Computed Tomography (CT) Automatic Exposure Controls (AEC) Testing Protocol Using a CelT Phantom
暂无分享,去创建一个
The purpose of this research was to set-up a protocol for
using the CeIT elliptical test phantom to test the
performance of Automatic Exposure Control (AEC) systems
on the CT scanners in use at Aberdeen Royal Infirmary (ARI).
These are the GE Lightspeed, GE Optima 6600 and Siemens
Somaton Definition in Radiology and the Philips Brilliance in
Radiotherapy treatment planning. The variation of image
noise and the tube current-time product (mAs) were
studied from images obtained from each scanner. Noise was
measured using the standard deviation of five selected
regions of interest in the images of the phantom obtained
from the CT scanners. Normalised percentage noise (noise
%) was then calculated to compare how the scanners dealt
with image noise with relation to the mAs. The results
showed an increase in mAs values (increase in dose) with
the phantom and regulation of the noise leading to
acquisition of quality images from all three scanners. Offcentering,
using the AP scout increased the dose to the
phantom when the patient table was above the isocenter
and reduced the dose to the phantom when the table was
below the isocenter. This shows the importance of patient
centering for effective AEC system ’ s dose regulation.
Different SPRs also affected the operations of the AEC
systems differently, with PA giving more dose followed by
LAT and AP in the GE and Philips scanners which were
studied on this aspect. The Philips D-DOM modulation kept
almost constant dose across the scanning process regardless
of the phantom size, hence D-DOM should be used with
care. CeLT phantom was useful in studying dose regulation
by different AEC systems, hence it is useful in quality control
(QC) tests of AEC systems on CT scanners as a testing
protocol was formulated from this study.