Template-guided vs. non-guided drilling in site preparation of dental implants

ObjectivesClinical success of oral implants is related to primary stability and osseointegration. These parameters are associated with delicate surgical techniques. We herein studied whether template-guided drilling has a significant influence on drillholes diameter and accuracy in an in vitro model.Materials and methodsFresh cadaveric porcine mandibles were used for drilling experiments of four experimental groups. Each group consisted of three operators, comparing guide templates for drilling with free-handed procedure. Operators without surgical knowledge were grouped together, contrasting highly experienced oral surgeons in other groups. A total of 180 drilling actions were performed, and diameters were recorded at multiple depth levels, with a precision measuring instrument.ResultsTemplate-guided drilling procedure improved accuracy on a very significant level in comparison with free-handed drilling operation (p ≤ 0.001). Inaccuracy of free-handed drilling became more significant in relation to measurement depth. High homogenic uniformity of template-guided drillholes was significantly stronger than unguided drilling operations by highly experienced oral surgeons (p ≤ 0.001).ConclusionTemplate-guided drilling procedure leads to significantly enhanced accuracy. Significant results compared to free-handed drilling actions were achieved, irrespective of the clinical experience level of the operator.Clinical relevanceTemplate-guided drilling procedures lead to a more predictable clinical diameter. It shows that any set of instruments has to be carefully chosen to match the specific implant system. The current in vitro study is implicating an improvement of implant bed preparation but needs to be confirmed in clinical studies.

[1]  Mihoko Atsumi,et al.  Methods used to assess implant stability: current status. , 2007, The International journal of oral & maxillofacial implants.

[2]  M. Zwahlen,et al.  A systematic review on the accuracy and the clinical outcome of computer-guided template-based implant dentistry. , 2009, Clinical oral implants research.

[3]  J. Jansen,et al.  Implant Surface Roughness and Bone Healing: a Systematic Review , 2006, Journal of dental research.

[4]  M Hillery,et al.  Forces Generated in Guide-Wires when Drilling Human Bone , 1995, Proceedings of the Institution of Mechanical Engineers. Part H, Journal of engineering in medicine.

[5]  L. Bancroft,et al.  Imaging characteristics of bone graft materials. , 2006, Radiographics : a review publication of the Radiological Society of North America, Inc.

[6]  R. Pereira,et al.  Comparative study of axial and femoral bone mineral density and parameters of mandibular bone quality in patients receiving dental implants , 2006, Osteoporosis International.

[7]  J. Dowell,et al.  Orthopaedic bone drills-can they be improved? Temperature changes near the drilling face. , 1996, The Journal of bone and joint surgery. British volume.

[8]  R H Ansell,et al.  A study of some factors which affect the strength of screws and their insertion and holding power in bone. , 1968, Journal of biomechanics.

[9]  Haw-Ming Huang,et al.  Factors influencing the resonance frequency of dental implants. , 2003, Journal of oral and maxillofacial surgery : official journal of the American Association of Oral and Maxillofacial Surgeons.

[10]  G. Krekeler,et al.  Atraumatic surgical technique and implant bed preparation. , 1992, Quintessence international.

[11]  J. Hirsch,et al.  Biological factors contributing to failures of osseointegrated oral implants. (II). Etiopathogenesis. , 1998, European journal of oral sciences.

[12]  S. Majumdar Current technologies in the evaluation of bone architecture , 2003, Current osteoporosis reports.

[13]  J. F. Kay,et al.  The tri-spade drill for endosseous dental implant installation. , 1991, The Journal of oral implantology.

[14]  G. Holt,et al.  Osseointegrated Titanium Implants , 1986 .

[15]  F. Savoie,et al.  Intracarpal Soft-Tissue Lesions Associated with an Intra-Articular Fracture of the Distal End of the Radius* , 1996, The Journal of bone and joint surgery. American volume.

[16]  L. Qin,et al.  A study of trabecular bones in ovariectomized goats with micro-computed tomography and peripheral quantitative computed tomography. , 2004, Bone.

[17]  R Jacobs,et al.  Accuracy of computer-aided implant placement. , 2012, Clinical oral implants research.

[18]  W. Bonfield,et al.  The temperature dependence of the deformation of bone. , 1968, Journal of biomechanics.

[19]  Troy MacAvelia,et al.  Force and torque modelling of drilling simulation for orthopaedic surgery , 2014, Computer methods in biomechanics and biomedical engineering.

[20]  J. Hirsch,et al.  Biological factors contributing to failures of osseointegrated oral implants. (I). Success criteria and epidemiology. , 1998, European journal of oral sciences.

[21]  C. Elias,et al.  The effects of superficial roughness and design on the primary stability of dental implants. , 2011, Clinical implant dentistry and related research.

[22]  Z. Lazić,et al.  Evaluation of primary stability of self-tapping and non-self-tapping dental implants. A 12-week clinical study. , 2013, Clinical implant dentistry and related research.

[23]  B. Al-Nawas,et al.  Primary stability of a conical implant and a hybrid, cylindric screw-type implant in vitro. , 2006, The International journal of oral & maxillofacial implants.

[24]  K L Wiggins,et al.  Drilling of bone. , 1976, Journal of biomechanics.

[25]  A. Piattelli,et al.  Correlation between implant stability quotient and bone-implant contact: a retrospective histological and histomorphometrical study of seven titanium implants retrieved from humans. , 2006, Clinical implant dentistry and related research.

