Mechanics of bone/PMMA composite structures: an in vitro study of human vertebrae.

The goal of this study was to provide material property data for the cement/bone composite resulting from the introduction of PMMA bone cement into human vertebral bodies. A series of quasistatic tensile and compressive mechanical tests were conducted using cement/bone composite structures machined from cement-infiltrated vertebral bodies. Experiments were performed both at room temperature and at body temperature. We found that the modulus of the composite structures was lower than bulk cement (p<0.0001). For compression at 37( composite function)C: composite =2.3+/-0.5GPa, cement =3.1+/-0.2GPa; at 23( composite function)C: composite =3.0+/-0.3GPa, cement =3.4+/-0.2GPa. Specimens tested at room temperature were stiffer than those tested at body temperature (p=0.0004). Yield and ultimate strength factors for the composite were all diminished (55-87%) when compared to cement properties. In general, computational models have assumed that cement/bone composite had the same modulus as cement. The results of this study suggest that computational models of cement infiltrated vertebrae and cemented arthroplasties could be improved by specifying different material properties for cement and cement/bone composite.

[1]  G. Lewis,et al.  Properties of acrylic bone cement: state of the art review. , 1997, Journal of biomedical materials research.

[2]  Y. Hochberg A sharper Bonferroni procedure for multiple tests of significance , 1988 .

[3]  A. Levine,et al.  Biomechanical efficacy of unipedicular versus bipedicular vertebroplasty for the management of osteoporotic compression fractures. , 1999, Spine.

[4]  W. Hayes,et al.  Compressive behavior of human bone-cement composites. , 1991, The Journal of arthroplasty.

[5]  T. Keaveny,et al.  Yield strain behavior of trabecular bone. , 1998, Journal of biomechanics.

[6]  Raj D. Rao,et al.  Painful osteoporotic vertebral fracture. Pathogenesis, evaluation, and roles of vertebroplasty and kyphoplasty in its management. , 2003, The Journal of bone and joint surgery. American volume.

[7]  G Van der Perre,et al.  Structural and material mechanical properties of human vertebral cancellous bone. , 1997, Medical engineering & physics.

[8]  D F Kallmes,et al.  Percutaneous polymethylmethacrylate vertebroplasty in the treatment of osteoporotic vertebral body compression fractures: technical aspects. , 1997, AJNR. American journal of neuroradiology.

[9]  T. Keaveny,et al.  Dependence of yield strain of human trabecular bone on anatomic site. , 2001, Journal of biomechanics.

[10]  Christoph Fankhauser,et al.  Augmentation of mechanical properties in osteoporotic vertebral bones – a biomechanical investigation of vertebroplasty efficacy with different bone cements , 2001, European Spine Journal.

[11]  W. J. Johnson,et al.  Elastic constants of composites formed from PMMA bone cement and anisotropic bovine tibial cancellous bone. , 1989, Journal of biomechanics.

[12]  S. Belkoff,et al.  An in vitro biomechanical evaluation of bone cements used in percutaneous vertebroplasty. , 1999, Bone.

[13]  S. Cowin Bone mechanics handbook , 2001 .