Authors' reply [to comments on "Theoretical model for an MRI radio frequency resonator"]

The authors thank the commentors for their remarks (see De Zanche et al., ibid., vol.49, p.494, 2002) and for the opportunity to clarify and correct some statements in the subject paper (see ibid., vol.47, p.535, 2000). The principal issue raised by De Zanche deals with a remark that appears the in original paper which seems to suggest that maxima of |E| and maxima of |H| are coincident in a transverse electromagnetic (TEM) resonator. The analysis presented in the original makes it clear that this cannot occur. The periodic nature of the TEM resonator fields is relevant to high-field coil design.

[1]  J. Wikswo,et al.  An Estimate of the Steady Magnetic Field Strength Required to Influence Nerve Conduction , 1980, IEEE Transactions on Biomedical Engineering.

[2]  P. Basser Interstitial pressure, volume, and flow during infusion into brain tissue. , 1992, Microvascular research.

[3]  I. Tasaki,et al.  Volume expansion of nonmyelinated nerve fibers during impulse conduction. , 1990, Biophysical journal.

[4]  Y. Fung Foundations of solid mechanics , 1965 .

[5]  D. B. Heppner,et al.  Considerations of quasi-stationarity in electrophysiological systems. , 1967, The Bulletin of mathematical biophysics.

[6]  J P Wikswo,et al.  A calculation of the magnetic field of a nerve action potential. , 1980, Biophysical journal.

[7]  A. Love A treatise on the mathematical theory of elasticity , 1892 .

[8]  A. Lauder,et al.  Magnetic susceptibility imaging for nondestructive evaluation (using SQUID magnetometer) , 1993, IEEE Transactions on Applied Superconductivity.

[9]  J P Wikswo,et al.  Noninvasive magnetic detection of cardiac mechanical activity: theory. , 1980, Medical physics.

[10]  J P Wikswo,et al.  Magnetic field of a nerve impulse: first measurements. , 1980, Science.

[11]  P. Basser,et al.  Estimation of the effective self-diffusion tensor from the NMR spin echo. , 1994, Journal of magnetic resonance. Series B.

[12]  S. H. Lam,et al.  Principles of Fluid Mechanics , 1964 .

[13]  M.R. Islam,et al.  A magneto-acoustic method for the noninvasive measurement of bioelectric currents , 1988, IEEE Transactions on Biomedical Engineering.

[14]  John P. Wikswo,et al.  Magnetic susceptibility tomography for three-dimensional imaging of diamagnetic and paramagnetic objects , 1994 .

[15]  B C Towe,et al.  Bioelectric current image reconstruction from magneto-acoustic measurements. , 1988, IEEE transactions on medical imaging.

[16]  R. Turner,et al.  Echo‐planar time course MRI of cat brain oxygenation changes , 1991, Magnetic resonance in medicine.