Calculation method based on a BGO+SiPM detector for ^13N gas coincidence detection efficiency

To reduce the lower detection limit of 13N gas, we design a γ-γ coincidence measuring instrument including two detectors, a sampling vessel, and a summation and coincidence circuit in this work. The detector uses a 13 cm × 5 cm oversized Bi4Ge3O12 (BGO) scintillator and 172 silicon photomultiplier (SiPM) tubes, which greatly reduce the instrument size while improving the coincidence detection efficiency. The 13N gas coincidence detection efficiency is improved by installing a multilayer metal absorber plate inside the cylindrical sampling vessel. Due to the short half-life of 13N gas, it cannot be stored for a long time; additionally, it is difficult to obtain. It is not possible to directly scale the coincidence detection efficiency of this instrument using a 13N gas source with known activity in engineering projects. Based on the spatial distribution of the relative efficiency of the sampling vessel and the absolute efficiency of the reference point, we use a combination of 22Na solid point source experiments and Monte Carlo simulations to calculate the coincidence detection efficiency of this instrument for 13N gas in this work; the coincidence detection efficiency is approximately 4%, which meets the engineering design requirements.

[1]  M. Kitaura,et al.  Characterization of imperfections in scintillator crystals using gamma-ray induced positron annihilation lifetime spectroscopy , 2022, Optical Materials: X.

[2]  P. Nambissan,et al.  Positron annihilation spectroscopy for defect characterization in nanomaterials , 2022, Design, Fabrication, and Characterization of Multifunctional Nanomaterials.

[3]  Design, Fabrication, and Characterization of Multifunctional Nanomaterials , 2022 .

[4]  A. Muñoz,et al.  HIGH-HEAT FLUX Cu-0.8Y ALLOYS INVESTIGATED BY POSITRON ANNIHILATION SPECTROSCOPY , 2021, Journal of Alloys and Compounds.

[5]  K. Yamaoka,et al.  Silicon photomultiplier (Si-PM) comparisons for low-energy gamma ray readouts with BGO and CsI (Tl) scintillators , 2020 .

[6]  J. Benlloch,et al.  In-depth evaluation of TOF-PET detectors based on crystal arrays and the TOFPET2 ASIC , 2020 .

[7]  E. Auffray,et al.  Pushing Cherenkov PET with BGO via coincidence time resolution classification and correction , 2020, Physics in medicine and biology.

[8]  M. Moszynski,et al.  Silicon photomultipliers in gamma spectroscopy with scintillators , 2019, Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment.

[9]  K. Ungar,et al.  The development of a digital gamma-gamma coincidence/anticoincidence spectrometer and its applications to monitor low-level atmospheric 22Na/7Be activity ratios in Resolute Bay, Canada. , 2018, Journal of environmental radioactivity.

[10]  D. Liang,et al.  Performance of a SiPM based semi-monolithic scintillator PET detector , 2017, Physics in medicine and biology.

[11]  M. Grodzicka-Kobylka,et al.  Comparison of SensL and Hamamatsu 4×4 channel SiPM arrays in gamma spectrometry with scintillators , 2017 .

[12]  R. Zhou,et al.  A PET detector module with monolithic crystal, single end readout, SiPM array and high depth-of-interaction resolution , 2016 .

[13]  Chen Zeng,et al.  Feasibility Study on Silicon Photomultiplier With Epitaxial Quenching Resistors as the Readout for PET Detectors , 2016, IEEE Transactions on Nuclear Science.

[14]  K. Joo,et al.  Performance characteristics of a silicon photomultiplier based compact radiation detector for Homeland Security applications , 2015 .

[15]  P. Mekarski,et al.  A gamma-gamma coincidence/anticoincidence spectrometer for low-level cosmogenic (22)Na/(7)Be activity ratio measurement. , 2014, Journal of environmental radioactivity.

[16]  J. Saegusa CREPT-MCNP code for efficiency calibration of HPGe detectors with the representative point method. , 2008, Applied radiation and isotopes : including data, instrumentation and methods for use in agriculture, industry and medicine.