Aortic Annular Sizing Using a Novel 3-Dimensional Echocardiographic Method: Use and Comparison With Cardiac Computed Tomography

Background—Previous studies have shown cross-sectional 3-dimensional (3D) transesophageal echocardiographic (TEE) measurements to severely underestimate multidetector row computed tomographic (MDCT) measurements for the assessment of aortic annulus before transcatheter aortic valve replacement. This study compares annulus measurements from 3D-TEE using off-label use of commercially available software with MDCT measurements and assesses their ability to predict paravalvular regurgitation. Methods and Results—One hundred patients with severe, symptomatic aortic stenosis who had both contrast MDCT and 3D-TEE for annulus assessment before balloon-expandable transcatheter aortic valve replacement were analyzed. Annulus area, perimeter, and orthogonal maximum and minimum diameters were measured. Receiver operating characteristic analysis was performed with mild or greater paravalvular regurgitation as the classification variable. Three-dimensional TEE and MDCT cross-sectional perimeter and area measurements were strongly correlated (r=0.93–0.94; P<0.0001); however, the small differences (⩽1%) were statistically significant (P=0.0002 and 0.0074, respectively). Discriminatory ability for ≥ mild paravalvular regurgitation was good for both MDCT (area under the curve for perimeter and area cover index=0.715 and 0.709, respectively) and 3D-TEE (area under the curve for perimeter and area cover index=0.709 and 0.694, respectively). Differences in receiver operating characteristic analysis between MDCT and 3D-TEE perimeter and area cover indexes were not statistically significant (P=0.15 and 0.35, respectively). Conclusions—Annulus measurements using a new method for analyzing 3D-TEE images closely approximate those of MDCT. Annulus measurements from both modalities predict mild or greater paravalvular regurgitation with equivalent accuracy.

[1]  A. Kirtane,et al.  Predicting paravalvular regurgitation following transcatheter valve replacement: utility of a novel method for three-dimensional echocardiographic measurements of the aortic annulus. , 2013, Journal of the American Society of Echocardiography : official publication of the American Society of Echocardiography.

[2]  S. Pocock,et al.  Transcatheter aortic-valve implantation for aortic stenosis in patients who cannot undergo surgery. , 2010, The New England journal of medicine.

[3]  Jeroen J. Bax,et al.  Noninvasive evaluation of the aortic root with multislice computed tomography implications for transcatheter aortic valve replacement. , 2008, JACC. Cardiovascular imaging.

[4]  Jeroen J. Bax,et al.  Comparison of Aortic Root Dimensions and Geometries Before and After Transcatheter Aortic Valve Implantation by 2- and 3-Dimensional Transesophageal Echocardiography and Multislice Computed Tomography , 2010, Circulation. Cardiovascular imaging.

[5]  Nikolaus Marx,et al.  Association of aortic valve calcification severity with the degree of aortic regurgitation after transcatheter aortic valve implantation. , 2011, International journal of cardiology.

[6]  C. da Silva,et al.  Transesophageal echocardiography measurements of aortic annulus diameter using biplane mode in patients undergoing transcatheter aortic valve implantation , 2013, Cardiovascular Ultrasound.

[7]  Antonio Colombo,et al.  Clinical ResearchInterventional CardiologyIncidence, Predictors, and Outcomes of Aortic Regurgitation After Transcatheter Aortic Valve Replacement: Meta-Analysis and Systematic Review of Literature , 2013 .

[8]  Friedhelm Beyersdorf,et al.  Preoperative assessment of aortic annulus dimensions: comparison of noninvasive and intraoperative measurement. , 2011, The Annals of thoracic surgery.

[9]  D. Andreini,et al.  Aortic annulus area assessment by multidetector computed tomography for predicting paravalvular regurgitation in patients undergoing balloon-expandable transcatheter aortic valve implantation: a comparison with transthoracic and transesophageal echocardiography. , 2012, American heart journal.

[10]  Susheel Kodali,et al.  Paravalvular leak after transcatheter aortic valve replacement: the new Achilles' heel? A comprehensive review of the literature. , 2013, Journal of the American College of Cardiology.

[11]  Michael N D'Ambra,et al.  Safety of transesophageal echocardiography. , 2010, Journal of the American Society of Echocardiography : official publication of the American Society of Echocardiography.

