Development of a sensor for temperature and water concentration in combustion gases using a single tunable diode laser

The water vapour spectrum in the 1–2 µm near-infrared region is systematically analysed to find the best absorption transitions for sensitive measurement of H2O concentration and temperature in combustion environments using a single tunable diode laser with typical distributed feedback single-mode scanning range (1 cm−1). The use of a single laser, even with relatively narrow tuning range, can offer distinct advantages over wavelength-multiplexing techniques. The strategy and spectroscopic criteria for selecting optimum wavelength regions and absorption line combinations are discussed. It should be stressed that no single figure of merit can be derived to simplify the selection process, and the optimum line pair should be chosen case by case. Our investigation reveals that the 1.8 µm spectral region is especially promising, and we have identified 10 of the best water line pairs in this spectral region for temperature measurements in flames. Based on these findings, a pair of H2O transitions near 1.8 µm was targeted for the design and development of an initial single-laser sensor for simultaneously measuring H2O concentration and temperature in atmospheric-pressure flames. As part of the sensor development effort, fundamental spectroscopic parameters including the line strength, line-centre frequency and lower state energies of the probed transitions were measured experimentally to improve the current databases. We conclude with demonstration results in a steady and a forced atmospheric-pressure laboratory combustor.

[1]  R. Hanson,et al.  Diode-Laser Absorption Sensor for Line-of-Sight Gas Temperature Distributions. , 2001, Applied optics.

[2]  R. Hanson,et al.  Diode-laser absorption sensor for measurements in pulse detonation engines , 2000 .

[3]  R. Hanson,et al.  Design of a diode-laser sensor to monitor water vapor in high-pressure combustion gases. , 1997, Applied optics.

[4]  R. Hanson,et al.  Multiplexed diode-laser sensor system for simultaneous H2O, O2, and temperature measurements. , 1994, Optics letters.

[6]  Ronald K. Hanson,et al.  Advanced diode laser absorption sensor for in situ combustion measurements of CO2, H2O, and gas temperature , 1998 .

[7]  Ronald K. Hanson,et al.  Diode-laser measurements of temperature-dependent half-widths of H2O transitions in the 1.4 μm region , 1997 .

[8]  M. Wooldridge,et al.  H2O absorption spectroscopy for determination of temperature and H2O mole fraction in high-temperature particle synthesis systems. , 2002, Applied optics.

[9]  S. R. Drayson Rapid computation of the Voigt profile , 1976 .

[10]  Jian Wang,et al.  In situ combustion measurements of CO, CO2, H2O and temperature using diode laser absorption sensors , 2000 .

[11]  M. P. Arroyo,et al.  Diode-laser absorption technique for simultaneous measurements of multiple gasdynamic parameters in high-speed flows containing water vapor. , 1994, Applied optics.

[12]  M. Allen,et al.  Diode laser absorption sensors for gas-dynamic and combustion flows. , 1998, Measurement science & technology.

[13]  P. R. Bevington,et al.  Data Reduction and Error Analysis for the Physical Sciences , 1969 .

[14]  X Ouyang,et al.  Selection of spectral lines for combustion diagnostics. , 1990, Applied optics.

[15]  Ellis E. Whiting,et al.  An empirical approximation to the Voigt profile , 1968 .

[16]  M. P. Arroyo,et al.  Absorption measurements of water-vapor concentration, temperature, and line-shape parameters using a tunable InGaAsP diode laser. , 1993, Applied optics.

[17]  Laurence S. Rothman,et al.  The HITRAN molecular spectroscopic database and HAWKS (HITRAN atmospheric workstation) , 1998, Defense, Security, and Sensing.

[18]  Laurence S. Rothman,et al.  Reprint of: The HITRAN molecular spectroscopic database and HAWKS (HITRAN Atmospheric Workstation): 1996 edition , 1998 .

[19]  Ronald K. Hanson,et al.  Laser diode wavelength-modulation spectroscopy for simultaneous measurement of temperature, pressure, and velocity in shock-heated oxygen flows. , 1993, Applied optics.

[20]  R. Hanson,et al.  Combustion control and monitoring using a multiplexed diode-laser sensor system , 1996 .

[21]  R. Hanson,et al.  Rapid tuning cw laser technique for measurements of gas velocity, temperature, pressure, density, and mass flux using NO. , 1991, Applied optics.