Calculation and analysis of DQE for some image detectors in mammography

The development and clinical testing of digital detector designs for mammography are making rapid progress and there is widespread interest in comparing the performance of these new detectors to that of analogue screen-film mammography systems. In this report, Monte Carlo calculations of the x-ray absorption characteristics (single-event distribution), the quantum absorption, Aq, and detective quantum, DQEq, efficiencies are made and compared to results from the literature. Detectors of CsI and Si of various thicknesses are compared to a state-ofthe art analogue, screen-film system (Gd2O2S) in the energy range 1-35 keV. The results show that 1.5 mm thick Si detectors will have the same DQEq as commonly used Gd2O2S fluorescent screens and that a CsI phosphor of 80 mm has similar DQEq as 1.0 mm Si. The total DQE (including added noise and inherent detector unsharpness) of fluorescent screen-film systems will be significantly reduced from this value due to the light scatter, film noise and the inherent limitations caused by the film characteristic curve. This indicates that also thinner Si detectors (0.3-0.5 mm), which do not suffer from these limitations but from a comparably low Aq may have a total performance DQE(f) comparable to that of traditional screen-film based, image detectors.

[1]  L. Mandel,et al.  Image fluctuations in cascade intensifiers , 1959 .

[2]  A. H. Wapstra,et al.  Nuclear Data Tables , 1961 .

[3]  Lellery Storm,et al.  Photon cross sections from 1 keV to 100 MeV for elements Z=1 to Z=100 , 1970 .

[4]  R. K. Swank Absorption and noise in x‐ray phosphors , 1973 .

[5]  J. H. Hubbell,et al.  Atomic form factors, incoherent scattering functions, and photon scattering cross sections , 1975 .

[6]  M Danos,et al.  Image information content and patient exposure. , 1978, Medical physics.

[7]  J. H. Hubbell,et al.  Relativistic atomic form factors and photon coherent scattering cross sections , 1979 .

[8]  C E Dick,et al.  Image information transfer properties of x-ray fluorescent screens. , 1981, Medical physics.

[9]  J A Rowlands,et al.  Absorption and noise in cesium iodide x-ray image intensifiers. , 1983, Medical physics.

[10]  K Doi,et al.  Studies of x-ray energy absorption and quantum noise properties of x-ray screens by use of Monte Carlo simulation. , 1984, Medical physics.

[11]  M J Yaffe,et al.  Model of the spatial-frequency-dependent detective quantum efficiency of phosphor screens. , 1990, Medical physics.

[12]  M J Yaffe,et al.  Effect of various noise sources on the detective quantum efficiency of phosphor screens. , 1990, Medical physics.

[13]  G. A. Carlsson,et al.  Influence on X-ray energy spectrum, contrasting detail and detector on the signal-to-noise ratio (SNR) and detective quantum efficiency (DQE) in projection radiography , 1992 .

[14]  C E Dick,et al.  Image information transfer properties of x-ray intensifying screens in the energy range from 17 to 320 keV. , 1993, Medical physics.

[15]  J A Rowlands,et al.  X-ray imaging with amorphous selenium: detective quantum efficiency of photoconductive receptors for digital mammography. , 1995, Medical physics.

[16]  Martin J. Yaffe,et al.  Solid state x-ray detectors for digital mammography , 1995, Medical Imaging.

[17]  J A Rowlands,et al.  X-ray imaging using amorphous selenium: inherent spatial resolution. , 1995, Medical physics.

[18]  J A Rowlands,et al.  X-ray detectors for digital radiography. , 1997, Physics in medicine and biology.