A comprehensive system for dosimetric commissioning and Monte Carlo validation for the small animal radiation research platform

Our group has constructed the small animal radiation research platform (SARRP) for delivering focal, kilo-voltage radiation to targets in small animals under robotic control using cone-beam CT guidance. The present work was undertaken to support the SARRP's treatment planning capabilities. We have devised a comprehensive system for characterizing the radiation dosimetry in water for the SARRP and have developed a Monte Carlo dose engine with the intent of reproducing these measured results. We find that the SARRP provides sufficient therapeutic dose rates ranging from 102 to 228 cGy min(-1) at 1 cm depth for the available set of high-precision beams ranging from 0.5 to 5 mm in size. In terms of depth-dose, the mean of the absolute percentage differences between the Monte Carlo calculations and measurement is 3.4% over the full range of sampled depths spanning 0.5-7.2 cm for the 3 and 5 mm beams. The measured and computed profiles for these beams agree well overall; of note, good agreement is observed in the profile tails. Especially for the smallest 0.5 and 1 mm beams, including a more realistic description of the effective x-ray source into the Monte Carlo model may be important.

[1]  Peter Kazanzides,et al.  High-resolution, small animal radiation research platform with x-ray tomographic guidance capabilities. , 2008, International journal of radiation oncology, biology, physics.

[2]  C Ross Schmidtlein,et al.  Precise radiochromic film dosimetry using a flat-bed document scanner. , 2005, Medical physics.

[3]  S Mutic,et al.  MicroRT-small animal conformal irradiator. , 2007, Medical physics.

[4]  Christopher H Contag,et al.  Design and evaluation of a variable aperture collimator for conformal radiotherapy of small animals using a microCT scanner. , 2007, Medical physics.

[5]  Peter Kazanzides,et al.  Image-guided small animal radiation research platform: calibration of treatment beam alignment , 2009, Physics in medicine and biology.

[6]  Aapm A protocol for the determination of absorbed dose from high-energy photon and electron beams. , 1983, Medical physics.

[7]  David A. Jaffray,et al.  WE‐D‐330A‐09: An Image‐Guided Irradiator for Pre‐Clinical Radiation Therapy Studies , 2006 .

[8]  J. H. Hubbell,et al.  XCOM: Photon cross sections on a personal computer , 1987 .

[9]  Frank Verhaegen,et al.  Monte Carlo modelling of external radiotherapy photon beams. , 2003, Physics in medicine and biology.

[10]  F Verhaegen,et al.  Monte Carlo modelling of radiotherapy kV x-ray units. , 1999, Physics in medicine and biology.

[11]  C. Ma,et al.  BEAM: a Monte Carlo code to simulate radiotherapy treatment units. , 1995, Medical physics.