A quantitative investigation of additive noise reduction for active matrix flat-panel imagers using compensation lines.

A quantitative investigation of a technique for reducing correlated noise in indirect detection active matrix flat-panel imagers has been reported. Correlated noise in such systems arises from the coupling of electronic noise, originating from fluctuations in external sources such as power supplies and ambient electromagnetic sources, to the imaging array via its address lines. The noise reduction technique involves the use of signals from columns of compensation line pixels located in relatively close proximity to the columns of normal imaging pixels on the array. Compensation line pixels are designed to be as sensitive to externally-coupled noise as columns of normal imaging pixels but are insensitive to incident radiation. For each imaging pixel, correlated noise is removed by subtracting from the imaging pixel signal a signal derived from compensation line pixels located on the same row. The effectiveness of various implementations of this correction has been examined through measurements of signal and noise from individual pixels as well as of noise power spectra. These measurements were performed both in the absence of radiation as well as with x rays. The effectiveness of the correction was also demonstrated qualitatively by means of an image of a hand phantom. It was found that the use of a single compensation line dramatically reduces external noise through removal of the correlated noise component. While this form of the correction increases non-radiation-related uncorrelated noise, the effect can be largely reduced through the introduction of multiple compensation lines. Finally, a position-dependent correction based on compensation lines on both sides of the array was found to be effective when the magnitude of the correlated noise varied linearly across the array.

[1]  M. Giger,et al.  Investigation of basic imaging properties in digital radiography. 2. Noise Wiener spectrum. , 1984, Medical physics.

[2]  J Yorkston,et al.  Demonstration of megavoltage and diagnostic x-ray imaging with hydrogenated amorphous silicon arrays. , 1992, Medical physics.

[3]  Weidong Huang,et al.  A data acquisition system for flat-panel imaging arrays , 1993 .

[4]  Norbert Conrads,et al.  Fluoroscopic x-ray imaging with amorphous silicon thin-film arrays , 1994, Medical Imaging.

[5]  M J Yaffe,et al.  Analysis of the spatial-frequency-dependent DQE of optically coupled digital mammography detectors. , 1994, Medical physics.

[6]  J Yorkston,et al.  Empirical investigation of the signal performance of a high-resolution, indirect detection, active matrix flat-panel imager (AMFPI) for fluoroscopic and radiographic operation. , 1997, Medical physics.

[7]  J Yorkston,et al.  Empirical and theoretical investigation of the noise performance of indirect detection, active matrix flat-panel imagers (AMFPIs) for diagnostic radiology. , 1997, Medical physics.

[8]  I. Blevis,et al.  Digital radiology using active matrix readout of amorphous selenium: construction and evaluation of a prototype real-time detector. , 1997, Medical physics.

[9]  J Yorkston,et al.  Initial performance evaluation of an indirect-detection, active matrix flat-panel imager (AMFPI) prototype for megavoltage imaging. , 1998, International journal of radiation oncology, biology, physics.

[10]  Mark A. Hartney,et al.  New amorphous-silicon image sensor for x-ray diagnostic medical imaging applications , 1998, Medical Imaging.

[11]  Isaias D. Job,et al.  Flat-panel imaging system for fluoroscopy applications , 1998, Medical Imaging.

[12]  B. Fallone,et al.  Preliminary study of a metal/a-Se-based portal detector. , 1998, Medical physics.

[13]  P. Munro,et al.  X-ray quantum limited portal imaging using amorphous silicon flat-panel arrays. , 1998, Medical physics.

[14]  Paul R. Granfors Performance characteristics of an amorphous silicon flat-panel x-ray imaging detector , 1999, Medical Imaging.

[15]  Martin Choquette,et al.  Direct conversion detector for digital mammography , 1999, Medical Imaging.

[16]  Isaias D. Job,et al.  Characterization of a third-generation multimode sensor panel , 1999, Medical Imaging.

[17]  Larry E. Antonuk,et al.  Beyond the limits of present active matrix flat-panel imagers (AMFPIs) for diagnostic radiology , 1999, Medical Imaging.

[18]  K. Lam,et al.  Relative dosimetry using active matrix flat-panel imager (AMFPI) technology. , 1999, Medical physics.

[19]  Larry E. Antonuk,et al.  An asynchronous, pipelined, electronic acquisition system for Active Matrix Flat-Panel Imagers (AMFPIs) , 1999 .

[20]  Robert A. Street,et al.  Technology and Applications of Amorphous Silicon , 2000 .

[21]  J H Siewerdsen,et al.  Strategies to improve the signal and noise performance of active matrix, flat-panel imagers for diagnostic x-ray applications. , 2000, Medical physics.

[22]  S Suryanarayanan,et al.  Full breast digital mammography with an amorphous silicon-based flat panel detector: physical characteristics of a clinical prototype. , 2000, Medical physics.

[23]  Larry E. Antonuk,et al.  A programmable, low noise, multichannel ASIC for readout of pixelated amorphous silicon arrays , 2000 .