Clinical results and pump analysis of the Gyro pump for long-term extracorporeal life support.

Rescuing patients in severe cardiac failure with extracorporeal support remains challenging. The Gyro pump is a centrifugal blood pump and was now used for cardiopulmonary bypass, although it was originally developed for long-term cardiac assist. Little is known about clinical experiences using this pump. Here, we report on the clinical results of long-term extracorporeal life support for over 4 days using the Gyro pump with Excelung, a hollow fiber oxygenator coated with silicone and heparin. Seven patients underwent extracorporeal life support with 15 pump and oxygenator combinations. Gyro and Excelung were used for venoarterial extracorporeal support in six patients and for right ventricular support in one patient. Patient characteristics, pump driving conditions, and blood chemistry were obtained retrospectively. All pumps were subsequently disassembled and examined macroscopically, with 6 of 15 pumps also examined by scanning electron microscopy (SEM). The patient mortality rate was 57.1%. Mean duration of support was 10.5 ± 7.2 days per pump and oxygenator combination. Lactate dehydrogenase and aspartate aminotransferase were generally maintained below 1000 and 100 IU/L, respectively, after the first 4 days of pump driving. Thrombi were found in two pumps, one used without anticoagulation and the other driven at a very slow rotational speed. SEM revealed no wear in the male bearings and very low wear and deformation (0.02 ± 0.03 mm) in the female bearings. The combination of Gyro and Excelung may be applicable for long-term biventricular and right ventricular support, although proper anticoagulation should be administrated to avoid thrombus formation inside the pump.

[1]  A. Agha,et al.  A 5-year experience with cardiopulmonary resuscitation using extracorporeal life support in non-postcardiotomy patients with cardiac arrest. , 2012, Resuscitation.

[2]  A. Haneya,et al.  Successful use of temporary right ventricular support to avoid implantation of biventricular long-term assist device: a transcutaneous approach. , 2011, ASAIO journal.

[3]  Sidney C. Smith,et al.  A 20-year experience with urgent percutaneous cardiopulmonary bypass for salvage of potential survivors of refractory cardiovascular collapse. , 2010, The Journal of thoracic and cardiovascular surgery.

[4]  Y Nosé,et al.  Effects of ultrathin silicone coating of porous membrane on gas transfer and hemolytic performance. , 2008, Artificial organs.

[5]  J. Linneweber,et al.  Evaluation of Floating Impeller Phenomena in a Gyro Centrifugal Pump , 2003, ASAIO journal.

[6]  T. Motomura,et al.  Antithrombogenicity of the Gyro permanently implantable pump with the RPM dynamic suspension system for the impeller. , 2003, Artificial organs.

[7]  J. Linneweber,et al.  Gyro pump wear and deformation analysis in vivo study: creep deformation. , 2000, Artificial organs.

[8]  I. Clarke,et al.  Particles released from the Gyro C1E3 during simulated extracorporeal circulation. , 2000, Artificial organs.

[9]  M. Okada,et al.  Usefulness of postoperative percutaneous cardiopulmonary support using a centrifugal pump: retrospective analysis of complications. , 1999, Artificial organs.

[10]  S. Kashiwazaki,et al.  Clinical evaluation of the Gyro Pump C1E3 as a cardiopulmonary bypass pump. , 1999, ASAIO journal.

[11]  K. Nakata,et al.  Clinical experiences of percutaneous cardiopulmonary support: its effectiveness and limit. , 1998, Artificial organs.

[12]  Y. Shomura,et al.  Clinical Evaluation of a Silicone Coated Hollow Fiber Oxygenator , 1997, ASAIO journal.

[13]  S. Kashiwazaki,et al.  A newly developed silicone-coated membrane oxygenator for long-term cardiopulmonary bypass and cardiac support. , 1997, Artificial organs.

[14]  Y. Ohara,et al.  Hemolytic characteristics of a pivot bearing supported Gyro centrifugal pump (C1E3) simulating various clinical applications. , 1996, Artificial organs.

[15]  Y. Nosé,et al.  Development of an antithrombogenic and antitraumatic blood pump: the Gyro C1E3. , 2000, ASAIO journal.

[16]  J. W. Ball,et al.  Bio-Medicus centrifugal ventricular support for postcardiotomy cardiac failure: a review of 129 cases. , 1996, The Annals of thoracic surgery.

[17]  G. Noon,et al.  Development of a pivot bearing supported sealless centrifugal pump for ventricular assist. , 1996, Artificial Organs.