Comparison of a cardiorespiratory device versus polysomnography for diagnosis of sleep apnoea

This study assessed the accuracy of a cardiorespiratory monitoring device versus polysomnography for the diagnosis of suspected sleep apnoea/hypoponea syndrome (SAS). A total of 86 patients (89% male, mean age 52 yrs) that had been referred to a sleep laboratory with a clinical diagnosis of SAS underwent cardiorespiratory polygraphy in an unattended mode using an ambulatory device (MERLIN). Analysis was carried out both automatically and manually. Conventional overnight full-channel polysomnography was performed simultaneously. Valid polygraphical recordings were obtained from 79 patients. The mean±sd apnoea/hypopnoea index (AHI) was 34.4±29.2. The results obtained with manual scoring were superior to automatic scoring for all AHI thresholds. For an AHI of ≥5, which is diagnostic SAS, the optimum cut-off value for the manual respiratory event index was 6.7 and the cardiorespiratory monitoring device had 97.1% sensitivity and 90.9% specificity. Correct classification according to the different cut-off points obtained via polysomnography and the corresponding cut-off points in the MERLIN manual index were confirmed in 90–96% of patients. The MERLIN device is a useful diagnostic approach for the initial assessment of adult patients with clinical suspicion of sleep apnoea/hypopnoea syndrome. Manual scoring is clearly better than automatic scoring in terms of agreement with the apnoea/hypopnoea index and to discern patients with sleep apnoea/hypopnoea syndrome.

[1]  D. Kripke,et al.  Comparisons of home sleep recordings and polysomnograms in older adults with sleep disorders. , 1981, Sleep.

[2]  S D Ross,et al.  Systematic review and meta-analysis of the literature regarding the diagnosis of sleep apnea. , 2000, Sleep.

[3]  E. Svanborg,et al.  A limited diagnostic investigation for obstructive sleep apnea syndrome. Oximetry and static charge sensitive bed. , 1990, Chest.

[4]  T. Verse,et al.  Validation of the POLY-MESAM seven-channel ambulatory recording unit. , 2000, Chest.

[5]  Guilleminault Suspicion of sleep-disordered breathing: which test to perform? , 2000, Sleep Medicine.

[6]  Martin Reite,et al.  ASDA standards of practice: Practice parameters for the use of portable recording in the assessment of obstructive sleep apnea , 1994 .

[7]  M A Sackner Indications and standards for cardiopulmonary sleep studies. , 1989, The American review of respiratory disease.

[8]  T. Young,et al.  The occurrence of sleep-disordered breathing among middle-aged adults. , 1993, The New England journal of medicine.

[9]  A J Block,et al.  Indications and standards for cardiopulmonary sleep studies. , 1985, Sleep.

[10]  C. Guilleminault,et al.  EEG arousals: scoring rules and examples: a preliminary report from the Sleep Disorders Atlas Task Force of the American Sleep Disorders Association. , 1992, Sleep.

[11]  N. Saunders,et al.  Evaluation of a microprocessor-based portable home monitoring system to measure breathing during sleep. , 1987, Sleep.

[12]  K P Strohl When, where and how to test for sleep apnea. , 2000, Sleep.

[13]  Bonnie K. Lind,et al.  Association of Sleep-Disordered Breathing, Sleep Apnea, and Hypertension in a Large Community-Based Study , 2000 .

[14]  F. B. Illa,et al.  Disponibilidad de recursos técnicos para el diagnóstico y tratamiento del síndrome de apnea obstructiva del sueño en los hospitales de la red pública del Estado , 1995 .

[15]  M. Littner,et al.  The indications for polysomnography and related procedures. , 1997, Sleep.

[16]  D. Altman,et al.  STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT , 1986, The Lancet.

[17]  R. Rodríguez-Roisín,et al.  Adequacy of prescribing nasal continuous positive airway pressure therapy for the sleep apnoea/hypopnoea syndrome on the basis of night time respiratory recording variables. , 1995, Thorax.

[18]  K. Clark,et al.  Association of Sleep-Disordered Breathing, Sleep Apnea, and Hypertension in a Large Community- Based Study , 2000 .

[19]  L. Ferini-Strambi,et al.  An unattended device for sleep-related breathing disorders: validation study in suspected obstructive sleep apnoea syndrome. , 1996, The European respiratory journal.

[20]  B. Raffestin,et al.  Evaluation of an ambulatory device, CID 102, in the diagnosis of obstructive sleep apnoea syndrome. , 1995, The European respiratory journal.

[21]  E. Wolpert A Manual of Standardized Terminology, Techniques and Scoring System for Sleep Stages of Human Subjects. , 1969 .

[22]  S. Redline,et al.  Measurement of sleep-related breathing disturbances in epidemiologic studies. Assessment of the validity and reproducibility of a portable monitoring device. , 1991, Chest.

[23]  A. Rechtschaffen,et al.  A manual of standardized terminology, technique and scoring system for sleep stages of human subjects , 1968 .

[24]  C. Infante-Rivard,et al.  Diagnostic accuracy of a portable recording device (MESAM IV) in suspected obstructive sleep apnoea. , 1996, The European respiratory journal.

[25]  D T Berry,et al.  What is hypopnea, anyway? , 1994, Chest.

[26]  AdrianJ Williams,et al.  Clinical value of polysomnography , 1992, The Lancet.

[27]  G. Man,et al.  Validation of a portable sleep apnea monitoring device. , 1995, Chest.

[28]  E. Bixler,et al.  Effects of age on sleep apnea in men: I. Prevalence and severity. , 1998, American journal of respiratory and critical care medicine.