From plane to spatial angles: PTB’s spatial angle autocollimator calibrator

Abstract Electronic autocollimators are utilised versatilely for non-contact angle measurements in applications like straightness measurements and profilometry. Yet, no calibration of the angle measurement of an autocollimator has been available when both its measurement axes are engaged. Additionally, autocollimators have been calibrated at fixed distances to the reflector, although its distance may vary during the use of an autocollimator. To extend the calibration capabilities of the Physikalisch-Technische Bundesanstalt (PTB) regarding spatial angles and variable distances, a novel calibration device has been set up: the spatial angle autocollimator calibrator (SAAC). In this paper, its concept and its mechanical realisation will be presented. The focus will be on the system’s mathematical modelling and its application in spatial angle calibrations. The model considers the misalignments of the SAAC’s components, including the non-orthogonalities of the measurement axes of the autocollimators and of the rotational axes of the tilting unit. It allows us to derive specific measurement procedures to determine the misalignments in situ and, in turn, to correct the measurements of the autocollimators. Finally, the realisation and the results of a traceable spatial angle calibration of an autocollimator will be presented. This is the first calibration of this type worldwide.

[1]  G. G. Stokes "J." , 1890, The New Yale Book of Quotations.

[2]  Mourad Idir,et al.  Modern developments in X-ray and neutron optics , 2008 .

[3]  P. Cochat,et al.  Et al , 2008, Archives de pediatrie : organe officiel de la Societe francaise de pediatrie.

[4]  Andreas Just,et al.  Calibration of high-resolution electronic autocollimators against an angle comparator , 2003 .

[5]  Garth J. Williams,et al.  Ultra-precise characterization of LCLS hard X-ray focusing mirrors by high resolution slope measuring deflectometry. , 2012, Optics express.