Design and fabrication of a multi-electrode array for spinal cord epidural stimulation

A detailed design, fabrication, characterization and test of a flexible multi-site platinum/polyimide based electrode array for electrical epidural stimulation in spinal cord prosthesis is described in this paper. Carefully designed 8.4 μm-thick structure fabrication flow achieves an electrode surface modification with 3.8 times enhanced effective surface area without extra process needed. Measured impedance and phase of two type of electrodes are 2.35±0.21 KΩ and 2.10±0.11 KΩ, -34.25±8.07° and -27.71±8.27° at 1K Hz, respectively. The fabricated arrays were then in-vitro tested by a multichannel neural stimulation system in physiological saline to validate the capability for electrical stimulation. The measured channel isolation on adjacent electrode is about -34dB. Randles cell model was used to investigate the charging waveforms, the model parameters were then extracted by various methods. The measured charge transfer resistance, double layer capacitance, and solution resistance are 1.9 KΩ, 220 nF and 15 KΩ, respectively. The results show that the fabricated array is applicable for electrical stimulation with well characterized parameters. Combined with a multichannel stimulator, this system provides a full solution for versatile neural stimulation applications.

[1]  R. Stephenson A and V , 1962, The British journal of ophthalmology.

[2]  Madhav Datta,et al.  Anodic dissolution of metals at high rates , 1993, IBM J. Res. Dev..

[3]  Hong Ying,et al.  Effect of temperature on etch rate of iridium and platinum in CF4/O2 , 2001 .

[4]  Alireza Zolfaghari,et al.  Capacitance of the double-layer at polycrystalline Pt electrodes bearing a surface-oxide film , 2002 .

[5]  Daniel R. Merrill,et al.  Electrical stimulation of excitable tissue: design of efficacious and safe protocols , 2005, Journal of Neuroscience Methods.

[6]  Hui Zhong,et al.  Plasticity of spinal cord reflexes after a complete transection in adult rats: relationship to stepping ability. , 2006, Journal of neurophysiology.

[7]  Sten Grillner,et al.  Biological Pattern Generation: The Cellular and Computational Logic of Networks in Motion , 2006, Neuron.

[8]  Igor A. Lavrov,et al.  Transformation of nonfunctional spinal circuits into functional states after the loss of brain input , 2009, Nature Neuroscience.

[9]  V. Reggie Edgerton,et al.  A parylene-based microelectrode array implant for spinal cord stimulation in rats , 2011, 2011 IEEE 24th International Conference on Micro Electro Mechanical Systems.

[10]  Yu-Chong Tai,et al.  Development of a multi-electrode array for spinal cord epidural stimulation to facilitate stepping and standing after a complete spinal cord injury in adult rats , 2013, Journal of neuroengineering and rehabilitation.

[11]  V. Reggie Edgerton,et al.  Using in vivo spinally-evoked potentials to assess functional connectivity along the spinal axis , 2013, 2013 6th International IEEE/EMBS Conference on Neural Engineering (NER).

[12]  Vincent Adrianus Henneken,et al.  Residue-free plasma etching of polyimide coatings for small pitch vias with improved step coverage , 2013 .

[13]  Laurent Francis,et al.  Polyimide as a versatile enabling material for microsystems fabrication: surface micromachining and electrodeposited nanowires integration , 2013 .

[14]  Wentai Liu,et al.  A Fully-Integrated High-Compliance Voltage SoC for Epi-Retinal and Neural Prostheses , 2013, IEEE Transactions on Biomedical Circuits and Systems.