REOPERATION FOR SUBOPTIMAL OUTCOMES AFTER DEEP BRAIN STIMULATION SURGERY

OBJECTIVETo examine a case series of reoperations for deep brain stimulation (DBS) leads in which clinical scenarios revealed suboptimal outcome from a previous operation. Suboptimally placed DBS leads are one potential reason for unsatisfactory results after surgery for Parkinson's disease (PD), essential tremor (ET), or dystonia. In a previous study of patients who experienced suboptimal results, 19 of 41 patients had misplaced leads. Similarly, another report commented that lead placement beyond a 2- to 3-mm window resulted in inadequate clinical benefit, and, in 1 patient, revision improved outcome. The goal of the current study was to perform an unblinded retrospective chart review of DBS patients with unsatisfactory outcomes who presented for reoperation. METHODSPatients who had DBS lead replacements after reoperation were assessed with the use of a retrospective review of an institutional review board-approved movement disorders database. Cases of reoperation for suboptimal clinical benefit were included, and cases of replacement of DBS leads caused by infection or hardware malfunction were excluded. Data points studied included age, disease duration, diagnosis, motor outcomes (the Unified Parkinson Disease Rating Scale III in PD, the Tremor Rating Scale in ET, and the Unified Dystonia Rating Scale in dystonia), quality of life (Parkinson's Disease Questionnaire-39 in PD), and the Clinician Global Impression scale. The data from before and after reoperation were examined to determine the estimated impact of repeat surgery. RESULTSThere were 11 patients with PD, 7 with ET, and 4 with dystonia. The average age of the PD group was 52 years, the disease duration was 10 years, and the average vector distance of the location of the active DBS contact was adjusted 5.5 mm. Six patients (54%) with PD had preoperative off medication on DBS Unified Parkinson Disease Rating Scale scores that could be compared with postoperative off medication on DBS scores. The average improvement across this group of patients was 24.4%. The Parkinson's Disease Questionnaire-39 improved in the areas of mobility (28.18), activities of daily living (14.77), emotion (14.72), stigma (17.61), and discomfort (17.42). The average age of the ET group was 66 years, the disease duration was 29 years, and the average adjusted distance was 6.1 mm. Five ET patients (83.3%) in the cohort had a prereplacement on DBS Tremor Rating Scale and a postreplacement on DBS Tremor Rating Scale with the average improvement of 60.4%. The average age of the dystonia group was 39 years, the average disease duration was 7 years, and the average adjusted lead distance was 6.7 mm. Three patients (75%) with dystonia had prereplacement on DBS Unified Dystonia Rating Scale and postreplacement on DBS Unified Dystonia Rating Scale scores. Across these 3 dystonia patients, the improvement was 12.8%. Clinician Global Impression scale scores (1, very much improved; 2, much improved; 3, minimally improved; 4, no change; 5, minimally worse; 6, much worse; 7, very much worse) after replacement revealed the following results in patients with PD: 1, 7 patients; 2, 3 patients; 3, 1 patient); with ET (1, 4 patients; 2, 3 patients); and with dystonia (1, 1 patient; 2, 2 patients; 3, 1 patient). The latency from original lead placement to reoperation (repositioning/revision) overall was 28.9 months (range, 2–104 mo); however, in leads referred from outside institutions (n = 11 patients), this latency was 48 months (range, 12–104 mo) compared with leads implanted by surgeons from the University of Florida (n = 11 patients), which was 9.7 months (range, 2–19 mo). The most common clinical history was failure to achieve a perceived outcome; however, history of an asymmetric benefit was present in 4 (18.2%) of 22 patients, and lead migration was present in 3 (13.6%) of 22 patients. CONCLUSIONThere are many potential causes of suboptimal benefit after DBS. Timely identification of suboptimal lead placements followed by reoperation and repositioning/replacement in a subset of patients may improve outcomes.

[1]  M. Okun,et al.  Lessons Learned in Deep Brain Stimulation for Movement and Neuropsychiatric Disorders , 2006, CNS Spectrums.

[2]  K. Lyons,et al.  Long term safety and efficacy of unilateral deep brain stimulation of the thalamus for parkinsonian tremor , 2001, Journal of neurology, neurosurgery, and psychiatry.

[3]  Michael S Okun,et al.  Lesion therapy for Parkinson's disease and other movement disorders: Update and controversies , 2004, Movement disorders : official journal of the Movement Disorder Society.

[4]  Robert E. Gross,et al.  Assessment of Brain Shift Related to Deep Brain Stimulation Surgery , 2007, Stereotactic and Functional Neurosurgery.

[5]  Y. Ben-Shlomo,et al.  Stimulation of the caudal zona incerta is superior to stimulation of the subthalamic nucleus in improving contralateral parkinsonism. , 2006, Brain : a journal of neurology.

[6]  Y. Hsin,et al.  Surgical and hardware complications in subthalamic nucleus deep brain stimulation , 2007, Journal of Clinical Neuroscience.

