Gelled versus nongelled phantom material for measurement of MRI-induced temperature increases with bioimplants

Measurements in phantoms are used to predict temperature changes that would occur in vivo for medical implants due to the radio frequency (RF) field in magnetic resonance imaging (MRI). In this study, the impact of concentration of the gelling agent in a saline-based phantom on the RF-induced temperature rise was measured using an apparatus that accurately reproduces the RF environment present in a 1.5-T whole-body MR system. The temperature was measured using fluoroptic thermometry at the electrode and other sites for a deep brain neurostimulation system. The average power deposition in the 30-kg phantom was about 1.5 W/kg. Four phantom formulations were evaluated, using different concentrations of polyacrylic acid (PAA) added to saline solution, with NaCl concentration adjusted to maintain an electrical conductivity near 0.24 S/m. The greatest temperature rises occurred at the electrode, ranging from 16.2/spl deg/C for greatest concentration of PAA to 2.9/spl deg/C for only saline solution. The temperature rise attained the maximal value for sufficient concentration of PAA. Similar behavior was observed in the temperature versus time relationship near a current-carrying resistor, immersed in gel and saline, which was used to model a localized heat source. The temperature rise for insufficient PAA concentration is reduced due to convection of phantom material. In conclusion, an appropriate gelling agent is required to accurately simulate the thermal properties of body tissues for measurements of RF-induced heating with medical implants.

[1]  Ashwini Sharan,et al.  Neurostimulation systems for deep brain stimulation: In vitro evaluation of magnetic resonance imaging–related heating at 1.5 tesla , 2002, Journal of magnetic resonance imaging : JMRI.

[2]  T W Athey A model of the temperature rise in the head due to magnetic resonance imaging procedures , 1989, Magnetic resonance in medicine.

[3]  Chris D. Smith Thermal effects of radiofrequency fields in magnetic resonance imaging on implanted medical devices , 2000 .

[4]  G Hartsgrove,et al.  Simulated biological materials for electromagnetic radiation absorption studies. , 1987, Bioelectromagnetics.

[5]  A. Guy,et al.  Formulas for preparing phantom muscle tissue at various radiofrequencies. , 1984, Bioelectromagnetics.

[6]  E. Larsson,et al.  MR Evaluation Ex Vivo and In Vivo of a Covered Stent‐Graft for Abdominal Aortic Aneurysms: Ferromagnetism, Heating, Artifacts, and Velocity Mapping , 2000, Journal of magnetic resonance imaging : JMRI.

[7]  Athey Tw A model of the temperature rise in the head due to magnetic resonance imaging procedures. , 1989 .

[8]  Georg Neubauer,et al.  Temperature measurement on neurological pulse generators during MR scans , 2002, Biomedical engineering online.

[9]  W. Nitz,et al.  On the heating of linear conductive structures as guide wires and catheters in interventional MRI , 2001, Journal of magnetic resonance imaging : JMRI.

[10]  M. J. Hagmann,et al.  Muscle-equivalent phantom materials for 10-100 MHz , 1992 .

[11]  B. Condon,et al.  Investigation of the factors responsible for burns during MRI , 2001, Journal of magnetic resonance imaging : JMRI.

[12]  Alexander V. Kildishev,et al.  Interactions of magnetic resonance imaging radio frequency magnetic fields with elongated medical implants , 2000 .

[13]  M. L. Wood,et al.  Magnetic Resonance Procedures: Health Effects and Safety , 2001 .