Adaptive Lung Ventilation (ALV) during Anesthesia for Pulmonary Surgery: Automatic Response to Transitions to and from One-Lung Ventilation

Adaptive lung ventilation is a novel closed-loop-controlled ventilation system. Based upon instantaneous breath-to-breath analyses, the ALV controller adjusts ventilation patterns automatically to momentary respiratory mechanics. Its goal is to provide a preset alveolar ventilation (V′ A) and, at the same time, minimize the work of breathing. Aims of our study were (1) to investigate changes in respiratory mechanics during transition to and from one-lung ventilation (OLV), (2) to describe the automated adaptation of the ventilatory pattern. Methods. With institutional approval and informed consent, 9 patients (33–72 y, 66–88 kg) underwent ALV during total intravenous anesthesia for pulmonary surgery. The ALV controller uses a pressure controlled ventilation mode. V′ A is preset by the anesthesiologist. Flow, pressure, and CO2 are continuously measured at the DLT connector. The signals were read into a IBM compatible PC and processed using a linear one-compartment model of the lung to calculate breath-by-breath resistance (R), compliance (C), respiratory time constant (TC), serial dead space (VdS) and V′ A. Based upon the results, the controller optimizes respiratory rate (RR) and tidal volume (VT) such as to achieve the preset V′ A with the minimum work of breathing. In addition to V′ A, only PEEP and FIO2 settings are at the anesthesiologist's discretion. All patients were ventilated using FIO2 = 1,0 and PEEP = 3 cm H2O. Parameters of respiratory mechanics, ventilation, and ABG were recorded during three 5-min periods: 10 min prior to OLV (I), 20 min after onset of OLV (II), and after chest closure (III). Data analyses used nonparametric comparisons of paired samples (Wilcoxon, Friedman) with Bonferroni's correction. Significance was assumed at p < 0.05. Values are given as medians (range). Results. 20 min after onset of OLV (II), resistance had approximately doubled compared with (I), compliance had decreased from 54 (36–81) to 50 (25–70) ml/cm H2O. TC remained stable at 1.4 (0.8–2.4) vs. 1.2 (0.9–1.6) s. Institution of OLV was followed by a reproducible response of the ALV controller. The sudden changes in respiratory mechanics caused a transient reduction in VT by 42 (8–59) %, with RR unaffected. In order to reestablish the preset V′ A, the controller increased inspiratory pressure in a stepwise fashion from 18 (14–23) to 27 (19–39) cm H2O, thereby increasing VT close to baseline (7.5 (6.6–9.0) ml/kg BW vs. 7.9 (5.4–11.7) ml/kg BW). The controller was, thus, effective in maintaining V′ A. The minimum PaO2 during phase II was 101 mmHg. After chest closure, respiratory mechanics had returned to baseline. Conclusions. Respiratory mechanics during transition to and from OLV are characterized by marked changes in R and C into opposite directions, leaving TC unaffected. The ALV controller manages these transitions successfully, and maintaines V′ A reliably without intervention by the anesthesiologist. VT during OLV was found to be consistently lower than recommended in the literature.

[1]  H RAHN,et al.  Mechanics of breathing in man. , 1950, Journal of applied physiology.

[2]  J. Mead,et al.  The control of respiratory frequency. , 1960, Annals of the New York Academy of Sciences.

[3]  H. Turndorf,et al.  Optimization of Arterial Oxygenation during One‐Lung Anesthesia , 1980, Anesthesia and analgesia.

[4]  J. Benumof,et al.  Improving Oxygenation during One‐lung Ventilation in Dogs: The Effects of Positive End‐Expiratory Pressure and Blood Flow Restriction to the Nonventilated Lung , 1981, Anesthesiology.

[5]  J. Benumof One‐Lung Ventilation and Hypoxic Pulmonary Vasoconstriction: Implications for Anesthetic Management , 1985, Anesthesia and analgesia.

[6]  R. Dellinger,et al.  Pulmonary Function Indices in Critical Care Patients , 1989 .

[7]  B. Marshall Hypoxic pulmonary vasoconstriction , 1990, Acta anaesthesiologica Scandinavica. Supplementum.

[8]  C. Mélot,et al.  Enhancement of hypoxic pulmonary vasoconstriction by metabolic acidosis in dogs. , 1990, Anesthesiology.

[9]  C. Mélot,et al.  Effects of acidosis and alkalosis on hypoxic pulmonary vasoconstriction in dogs. , 1990, The American journal of physiology.

[10]  E. Engelman,et al.  Continuous spirometry for detection of double-lumen endobronchial tube displacement. , 1993, British journal of anaesthesia.

[11]  W. Heinrichs,et al.  An adaptive lung ventilation controller , 1994, IEEE Transactions on Biomedical Engineering.

[12]  P. Rocmans,et al.  On-line expiratory flow-volume curves during thoracic surgery: occurrence of auto-PEEP. , 1994, British journal of anaesthesia.

[13]  N. Weiler,et al.  The AVL-mode: A safe closed loop algorithm for ventilation during total intravenous anesthesia , 1994, International journal of clinical monitoring and computing.

[14]  G. Iotti,et al.  Simple method to measure total expiratory time constant based on the passive expiratory flow-volume curve. , 1995, Critical care medicine.

[15]  J. Yernault,et al.  Intrinsic positive end-expiratory pressure during one-lung ventilation for thoracic surgery. The influence of preoperative pulmonary function. , 1996, Chest.

[16]  A. Sari,et al.  Auto-Positive End-Expiratory Pressure During One-Lung Ventilation Using a Double-Lumen Endobronchial Tube , 1996, Anesthesia and Analgesia.

[17]  N. Weiler,et al.  [Adaptive lung ventilation (AVL). Evaluation of new closed loop regulated respiration algorithm for operation in the hyperextended lateral position]. , 1996, Der Anaesthesist.

[18]  J. Eisenkraft,et al.  Positive end-expiratory pressure during one-lung ventilation improves oxygenation in patients with low arterial oxygen tensions. , 1996, Journal of cardiothoracic and vascular anesthesia.

[19]  Adaptive Lung Ventilation (ALV) Evaluierung eines neuen closed loop-gesteuerten Beatmungsalgorithmus bei Eingriffen in überstreckter Seitenlage , 1996, Der Anaesthesist.

[20]  R. Veronesi,et al.  Respiratory mechanics by least squares fitting in mechanically ventilated patients: Applications during paralysis and during pressure support ventilation , 1995, Intensive Care Medicine.