Limited-angle effect compensation for respiratory binned cardiac SPECT.

PURPOSE In cardiac single photon emission computed tomography (SPECT), respiratory-binned study is used to combat the motion blur associated with respiratory motion. However, owing to the variability in respiratory patterns during data acquisition, the acquired data counts can vary significantly both among respiratory bins and among projection angles within individual bins. If not properly accounted for, such variation could lead to artifacts similar to limited-angle effect in image reconstruction. In this work, the authors aim to investigate several reconstruction strategies for compensating the limited-angle effect in respiratory binned data for the purpose of reducing the image artifacts. METHODS The authors first consider a model based correction approach, in which the variation in acquisition time is directly incorporated into the imaging model, such that the data statistics are accurately described among both the projection angles and respiratory bins. Afterward, the authors consider an approximation approach, in which the acquired data are rescaled to accommodate the variation in acquisition time among different projection angles while the imaging model is kept unchanged. In addition, the authors also consider the use of a smoothing prior in reconstruction for suppressing the artifacts associated with limited-angle effect. In our evaluation study, the authors first used Monte Carlo simulated imaging with 4D NCAT phantom wherein the ground truth is known for quantitative comparison. The authors evaluated the accuracy of the reconstructed myocardium using a number of metrics, including regional and overall accuracy of the myocardium, uniformity and spatial resolution of the left ventricle (LV) wall, and detectability of perfusion defect using a channelized Hotelling observer. As a preliminary demonstration, the authors also tested the different approaches on five sets of clinical acquisitions. RESULTS The quantitative evaluation results show that the three compensation methods could all, but to different extents, reduce the reconstruction artifacts over no compensation. In particular, the model based approach reduced the mean-squared-error of the reconstructed myocardium by as much as 40%. Compared to the approach of data rescaling, the model based approach further improved both the overall and regional accuracy of the myocardium; it also further improved the lesion detectability and the uniformity of the LV wall. When ML reconstruction was used, the model based approach was notably more effective for improving the LV wall; when MAP reconstruction was used, the smoothing prior could reduce the noise level and artifacts with little or no increase in bias, but at the cost of a slight resolution loss of the LV wall. The improvements in image quality by the different compensation methods were also observed in the clinical acquisitions. CONCLUSIONS Compensating for the uneven distribution of acquisition time among both projection angles and respiratory bins can effectively reduce the limited-angle artifacts in respiratory-binned cardiac SPECT reconstruction. Direct incorporation of the time variation into the imaging model together with a smoothing prior in reconstruction can lead to the most improvement in the accuracy of the reconstructed myocardium.

[1]  Yongyi Yang,et al.  Effects of piecewise smoothing on cardiac SPECT reconstruction , 2011, 2011 18th IEEE International Conference on Image Processing.

[2]  T Kumazaki,et al.  Development of respiratory gated myocardial SPECT system , 1999, Journal of nuclear cardiology : official publication of the American Society of Nuclear Cardiology.

[3]  William Paul Segars,et al.  Development of a new dynamic NURBS-based cardiac-torso (NCAT) phantom , 2001 .

[4]  P. Börnert,et al.  Free-breathing cardiac MR imaging: study of implications of respiratory motion--initial results. , 2001, Radiology.

[5]  J. Cooper,et al.  Effect of patient motion on tomographic myocardial perfusion imaging. , 1992, Journal of nuclear medicine : official publication, Society of Nuclear Medicine.

[6]  D. Low,et al.  A comparison between amplitude sorting and phase-angle sorting using external respiratory measurement for 4D CT. , 2006, Medical physics.

[7]  A J Reader,et al.  List-mode-based reconstruction for respiratory motion correction in PET using non-rigid body transformations , 2007, Physics in medicine and biology.

[8]  Michael A. King,et al.  Compensating for limited-angle effect in respiratory-gated cardiac SPECT , 2014, 2014 IEEE 11th International Symposium on Biomedical Imaging (ISBI).

[9]  Simon Rit,et al.  Comparison of Analytic and Algebraic Methods for Motion-Compensated Cone-Beam CT Reconstruction of the Thorax , 2009, IEEE Transactions on Medical Imaging.

[10]  Yongyi Yang,et al.  Effects of motion, attenuation, and scatter corrections on gated cardiac SPECT reconstruction. , 2011, Medical physics.

[11]  Yongyi Yang,et al.  A quantitative evaluation study of four-dimensional gated cardiac SPECT reconstruction. , 2009, Physics in medicine and biology.

[12]  Jeffrey A. Fessler,et al.  Globally convergent image reconstruction for emission tomography using relaxed ordered subsets algorithms , 2003, IEEE Transactions on Medical Imaging.

[13]  Debasis Mitra,et al.  SinoCor: motion correction in SPECT , 2012, Medical Imaging.

[14]  M. Ljungberg,et al.  A Monte Carlo program for the simulation of scintillation camera characteristics. , 1989, Computer methods and programs in biomedicine.

[15]  J. E. Hansen,et al.  Updated imaging guidelines for nuclear cardiology procedures, part 1. , 2001, Journal of nuclear cardiology : official publication of the American Society of Nuclear Cardiology.

[16]  Michael A. King,et al.  Effects of Piecewise Spatial Smoothing in 4-D SPECT Reconstruction , 2014, IEEE Transactions on Nuclear Science.

[17]  Jeffrey A. Fessler,et al.  Joint estimation of respiratory motion and activity in 4D PET using CT side information , 2006, 3rd IEEE International Symposium on Biomedical Imaging: Nano to Macro, 2006..

[18]  H H Barrett,et al.  Addition of a channel mechanism to the ideal-observer model. , 1987, Journal of the Optical Society of America. A, Optics and image science.

[19]  Xiaofeng Niu,et al.  A quantitative study of motion estimation methods on 4D cardiac gated SPECT reconstruction. , 2012, Medical physics.

[20]  G. J. Klein,et al.  Real-time system for respiratory-cardiac gating in positron tomography , 1997 .

[21]  W. Segars,et al.  Study of the efficacy of respiratory gating in myocardial SPECT using the new 4-D NCAT phantom , 2001 .

[22]  Michael A. King,et al.  AN INVESTIGATION OF THE FILTERING OF TEW SCATTER ESTIMATES USED TO COMPENSATE FOR SCATTER WITH ORDERED SUBSET RECONSTRUCTIONS , 1996 .

[23]  Katsuhiko Ueda,et al.  Correction of liver displacement due to irregular respiration for SPECT images obtained using a multiple short-time acquisition with breath-holding technique , 2011, Radiological Physics and Technology.

[24]  Jason G Parker,et al.  Respiratory motion correction in gated cardiac SPECT using quaternion-based, rigid-body registration. , 2009, Medical physics.

[25]  Joyoni Dey,et al.  Estimation and correction of cardiac respiratory motion in SPECT in the presence of limited-angle effects due to irregular respiration. , 2010, Medical physics.