[7]  Clement Hamani,et al.  Hardware-Related Complications of Deep Brain Stimulation: A Review of the Published Literature , 2006, Stereotactic and Functional Neurosurgery.

[8]  Jens Volkmann,et al.  Update on surgery for Parkinson's disease , 2007, Current opinion in neurology.

[9]  P. Starr,et al.  Implantation of deep brain stimulators into the subthalamic nucleus: technical approach and magnetic resonance imaging-verified lead locations. , 2002, Journal of neurosurgery.

[10]  G. Deuschl,et al.  Subthalamic nucleus deep brain stimulation: Summary and meta‐analysis of outcomes , 2006, Movement disorders : official journal of the Movement Disorder Society.

[11]  Yunling E. Du,et al.  Subthalamic stimulation for Parkinson disease: determination of electrode location necessary for clinical efficacy. , 2005, Neurosurgical focus.

[12]  M. Hariz What is deep brain stimulation "failure" and how do we manage our own failures? , 2005, Archives of neurology.

[13]  M. Okun,et al.  Identifying candidates for deep brain stimulation in Parkinson's disease: the role of the primary care physician. , 2007, Geriatrics.

[14]  D. Bowers,et al.  Deep Brain Stimulation and the Role of the Neuropsychologist , 2007, The Clinical neuropsychologist.

[15]  J Meixensberger,et al.  The first evaluation of brain shift during functional neurosurgery by deformation field analysis , 2005, Journal of Neurology, Neurosurgery & Psychiatry.

[16]  R. Uitti,et al.  Deep brain stimulation lead fixation: a comparative study of the Navigus and Medtronic burr hole fixation device , 2005, Clinical Neurology and Neurosurgery.

[17]  S. Gill,et al.  MAGNETIC RESONANCE IMAGING‐DIRECTED METHOD FOR FUNCTIONAL NEUROSURGERY USING IMPLANTABLE GUIDE TUBES , 2007, Neurosurgery.

[18]  G. Deuschl,et al.  Thirty days complication rate following surgery performed for deep‐brain‐stimulation , 2007, Movement disorders : official journal of the Movement Disorder Society.

[19]  Philip A. Starr,et al.  Placement of Deep Brain Stimulators into the Subthalamic Nucleus or Globus pallidus internus: Technical Approach , 2003, Stereotactic and Functional Neurosurgery.

[20]  Elena Moro,et al.  Criteria for deep-brain stimulation in Parkinson’s disease: review and analysis , 2006, Expert review of neurotherapeutics.

[21]  G. Friehs,et al.  Development and initial validation of a screening tool for Parkinson disease surgical candidates , 2004, Neurology.

[22]  M. Hariz,et al.  Do Microelectrode Techniques Increase Accuracy or Decrease Risks in Pallidotomy and Deep Brain Stimulation? , 2000, Stereotactic and Functional Neurosurgery.

[23]  Tarek A. Yousry,et al.  A Quick and Universal Method for Stereotactic Visualization of the Subthalamic Nucleus before and after Implantation of Deep Brain Stimulation Electrodes , 2004, Stereotactic and Functional Neurosurgery.

[24]  Hakan Widner,et al.  Deep brain stimulation for Parkinson's disease: Patient selection and evaluation , 2002, Movement disorders : official journal of the Movement Disorder Society.

[25]  Alastair J. Martin,et al.  Magnetic Resonance Imaging of Implanted Deep Brain Stimulators: Experience in a Large Series , 2007, Stereotactic and Functional Neurosurgery.

[26]  Anthony Davidson,et al.  Short-term and long-term safety of deep brain stimulation in the treatment of movement disorders. , 2007, Journal of neurosurgery.

[27]  M. Okun,et al.  Management of referred deep brain stimulation failures: a retrospective analysis from 2 movement disorders centers. , 2005, Archives of neurology.

[28]  Xuguang Liu,et al.  Operative and hardware complications of deep brain stimulation for movement disorders , 2006, British journal of neurosurgery.

[29]  Jerrold L Vitek,et al.  Patient selection and assessment recommendations for deep brain stimulation in Tourette syndrome , 2006, Movement disorders : official journal of the Movement Disorder Society.

[30]  M. Álvarez-Vega,et al.  Complications in subthalamic nucleus stimulation surgery for treatment of Parkinson’s disease. Review of 272 procedures , 2007, Acta Neurochirurgica.

[31]  T. Aziz,et al.  Increased risk of lead fracture and migration in dystonia compared with other movement disorders following deep brain stimulation , 2004, Journal of Clinical Neuroscience.

[32]  Alon Mogilner,et al.  Complications of Deep Brain Stimulation Surgery , 2002, Stereotactic and Functional Neurosurgery.

[33]  Rajesh Pahwa,et al.  Surgical and hardware complications of subthalamic stimulation , 2004, Neurology.

[34]  Marwan I. Hariz,et al.  Are Complications Less Common in Deep Brain Stimulation than in Ablative Procedures for Movement Disorders? , 2004, Stereotactic and Functional Neurosurgery.