[12]  P. Serruys,et al.  Correlates on MSCT of paravalvular aortic regurgitation after transcatheter aortic valve implantation using the medtronic corevalve prosthesis , 2011, Catheterization and cardiovascular interventions : official journal of the Society for Cardiac Angiography & Interventions.

[13]  W. Keeling,et al.  Transcatheter Aortic-Valve Implantation for Aortic Stenosis in Patients Who Cannot Undergo Surgery , 2011 .

[14]  Robert Bauernschmitt,et al.  Aortic annulus evaluation in transcatheter aortic valve implantation , 2010, Catheterization and cardiovascular interventions : official journal of the Society for Cardiac Angiography & Interventions.

[15]  Fabien Hyafil,et al.  Multimodal assessment of the aortic annulus diameter: implications for transcatheter aortic valve implantation. , 2010, Journal of the American College of Cardiology.

[16]  Amit R. Patel,et al.  Accuracy of aortic annular measurements obtained from three-dimensional echocardiography, CT and MRI: human in vitro and in vivo studies , 2012, Heart.

[17]  O. Alfieri,et al.  Periprocedural and short-term outcomes of transfemoral transcatheter aortic valve implantation with the Sapien XT as compared with the Edwards Sapien valve. , 2011, JACC. Cardiovascular interventions.

[18]  Lukas Altwegg,et al.  Prevention and management of transcatheter balloon‐expandable aortic valve malposition , 2008, Catheterization and cardiovascular interventions : official journal of the Society for Cardiac Angiography & Interventions.

[19]  H. Barnhart,et al.  An Overview on Assessing Agreement with Continuous Measurements , 2007, Journal of biopharmaceutical statistics.

[20]  Antonio Colombo,et al.  Computed tomography-based evaluation of aortic annulus, prosthesis size and impact on early residual aortic regurgitation after transcatheter aortic valve implantation. , 2013, European journal of cardio-thoracic surgery : official journal of the European Association for Cardio-thoracic Surgery.

[21]  S. Shaughnessy,et al.  Do No Harm: Health Systems’ Duty to Promote Clinician Well-Being , 2022, American Journal of Hospital Medicine.

[22]  C. Hengstenberg,et al.  Prosthesis sizing for transcatheter aortic valve implantation--comparison of three dimensional transesophageal echocardiography with multislice computed tomography. , 2013, International journal of cardiology.

[23]  Robert H. Anderson,et al.  Anatomy of the Aortic Valvar Complex and Its Implications for Transcatheter Implantation of the Aortic Valve , 2008, Circulation. Cardiovascular interventions.

[24]  Michael Becker,et al.  Comparison of two-dimensional and three-dimensional imaging techniques for measurement of aortic annulus diameters before transcatheter aortic valve implantation , 2011, Heart.

[25]  S. Achenbach,et al.  CT predictors of post-procedural aortic regurgitation in patients referred for transcatheter aortic valve implantation: an analysis of 105 patients , 2013, The International Journal of Cardiovascular Imaging.

[26]  A. Mahnken,et al.  Evaluation of aortic root for definition of prosthesis size by magnetic resonance imaging and cardiac computed tomography: implications for transcatheter aortic valve implantation , 2011, International journal of cardiology.

[27]  Adnan Kastrati,et al.  Standardized imaging for aortic annular sizing: implications for transcatheter valve selection. , 2013, JACC. Cardiovascular imaging.

[28]  J. Zamorano,et al.  Three-dimensional echocardiography in paravalvular aortic regurgitation assessment after transcatheter aortic valve implantation. , 2012, Journal of the American Society of Echocardiography : official publication of the American Society of Echocardiography.

[29]  Jeroen J. Bax,et al.  Accuracy of three-dimensional versus two-dimensional echocardiography for quantification of aortic regurgitation and validation by three-dimensional three-directional velocity-encoded magnetic resonance imaging. , 2013, The American journal of cardiology.

[30]  P. Serruys,et al.  Assessment of the aortic annulus by multislice computed tomography, contrast aortography, and trans‐thoracic echocardiography in patients referred for transcatheter aortic valve implantation , 2011, Catheterization and cardiovascular interventions : official journal of the Society for Cardiac Angiography & Interventions.