[26]  J A Hobkirk,et al.  Investigation of variable factors in drilling bone. , 1977, Journal of oral surgery.

[27]  Han-Joo Lee,et al.  The influence of drill wear on cutting efficiency and heat production during osteotomy preparation for dental implants: a study of drill durability. , 2004, The International journal of oral & maxillofacial implants.

[28]  J A Albright,et al.  Surgical drilling: design and performance of an improved drill. , 1982, Journal of biomechanical engineering.

[29]  E. Hunziker,et al.  Effect of surface topology on the osseointegration of implant materials in trabecular bone. , 1995, Journal of biomedical materials research.

[30]  J. Li Bone-implant interface and remaining tissues on the implant surface after push-out test: an SEM observation. , 1997, Bio-medical materials and engineering.

[31]  Marc Quirynen,et al.  Tolerance within the sleeve inserts of different surgical guides for guided implant surgery. , 2013, Clinical oral implants research.

[32]  J. Gaughan,et al.  Effect of implant design on initial stability of tapered implants. , 2009, The Journal of oral implantology.

[33]  G Watzek,et al.  Effects of drill cooling and bone structure on IMZ implant fixation. , 1993, The International journal of oral & maxillofacial implants.

[34]  Z Majzoub,et al.  Heat generation during implant site preparation: an in vitro study. , 1997, The International journal of oral & maxillofacial implants.

[35]  Murat Akkocaoglu,et al.  Implant design and intraosseous stability of immediately placed implants: a human cadaver study. , 2005, Clinical oral implants research.

[36]  Gabriëlle J. M. Tuijthof,et al.  Influence of tool geometry on drilling performance of cortical and trabecular bone. , 2013, Medical engineering & physics.

[37]  H. Hansson,et al.  Osseointegrated titanium implants. Requirements for ensuring a long-lasting, direct bone-to-implant anchorage in man. , 1981, Acta orthopaedica Scandinavica.

[38]  S. Yerby,et al.  Effect of Cutting Flute Design on Cortical Bone Screw Insertion Torque and Pullout Strength , 2001, Journal of orthopaedic trauma.

[39]  Weiss Cm,et al.  Tissue integration of dental endosseous implants: description and comparative analysis of the fibro-osseous integration and osseous integration systems. , 1986 .

[40]  R Jacobs,et al.  The use of CT scan based planning for oral rehabilitation by means of implants and its transfer to the surgical field: a critical review on accuracy. , 2008, Journal of oral rehabilitation.

[41]  Shu-Wei Wu,et al.  The effects of flute shape and thread profile on the insertion torque and primary stability of dental implants. , 2012, Medical engineering & physics.

[42]  J. Jansen,et al.  In vitro osteogenic potential of bone debris resulting from placement of titanium screw-type implants. , 2008, Clinical oral implants research.

[43]  Rogério Margonar,et al.  Evaluation of immediate bone-cell viability and of drill wear after implant osteotomies: immunohistochemistry and scanning electron microscopy analysis. , 2008, Journal of oral and maxillofacial surgery : official journal of the American Association of Oral and Maxillofacial Surgeons.

[44]  C. Weiss Tissue integration of dental endosseous implants: description and comparative analysis of the fibro-osseous integration and osseous integration systems. , 1986, The Journal of oral implantology.

[45]  Troy MacAvelia,et al.  Biomechanical measurements of surgical drilling force and torque in human versus artificial femurs. , 2012, Journal of biomechanical engineering.

[46]  R. G. Richards,et al.  Animal models for implant biomaterial research in bone: a review. , 2007, European cells & materials.

[47]  G. Helgesson,et al.  Oral rehabilitation of tooth loss: a systematic review of quantitative studies of OHRQoL. , 2012, The International journal of prosthodontics.

[48]  F. Clubb,et al.  Swine as Models in Biomedical Research and Toxicology Testing , 2012, Veterinary pathology.

[49]  B. Al-Nawas,et al.  Primary stability of a hybrid self-tapping implant compared to a cylindrical non-self-tapping implant with respect to drilling protocols in an ex vivo model. , 2011, Clinical implant dentistry and related research.

[50]  I. Tekdemir,et al.  Biomechanical aspects of initial intraosseous stability and implant design: a quantitative micro-morphometric analysis. , 2006, Clinical oral implants research.

[51]  S H Tehemar,et al.  Factors affecting heat generation during implant site preparation: a review of biologic observations and future considerations. , 1999, The International journal of oral & maxillofacial implants.

[52]  Brisman Dl,et al.  The effect of speed, pressure, and time on bone temperature during the drilling of implant sites. , 1996 .

[53]  K. Smetana,et al.  The Miniature Pig as an Animal Model in Biomedical Research , 2005, Annals of the New York Academy of Sciences.

[54]  B. Aksakal,et al.  Influence of orthopaedic drilling parameters on temperature and histopathology of bovine tibia: an in vitro study. , 2011, Medical engineering & physics.

[55]  J B Brunski,et al.  Biomechanical factors affecting the bone-dental implant interface. , 1992, Clinical materials.

[56]  Lars Sennerby,et al.  Influence of implant taper on the primary and secondary stability of osseointegrated titanium implants. , 2004, Clinical oral implants research.

[57]  J. Medige,et al.  Ultrasound velocity and broadband attenuation over a wide range of bone mineral density , 2005, Osteoporosis International.

[58]  P Missika,et al.  Optimal implant stabilization in low density bone. , 2001, Clinical oral implants research.