Cosima – High Resolution Time-of-Flight Secondary Ion Mass Spectrometer for the Analysis of Cometary Dust Particles onboard Rosetta

The ESA mission Rosetta, launched on March 2nd, 2004, carries an instrument suite to the comet 67P/Churyumov-Gerasimenko. The COmetary Secondary Ion Mass Anaylzer – COSIMA – is one of three cometary dust analyzing instruments onboard Rosetta. COSIMA is based on the analytic measurement method of secondary ion mass spectrometry (SIMS). The experiment’s goal is in-situ analysis of the elemental composition (and isotopic composition of key elements) of cometary grains. The chemical characterization will include the main organic components, present homologous and functional groups, as well as the mineralogical and petrographical classification of the inorganic phases. All this analysis is closely related to the chemistry and history of the early solar system. COSIMA covers a mass range from 1 to 3500 amu with a mass resolution m/Δm @ 50% of 2000 at mass 100 amu. Cometary dust is collected on special, metal covered, targets, which are handled by a target manipulation unit. Once exposed to the cometary dust environment, the collected dust grains are located on the target by a microscopic camera. A pulsed primary indium ion beam (among other entities) releases secondary ions from the dust grains. These ions, either positive or negative, are selected and accelerated by electrical fields and travel a well-defined distance through a drift tube and an ion reflector. A microsphere plate with dedicated amplifier is used to detect the ions. The arrival times of the ions are digitized, and the mass spectra of the secondary ions are calculated from these time-of-flight spectra. Through the instrument commissioning, COSIMA took the very first SIMS spectra of the targets in space. COSIMA will be the first instrument applying the SIMS technique in-situ to cometary grain analysis as Rosetta approaches the comet 67P/Churyumov-Gerasimenko, after a long journey of 10 years, in 2014.

[1]  Thomas Stephan,et al.  TOF-SIMS in Cosmochemistry , 2001 .

[2]  Peter F. Arndt,et al.  Properties of Interplanetary Dust: Information from Collected Samples , 2001 .

[3]  F. R. Krueger,et al.  The organic component in dust from comet Halley as measured by the PUMA mass spectrometer on board Vega 1 , 1987, Nature.

[4]  K. Varmuza,et al.  Organic substances in cometary grains: comparison of secondary ion mass spectral data and californium-252 plasma desorption data from reference compounds , 1999 .

[5]  K. Varmuza,et al.  Evaluation of mass spectra from organic compounds assumed to be present in cometary grains. Exploratory data analysis , 2002 .

[6]  T. Stephan,et al.  Correction of dead time effects in time‐of‐flight mass spectrometry , 1994 .

[7]  F. R. Krueger,et al.  Chemometric evaluation of time-of-flight secondary ion mass spectrometry data of minerals in the frame of future in situ analyses of cometary material by COSIMA onboard ROSETTA. , 2006, Rapid communications in mass spectrometry : RCM.

[8]  Andrew Steele,et al.  Comet 81P/Wild 2 Under a Microscope , 2006, Science.

[9]  R. L. Newburn,et al.  Comets in the post-Halley era , 1991 .

[10]  F. R. Krueger,et al.  The organic matter of comet Halley as inferred by joint gas phase and solid phase analyses , 1991 .

[11]  Y. Langevin,et al.  Composition of comet Halley dust particles from Vega observations , 1986 .

[12]  Yu. A. Gur'yan,et al.  Dust in comet P/Halley from Vega observations , 1988 .

[13]  J. Kissel,et al.  CoMA — An advanced space experiment forin situ analysis of cometary matter , 1992 .