mathematical biology.

[31]  Danny Bluestein,et al.  Lagrangian methods for blood damage estimation in cardiovascular devices - How numerical implementation affects the results , 2016, Expert review of medical devices.

[32]  Klaus Affeld,et al.  Numerical Analysis of Blood Damage Potential of the HeartMate II and HeartWare HVAD Rotary Blood Pumps. , 2015, Artificial organs.

[33]  Jiafeng Zhang,et al.  Quantification of Shear-Induced Platelet Activation: High Shear Stresses for Short Exposure Time. , 2015, Artificial organs.

[34]  Ina Laura Pieper,et al.  The CentriMag Centrifugal Blood Pump as a Benchmark for In Vitro Testing of Hemocompatibility in Implantable Ventricular Assist Devices , 2014, Artificial organs.

[35]  Steven W Boyce,et al.  Current status of left ventricular assist device technology. , 2013, Seminars in thoracic and cardiovascular surgery.

[36]  M. Slaughter,et al.  Blood trauma testing of CentriMag and RotaFlow centrifugal flow devices: a pilot study. , 2012, Artificial organs.

[37]  Tao Zhang,et al.  A quantitative comparison of mechanical blood damage parameters in rotary ventricular assist devices: shear stress, exposure time and hemolysis index. , 2012, Journal of biomechanical engineering.

[38]  Tao Zhang,et al.  Computational model-based design of a wearable artificial pump-lung for cardiopulmonary/respiratory support. , 2012, Artificial organs.

[39]  V. L. Rayz,et al.  Flow Residence Time and Regions of Intraluminal Thrombus Deposition in Intracranial Aneurysms , 2010, Annals of Biomedical Engineering.

[40]  James F Antaki,et al.  Computational fluid dynamics analysis of blade tip clearances on hemodynamic performance and blood damage in a centrifugal ventricular assist device. , 2009, Artificial organs.

[41]  Yi Qian,et al.  Estimation of changes in dynamic hydraulic force in a magnetically suspended centrifugal blood pump with transient computational fluid dynamics analysis , 2009, Journal of Artificial Organs.

[42]  Alexandrina Untaroiu,et al.  Fluid force predictions and experimental measurements for a magnetically levitated pediatric ventricular assist device. , 2007, Artificial organs.

[43]  Bartley P Griffith,et al.  Computational and experimental evaluation of the fluid dynamics and hemocompatibility of the CentriMag blood pump. , 2006, Artificial organs.

[44]  William R Wagner,et al.  Elimination of adverse leakage flow in a miniature pediatric centrifugal blood pump by computational fluid dynamics-based design optimization. , 2005, ASAIO journal.

[45]  Marcus Hormes,et al.  Comparison of hydraulic and hemolytic properties of different impeller designs of an implantable rotary blood pump by computational fluid dynamics. , 2004, Artificial organs.

[46]  Mitsuo Umezu,et al.  Effects of Turbulent Stresses upon Mechanical Hemolysis: Experimental and Computational Analysis , 2004, ASAIO journal.

[47]  Xinwei Song,et al.  A prototype HeartQuest ventricular assist device for particle image velocimetry measurements. , 2002, Artificial organs.

[48]  Y Nosé,et al.  Impeller design for a miniaturized centrifugal blood pump. , 2000, Artificial organs.

[49]  T. Masuzawa,et al.  Magnetically suspended rotary blood pump with radial type combined motor-bearing. , 2000, Artificial organs.

[50]  K S Sakariassen,et al.  Shear-induced platelet activation and platelet microparticle formation at blood flow conditions as in arteries with a severe stenosis. , 1997, Arteriosclerosis, thrombosis, and vascular biology.

[51]  O. Bodger,et al.  In Vitro Benchmarking Study of Ventricular Assist Devices in Current Clinical Use. , 2019, Journal of cardiac failure.

[52]  M. Nishida,et al.  Development of a real-time and quantitative thrombus sensor for an extracorporeal centrifugal blood pump by near-infrared light. , 2018, Biomedical optics express.

[53]  A. Aliseda,et al.  LVAD Outflow Graft Angle and Thrombosis Risk , 2017, ASAIO journal.

[54]  S. Woo Contribution of biomechanics to clinical practice in orthopaedics , 2004, The 26th Annual International Conference of the IEEE Engineering in Medicine and Biology Society.