Can the effective number of bits be useful to assess the measurement uncertainty?

The effective number of bits (ENOB) is a well know parameter, employed to characterize the overall behavior of an A/D converter. In order to verify if the ENOB can be also used to assess the measurement uncertainty of the A/D conversion based instruments, we use a developed software tool. The results show that using the ENOB can lead to an underestimate of the uncertainties.