Investigating possible improvements in image quality with energy-weighting photon-counting breast CT

In an effort to improve the early stage detection and diagnosis of breast cancer, a number of research groups have been investigating the use of x-ray computerized tomography (CT) systems dedicated for use in imaging the breast. For a number of reasons, the performance of energy integrating detectors are sub-optimal for use in CT imaging of the breast. It is expected that the next generation of x-ray detectors for digital radiography and CT will have the capability of counting individually measured photons and recording their energy. In this paper, we used computer simulations to evaluate improvements in image quality that can be attained using energy weighting photon counting detectors for breast CT and a lower kVp settings. Results from this study suggest that improvements in SNR performance can be attained with photon counting detectors as compared to energy integrating detectors.

[1]  J. Boone,et al.  Dedicated breast CT: radiation dose and image quality evaluation. , 2001, Radiology.

[2]  John M Boone,et al.  Experimentally determined spectral optimization for dedicated breast computed tomography. , 2011, Medical physics.

[3]  R. F. Wagner,et al.  SNR and DQE analysis of broad spectrum X-ray imaging , 1985 .

[4]  Samta Thacker,et al.  Evaluating the impact of X-ray spectral shape on image quality in flat-panel CT breast imaging. , 2007, Medical physics.

[5]  P. C. Johns,et al.  X-ray characterisation of normal and neoplastic breast tissues. , 1987, Physics in medicine and biology.

[6]  Willi A Kalender,et al.  Spectral optimization for dedicated breast CT. , 2010, Medical physics.

[7]  Chris C. Shaw,et al.  Cone-beam CT breast imaging with a flat panel detector: a simulation study , 2005, SPIE Medical Imaging.

[8]  Selin Carkaci,et al.  Dedicated cone-beam breast CT: feasibility study with surgical mastectomy specimens. , 2007, AJR. American journal of roentgenology.

[9]  Ehsan Samei,et al.  Design and development of a fully 3D dedicated x-ray computed mammotomography system , 2005, SPIE Medical Imaging.

[10]  Stephen J Glick,et al.  Breast CT. , 2007, Annual review of biomedical engineering.

[11]  Biao Chen,et al.  Cone-beam volume CT breast imaging: feasibility study. , 2002, Medical physics.

[12]  Martin P. Tornai,et al.  Investigation of cone-beam acquisitions implemented using a novel dedicated mammotomography system with unique arbitrary orbit capability (Honorable Mention Poster Award) , 2005, SPIE Medical Imaging.

[13]  L. Feldkamp,et al.  Practical cone-beam algorithm , 1984 .

[14]  M J Yaffe,et al.  The myth of the 50-50 breast. , 2009, Medical physics.

[15]  M Danielsson,et al.  Detective quantum efficiency dependence on x-ray energy weighting in mammography. , 1999, Medical physics.

[16]  P. Shikhaliev Energy-resolved computed tomography: first experimental results , 2008, Physics in medicine and biology.

[17]  Sabee Molloi,et al.  Radiation dose reduction using a CdZnTe-based computed tomography system: comparison to flat-panel detectors. , 2010, Medical physics.

[18]  Ruola Ning,et al.  Cone-beam volume CT mammographic imaging: feasibility study , 2001, SPIE Medical Imaging.

[19]  Aruna A. Vedula,et al.  A computer simulation study comparing lesion detection accuracy with digital mammography, breast tomosynthesis, and cone-beam CT breast imaging. , 2006, Medical physics.

[20]  Taly Gilat Schmidt,et al.  Optimal "image-based" weighting for energy-resolved CT. , 2009, Medical physics.

[21]  Stephen J. Glick,et al.  Computer simulation of CT mammography using a flat-panel imager , 2003, SPIE Medical Imaging.

[22]  Willi A. Kalender X-Ray computed tomography: state of the art , 1993, Other Conferences.