[31]  Roberto M. Lang,et al.  Guidelines for Performing a Comprehensive Transesophageal Echocardiographic Examination: Recommendations from the American Society of Echocardiography and the Society of Cardiovascular Anesthesiologists , 2014, Anesthesia and analgesia.

[32]  Jeroen J. Bax,et al.  Location and severity of aortic valve calcium and implications for aortic regurgitation after transcatheter aortic valve implantation. , 2011, The American journal of cardiology.

[33]  Stuart J Pocock,et al.  Transcatheter versus surgical aortic-valve replacement in high-risk patients. , 2011, The New England journal of medicine.

[34]  S. Achenbach,et al.  Multi-detector computed tomography is equivalent to trans-oesophageal echocardiography for the assessment of the aortic annulus before transcatheter aortic valve implantation , 2012, European Radiology.

[35]  Wen Cheng,et al.  Aortic annular sizing for transcatheter aortic valve replacement using cross-sectional 3-dimensional transesophageal echocardiography. , 2013, Journal of the American College of Cardiology.

[36]  R. Erbel,et al.  Measurement of the aortic annulus size by real-time three-dimensional transesophageal echocardiography , 2011, Minimally invasive therapy & allied technologies : MITAT : official journal of the Society for Minimally Invasive Therapy.

[37]  J. Zamorano,et al.  Prosthesis/annulus discongruence assessed by three-dimensional transoesophageal echocardiography: a predictor of significant paravalvular aortic regurgitation after transcatheter aortic valve implantation. , 2012, European heart journal cardiovascular Imaging.

[38]  Jeroen J. Bax,et al.  Intraoperative 2D and 3D transoesophageal echocardiographic predictors of aortic regurgitation after transcatheter aortic valve implantation , 2012, Heart.

[39]  E. DeLong,et al.  Comparing the areas under two or more correlated receiver operating characteristic curves: a nonparametric approach. , 1988, Biometrics.

[40]  J. Leipsic,et al.  A practical guide to multimodality imaging of transcatheter aortic valve replacement. , 2012, JACC. Cardiovascular imaging.

[41]  J. Fleiss,et al.  Intraclass correlations: uses in assessing rater reliability. , 1979, Psychological bulletin.

[42]  N. Nanda,et al.  Assessment of Aortic Regurgitation by Live Three‐Dimensional Transthoracic Echocardiographic Measurements of Vena Contracta Area: Usefulness and Validation , 2005, Echocardiography.

[43]  Susheel Kodali,et al.  Transcatheter aortic-valve replacement for inoperable severe aortic stenosis. , 2012, The New England journal of medicine.

[44]  F. Mohr,et al.  Transesophageal echocardiographic scoring for transcatheter aortic valve implantation: impact of aortic cusp calcification on postoperative aortic regurgitation. , 2011, The Journal of thoracic and cardiovascular surgery.

[45]  J. Zamorano,et al.  3D color-Doppler echocardiography and chronic aortic regurgitation: a novel approach for severity assessment. , 2013, International journal of cardiology.

[46]  Ronen Gurvitch,et al.  Aortic annulus diameter determination by multidetector computed tomography: reproducibility, applicability, and implications for transcatheter aortic valve implantation. , 2011, JACC. Cardiovascular interventions.

[47]  A. Cribier,et al.  Prospective analysis of 30-day safety and performance of transfemoral transcatheter aortic valve implantation with Edwards SAPIEN XT versus SAPIEN prostheses. , 2012, Archives of cardiovascular diseases.

[48]  S. Achenbach,et al.  3-dimensional aortic annular assessment by multidetector computed tomography predicts moderate or severe paravalvular regurgitation after transcatheter aortic valve replacement: a multicenter retrospective analysis. , 2012, Journal of the American College of Cardiology.

[49]  M. Leon,et al.  Cross-sectional computed tomographic assessment improves accuracy of aortic annular sizing for transcatheter aortic valve replacement and reduces the incidence of paravalvular aortic regurgitation. , 2012, Journal of the American College of Cardiology.

[50]  C. Hengstenberg,et al.  Impact of three‐dimensional transesophageal echocardiography on prosthesis sizing for transcatheter aortic valve implantation , 2012, Catheterization and cardiovascular interventions : official journal of the Society for Cardiac Angiography & Interventions.