Methodology-Dependent Variations in Reticulocyte Counts Using a Manual and Two Different Flow Cytometric Procedures

This paper describes a comparison between the microscopic brilliant cresyl blue and the flow cytometric thiazol orange (FACScan) and auramine O (Sysmex) methods for enumeration of reticulocytes. The mean intra-assay coefficients of variation for the microscopic, FACScan and Sysmex methods were established to be 33.9 and 5% respectively. A rather poor correlation was observed between the microscopic count and both flow cytometric methods (FACScan r = 0.61; Sysmex r = 0.57). However, the correlation between the flow cytometric methods was satisfactory (r = 0.79). Reference ranges for the reticulocyte count, corresponding with the 2.5th and 97.5th percentile, were determined for the microscopic (8-30/1000), FACScan (11-27/1000) and Sysmex (8-18/1000) methods. Sysmex R-3000 methodology definitely revealed the lowest and narrowest reference range. In conclusion, because of higher reproducibility, flow cytometric analysis of reticulocytes is an attractive alternative procedure for the microscopic enumeration method.

[1]  G. Demetri,et al.  Recombinant human erythropoietin for the treatment of the anaemia associated with autologous bone marrow transplantation , 1994, British journal of haematology.

[2]  C. Brugnara,et al.  Effects of subcutaneous recombinant human erythropoietin in normal subjects: development of decreased reticulocyte hemoglobin content and iron-deficient erythropoiesis. , 1994, The Journal of laboratory and clinical medicine.

[3]  M. Kohnke,et al.  Evaluation of automated reticulocyte counts and their reliability in the presence of Howell-Jolly bodies. , 1994, American journal of clinical pathology.

[4]  J. Phair,et al.  Recombinant human erythropoietin treatment: investigational new drug protocol for the anemia of the acquired immunodeficiency syndrome. Overall results. , 1993, Archives of internal medicine.

[5]  J. Kanold,et al.  Evaluation of erythropoietic/hematopoietic reconstitution after BMT by highly fluorescent reticulocyte counts compares favorably with traditional peripheral blood cell counting. , 1993, Bone marrow transplantation.

[6]  B. Davis,et al.  Flow Cytometric Reticulocyte Analysis and the Reticulocyte Maturity Index , 1993, Annals of the New York Academy of Sciences.

[7]  M. Langweiler,et al.  Proposal for standardization of flow cytometric reticulocyte maturity index (RMI) measurements. , 1993, Cytometry.

[8]  G. Hansson,et al.  Flow cytometric analysis of reticulocytes using an RNA-binding fluorochrome. , 1992, Scandinavian journal of clinical and laboratory investigation.

[9]  H. Lazarus,et al.  Reticulocyte quantification by flow cytometry, image analysis, and manual counting. , 1992, Cytometry.

[10]  D. Bowen,et al.  Comparison of a modified thiazole orange technique with a fully automated analyser for reticulocyte counting. , 1991, Journal of clinical pathology.

[11]  R. Verwilghen,et al.  Reticulocyte count using thiazole orange. A flow cytometry method. , 2008, Clinical and laboratory haematology.

[12]  A. Montloin [Recombinant human erythropoietin]. , 1990, Revue de l'infirmiere.

[13]  P. Nobes,et al.  Reticulocyte counting using flow cytometry. , 1990, Journal of clinical pathology.

[14]  D. Ferguson,et al.  Evaluation of reticulocyte counts by flow cytometry in a routine laboratory , 1990, American journal of hematology.

[15]  A. Gratwohl,et al.  Evaluation of the Sysmex R-1000. An automated reticulocyte analyzer. , 1990, American journal of clinical pathology.

[16]  R. Koene,et al.  [Recombinant human erythropoietin]. , 1988, Nederlands tijdschrift voor geneeskunde.

[17]  L. G. Lee,et al.  Thiazole orange: a new dye for reticulocyte analysis. , 1986, Cytometry.

[18]  N Gochman,et al.  Incorrect least-squares regression coefficients in method-comparison analysis. , 1979, Clinical chemistry.