Fixation of long bone segmental defects: a biomechanical study.

OBJECTIVES Obtaining stable fixation in cases of long bone non-union with segmental bone defects can be challenging. Bone quality is often sub-optimal. Locking plates and structural allografts have both been used clinically in these cases. The objective of this study was to determine the biomechanical characteristics of three constructs that have been employed in this context. METHODS A biomechanical study was performed using 3rd Generation Composite Femurs as specimens. A diaphyseal segmental defect was created and fixed with one of three constructs: (1) lateral locking plate (LP); (2) lateral non-locking plate and medial allograft strut (S); (3) lateral non-locking plate and intramedullary fibula allograft (F). The "allografts" were fashioned from 3rd generation composite bones. Axial, torsional and bending stiffness as well as load to failure were determined using a materials testing machine. RESULTS Overall, construct S was the stiffest and construct LP was the least stiff. Construct F had intermediate characteristics. Axial load to failure for construct S (6108N) and for construct F (5344N) was significantly greater than for construct LP (2855N). CONCLUSION When maximal stiffness is desired, a construct with a structural allograft should be chosen over a locking plate. However, biological and anatomic factors must also be taken into account when using these constructs clinically.

[1]  G. Zych,et al.  Salvage of Humeral Nonunions With Onlay Bone Plate Allograft Augmentation , 2001, Clinical orthopaedics and related research.

[2]  Erik N. Kubiak,et al.  Biomechanics of Locked Plates and Screws , 2004, Journal of orthopaedic trauma.

[3]  M. Bosse,et al.  Unstable Proximal Extraarticular Tibia Fractures: A Biomechanical Evaluation of Four Methods of Fixation , 2004, Journal of orthopaedic trauma.

[4]  R. Zdero,et al.  Cortical screw pullout strength and effective shear stress in synthetic third generation composite femurs. , 2007, Journal of biomechanical engineering.

[5]  Erik N. Kubiak,et al.  Fixation of diaphyseal fractures with a segmental defect: a biomechanical comparison of locked and conventional plating techniques. , 2006, The Journal of trauma.

[6]  T. Wright,et al.  A CLINICAL AND BIOMECHANICAL STUDY , 1993 .

[7]  D. Ring,et al.  The functional outcome of operative treatment of ununited fractures of the humeral diaphysis in older patients. , 1999, The Journal of bone and joint surgery. American volume.

[8]  M. McKee,et al.  Treatment of osteopenic humeral shaft nonunion with compression plating, humeral cortical allograft struts, and bone grafting. , 2005, Journal of orthopaedic trauma.

[9]  P. Kloen,et al.  Locking Compression Plates for Osteoporotic Nonunions of the Diaphyseal Humerus , 2004, Clinical orthopaedics and related research.

[10]  P. Cole Endosteal Allograft Plating for the Treatment of Recalcitrant Nonunions , 2003 .

[11]  T D Brown,et al.  Structural properties of a new design of composite replicate femurs and tibias. , 2001, Journal of biomechanics.

[12]  L. Weng,et al.  Treatment of Distal Femoral Nonunion with Internal Fixation, Cortical Allograft Struts, and Autogenous Bone-Grafting , 2003, The Journal of bone and joint surgery. American volume.

[13]  D. Berry,et al.  Periprosthetic Femoral Fractures Around Well-Fixed Implants: Use of Cortical Onlay Allografts with or without a Plate , 2002, The Journal of bone and joint surgery. American volume.

[14]  M. Mcguire,et al.  Humeral shaft nonunions treated with fibular allograft and compression plating. , 2000, American journal of orthopedics.

[15]  R Zdero,et al.  The biomechanics of human femurs in axial and torsional loading: comparison of finite element analysis, human cadaveric femurs, and synthetic femurs. , 2007, Journal of biomechanical engineering.

[16]  L Cristofolini,et al.  Mechanical validation of whole bone composite femur models. , 1996, Journal of biomechanics.