Verification of respiratory-gated radiotherapy with new real-time tumour-tracking radiotherapy system using cine EPID images and a log file

A combined system comprising the TrueBeam linear accelerator and a new real-time tumour-tracking radiotherapy system, SyncTraX, was installed at our institution. The objectives of this study are to develop a method for the verification of respiratory-gated radiotherapy with SyncTraX using cine electronic portal image device (EPID) images and a log file and to verify this treatment in clinical cases. Respiratory-gated radiotherapy was performed using TrueBeam and the SyncTraX system. Cine EPID images and a log file were acquired for a phantom and three patients during the course of the treatment. Digitally reconstructed radiographs (DRRs) were created for each treatment beam using a planning CT set. The cine EPID images, log file, and DRRs were analysed using a developed software. For the phantom case, the accuracy of the proposed method was evaluated to verify the respiratory-gated radiotherapy. For the clinical cases, the intra- and inter-fractional variations of the fiducial marker used as an internal surrogate were calculated to evaluate the gating accuracy and set-up uncertainty in the superior-inferior (SI), anterior-posterior (AP), and left-right (LR) directions. The proposed method achieved high accuracy for the phantom verification. For the clinical cases, the intra- and inter-fractional variations of the fiducial marker were  ⩽3 mm and  ±3 mm in the SI, AP, and LR directions. We proposed a method for the verification of respiratory-gated radiotherapy with SyncTraX using cine EPID images and a log file and showed that this treatment is performed with high accuracy in clinical cases.

[1]  George Starkschall,et al.  Respiratory gating with EPID-based verification: the MDACC experience , 2009, Physics in medicine and biology.

[2]  Floris Ernst,et al.  A comparison of two clinical correlation models used for real-time tumor tracking of semi-periodic motion: A focus on geometrical accuracy in lung and liver cancer patients. , 2015, Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology.

[3]  Jiajia Ge,et al.  Planning 4-dimensional computed tomography (4DCT) cannot adequately represent daily intrafractional motion of abdominal tumors. , 2012, International journal of radiation oncology, biology, physics.

[4]  E. Larsen,et al.  A method for incorporating organ motion due to breathing into 3D dose calculations. , 1999, Medical physics.

[5]  Steve B Jiang,et al.  A technique for respiratory-gated radiotherapy treatment verification with an EPID in cine mode , 2005 .

[6]  K. Langen,et al.  Organ motion and its management. , 2001, International journal of radiation oncology, biology, physics.

[7]  R. Emery,et al.  Clinical experience using respiratory gated radiation therapy: comparison of free-breathing and breath-hold techniques. , 2004, International journal of radiation oncology, biology, physics.

[8]  Steve B. Jiang,et al.  Residual motion of lung tumours in gated radiotherapy with external respiratory surrogates , 2005, Physics in medicine and biology.

[9]  Mitsuhiro Nakamura,et al.  Positional reproducibility of pancreatic tumors under end-exhalation breath-hold conditions using a visual feedback technique. , 2011, International journal of radiation oncology, biology, physics.

[10]  Margrit Betke,et al.  The correlation between internal and external markers for abdominal tumors: implications for respiratory gating. , 2003, International journal of radiation oncology, biology, physics.

[11]  M. V. van Herk,et al.  Physical aspects of a real-time tumor-tracking system for gated radiotherapy. , 2000, International journal of radiation oncology, biology, physics.

[12]  C. Ramsey,et al.  A comparison of beam characteristics for gated and nongated clinical x-ray beams. , 1999, Medical physics.

[13]  Hiroki Shirato,et al.  Investigation of the change in marker geometry during respiration motion: a preliminary study for dynamic-multi-leaf real-time tumor tracking , 2012, Radiation oncology.

[14]  Paul J Keall,et al.  Retrospective analysis of artifacts in four-dimensional CT images of 50 abdominal and thoracic radiotherapy patients. , 2008, International journal of radiation oncology, biology, physics.

[15]  C. Ling,et al.  Evaluation of respiratory movement during gated radiotherapy using film and electronic portal imaging. , 2002, International journal of radiation oncology, biology, physics.

[16]  Geoffrey D Hugo,et al.  Interfractional positional variability of fiducial markers and primary tumors in locally advanced non-small-cell lung cancer during audiovisual biofeedback radiotherapy. , 2012, International journal of radiation oncology, biology, physics.

[17]  K. Shibuya,et al.  Evaluation of a combined respiratory-gating system comprising the TrueBeam linear accelerator and a new real-time tumor-tracking radiotherapy system: a preliminary study. , 2016, Journal of applied clinical medical physics.

[18]  M. Hoogeman,et al.  Clinical accuracy of the respiratory tumor tracking system of the cyberknife: assessment by analysis of log files. , 2009, International journal of radiation oncology, biology, physics.