Rib-cage-movement measurements as a potential new trigger signal in non-invasive mechanical ventilation

Non-invasive ventilation performed through an oronasal mask is a standard in clinical and homecare mechanical ventilation. Besides all its advantages, inevitable leaks through the mask cause errors in the feedback information provided by the airflow sensor and, hence, patient-ventilator asynchrony with multiple negative consequences. Here we investigate a new way to provide a trigger to the ventilator. The method is based on the measurement of rib cage movement at the onset of inspiration and during breathing by fibre-optic sensors. In a series of simultaneous measurements by a long-period fibre grating sensor and pneumotachograph we provide the statistical evidence of the 200 ms lag of the pneumo with respect the fibre-optic signal. The lag is registered consistently across three independent delay metrics. Further, we discuss exceptions from this trend and identify the needed improvements to the proposed fibre-sensing scheme.

[1]  S. Epstein How Often Does Patient-Ventilator Asynchrony Occur and What Are the Consequences? , 2011, Respiratory Care.

[2]  D. Hess Patient-Ventilator Interaction During Noninvasive Ventilation , 2011, Respiratory Care.

[3]  S. James,et al.  Optical fibre long-period grating sensors: characteristics and application , 2003 .

[4]  Philippe Jolliet,et al.  Performance of noninvasive ventilation modes on ICU ventilators during pressure support: a bench model study , 2007, Intensive Care Medicine.

[5]  P. Jorens,et al.  Neurally Adjusted Ventilatory Assist: A Ventilation Tool or a Ventilation Toy? , 2011, Respiratory Care.

[6]  M. Antonelli,et al.  Predictors of failure of noninvasive positive pressure ventilation in patients with acute hypoxemic respiratory failure: a multi-center study , 2001, Intensive Care Medicine.

[7]  Ian Bennion,et al.  Application of long-period-grating sensors to respiratory plethysmography. , 2007, Journal of biomedical optics.

[8]  M. Gentile Cycling of the Mechanical Ventilator Breath , 2011, Respiratory Care.

[9]  Martin Albert,et al.  Patient-ventilator interaction during pressure support ventilation and neurally adjusted ventilatory assist* , 2010, Critical care medicine.

[10]  L. Lloyd,et al.  pplication of long-period-grating sensors to respiratory lethysmography , 2007 .

[11]  B Bojović,et al.  Non-invasive respiratory monitoring using long-period fiber grating sensors. , 2014, Biomedical optics express.

[12]  B. Make,et al.  Non-invasive ventilation: From the past to the present , 2010, Non-Invasive Ventilation and Weaning.

[13]  Laurent Brochard,et al.  Nonconventional ventilation techniques , 2013, Current Opinion in Critical Care.