Intraoperative neurophysiological monitoring and spinal cord stimulator implantation

Introduction Spinal cord injury (SCI) is one of the most dreaded complications after spinal cord stimulation (SCS) implantation surgery. As a result, intraoperative neurophysiological monitoring (IONM) has been proposed to avoid accidental damage to nervous structures under anesthesia and confirm positioning for optimal stimulation. Our study uses a large administrative claims database to determine the 30-day risk of SCI after SCS implantation. Methods This retrospective cohort study used the IBM MarketScan Commercial and Medicare Supplemental Databases from 2016 to 2019. Adult patients undergoing SCS surgical procedures with at least 90 days of follow-up, IONM use, the type of sedation used during the procedure, and subsequent SCI were identified using administrative codes. In addition, logistic regression was used to examine the relationship between various risk factors and subsequent SCI. Results A total of 9676 patients underwent SCS surgery (64.7% percutaneous implants) during the study period. Nine hundred and forty-four (9.75%) patients underwent SCS implantation with IONM. Conscious sedation, Monitored Anesthesia Care anesthesia, and general anesthesia were used in patients with 0.9%, 60.2%, and 28.6%, respectively. Eighty-one (0.8%) patients developed SCI within 30 days after SCS implant surgery. The SCI rate was higher in the group that underwent IONM (2% vs 0.7%, p value <0.001) during the implantation procedure, reflecting the underlying risk. After adjustment for other factors, the OR of SCI is 2.39 (95% CI: 1.33 to 4.14, p value=0.002) times higher for those with IONM than those without IONM. Conclusions Increased SCI risk among patients with IONM likely reflects higher baseline risk, and further research is needed for risk mitigation.

[1]  Safdar N. Khan,et al.  Evaluating the incidence of spinal cord injury after spinal cord stimulator implant: an updated retrospective review , 2022, Regional Anesthesia & Pain Medicine.

[2]  John L. Parker,et al.  A Prospective Multicenter Case Series Utilizing Intraoperative Neuromonitoring With Evoked Compound Action Potentials to Confirm Spinal Cord Stimulation Lead Placement. , 2022, Neuromodulation : journal of the International Neuromodulation Society.

[3]  Mark Rubin,et al.  When to adjust alpha during multiple testing: a consideration of disjunction, conjunction, and individual testing , 2021, Synthese.

[4]  Dave A. Thomas,et al.  Practice , 2019, IEEE Softw..

[5]  Susan H. Morris,et al.  Correction to: Is the new ASNM intraoperative neuromonitoring supervision “guideline” a trustworthy guideline? A commentary , 2019, Journal of Clinical Monitoring and Computing.

[6]  A. Sdrulla,et al.  Spinal Cord Stimulation: Clinical Efficacy and Potential Mechanisms , 2018, Pain practice : the official journal of World Institute of Pain.

[7]  E. Ritzl,et al.  Practice guidelines for the supervising professional: intraoperative neurophysiological monitoring , 2018, Journal of Clinical Monitoring and Computing.

[8]  J. Arle,et al.  Neuromonitoring for Spinal Cord Stimulation Lead Placement Under General Anesthesia , 2018, Journal of clinical neurology.

[9]  N. Bedforth,et al.  Spinal cord injury arising in anaesthesia practice , 2018, Anaesthesia.

[10]  L. Álvarez,et al.  Osteoporosis and the Management of Spinal Degenerative Disease (II). , 2017, The archives of bone and joint surgery.

[11]  Tyler J. VanderWeele,et al.  Sensitivity Analysis in Observational Research: Introducing the E-Value , 2017, Annals of Internal Medicine.

[12]  K. Slavin,et al.  The Neurostimulation Appropriateness Consensus Committee (NACC) Safety Guidelines for the Reduction of Severe Neurological Injury , 2017, Neuromodulation : journal of the International Neuromodulation Society.

[13]  S Harrison Farber,et al.  The Incidence of Spinal Cord Injury in Implantation of Percutaneous and Paddle Electrodes for Spinal Cord Stimulation , 2016, Neuromodulation : journal of the International Neuromodulation Society.

[14]  E. Buchser,et al.  Complications of Spinal Cord Stimulation and Peripheral Nerve Stimulation Techniques: A Review of the Literature. , 2015, Pain medicine.

[15]  S. Tummala,et al.  Intraoperative neurophysiologic monitoring with Hoffmann reflex during thoracic spine surgery , 2015, Journal of Clinical Neuroscience.

[16]  K. Slavin,et al.  The Appropriate Use of Neurostimulation of the Spinal Cord and Peripheral Nervous System for the Treatment of Chronic Pain and Ischemic Diseases: The Neuromodulation Appropriateness Consensus Committee , 2014, Neuromodulation : journal of the International Neuromodulation Society.

[17]  J. Shils,et al.  Intraoperative Neurophysiologic Methods for Spinal Cord Stimulator Placement Under General Anesthesia , 2012, Neuromodulation : journal of the International Neuromodulation Society.

[18]  K. Slavin,et al.  Incidence and Avoidance of Neurologic Complications with Paddle Type Spinal Cord Stimulation Leads , 2011, Neuromodulation : journal of the International Neuromodulation Society.

[19]  A. Sestokas,et al.  Awake vs. Asleep Placement of Spinal Cord Stimulators: A Cohort Analysis of Complications Associated With Placement , 2011, Neuromodulation : journal of the International Neuromodulation Society.

[20]  Clark C Smith,et al.  A report of paraparesis following spinal cord stimulator trial, implantation and revision. , 2010, Pain physician.

[21]  P. Austin,et al.  A Modification of the Elixhauser Comorbidity Measures Into a Point System for Hospital Death Using Administrative Data , 2009, Medical care.

[22]  Y. Hirabayashi,et al.  The lumbar extradural structure changes with increasing age. , 1997, British journal of anaesthesia.

[23]  Y. Hirabayashi,et al.  Inflammatory changes after extradural anaesthesia may affect the spread of local anaesthetic within the extradural space. , 1996, British journal of anaesthesia.

[24]  J. Lopez,et al.  Intraoperative Neurophysiological Monitoring , 1996, International anesthesiology clinics.