Automatic external defibrillators for public access defibrillation: recommendations for specifying and reporting arrhythmia analysis algorithm performance, incorporating new waveforms, and enhancing safety. A statement for health professionals from the American Heart Association Task Force on Automa

These recommendations are presented to enhance the safety and efficacy of AEDs intended for public access. The task force recommends that manufacturers present developmental and validation data on their own devices, emphasizing high sensitivity for shockable rhythms and high specificity for nonshockable rhythms. Alternative defibrillation waveforms may reduce energy requirements, reducing the size and weight of the device. The highest levels of safety for public access defibrillation are needed. Safe and effective use of AEDs that are widely available and easily handled by nonmedical personnel has the potential to dramatically increase survival from cardiac arrest.

[1]  George A. Johnson,et al.  Truncated biphasic pulses for transthoracic defibrillation. , 1995, Circulation.

[2]  J. Schuder,et al.  Defibrillation of 100 kg calves with asymmetrical, bidirectional, rectangular pulses. , 1984, Cardiovascular research.

[3]  J. Ornato,et al.  American Heart Association Report on the Public Access Defibrillation Conference December 8-10, 1994. Automatic External Defibrillation Task Force. , 1995, Circulation.

[4]  R. Kerber,et al.  Overlapping sequential pulses. A new waveform for transthoracic defibrillation. , 1994, Circulation.

[5]  R. Fletcher,et al.  Increasing fibrillation duration enhances relative asymmetrical biphasic versus monophasic defibrillator waveform efficacy. , 1990, Circulation research.

[6]  R. Kronmal,et al.  Multicenter comparison of truncated biphasic shocks and standard damped sine wave monophasic shocks for transthoracic ventricular defibrillation. Transthoracic Investigators. , 1996, Circulation.

[7]  W C Blackwelder,et al.  "Proving the null hypothesis" in clinical trials. , 1981, Controlled clinical trials.

[8]  Jerry H. Gold,et al.  Transthoracic Ventricular Defibrillation in the 100 kg Calf with Symmetrical One-Cycle Bidirectional Rectangular Wave Stimuli , 1983, IEEE Transactions on Biomedical Engineering.

[9]  P. D. Chapman,et al.  Comparison of monophasic and biphasic defibrillating pulse waveforms for transthoracic cardioversion. Biphasic Waveform Defibrillation Investigators. , 1995, The American journal of cardiology.