A gamma camera count rate saturation correction method for whole-body planar imaging
暂无分享,去创建一个
Eric C Frey | George Sgouros | Robert F Hobbs | Sébastien Baechler | Srinivasan Senthamizhchelvan | Andrew R Prideaux | Caroline E Esaias | Melvin Reinhardt | David M Loeb | E. Frey | G. Sgouros | S. Senthamizhchelvan | S. Baechler | A. Prideaux | R. Hobbs | Caroline E. Esaias | D. Loeb | Melvin J Reinhardt
[1] Manuel Bardiès,et al. Comparison of four scatter correction methods for patient whole‐body imaging during therapeutic trials with iodine‐131 , 2002, Cancer.
[2] J. Eary,et al. Preliminary validation of the opposing view method for quantitative gamma camera imaging. , 1989, Medical physics.
[3] D. Loeb,et al. Dose-finding study of 153Sm-EDTMP in patients with poor prognosis osteosarcoma , 2008 .
[4] I Buvat,et al. Comparative assessment of nine scatter correction methods based on spectral analysis using Monte Carlo simulations. , 1995, Journal of nuclear medicine : official publication, Society of Nuclear Medicine.
[5] R. Jaszczak,et al. Improved SPECT quantification using compensation for scattered photons. , 1984, Journal of nuclear medicine : official publication, Society of Nuclear Medicine.
[6] Michael Ljungberg,et al. An activity quantification method based on registration of CT and whole-body scintillation camera images, with application to 131I. , 2002, Journal of nuclear medicine : official publication, Society of Nuclear Medicine.
[7] J A Sorenson. Deadtime characteristics of Anger cameras. , 1975, Journal of nuclear medicine : official publication, Society of Nuclear Medicine.
[8] Michael Lassmann,et al. The impact of PET and SPECT on dosimetry for targeted radionuclide therapy. , 2006, Zeitschrift fur medizinische Physik.
[9] C. Mulhern,et al. External imaging techniques for quantitation of distribution of I-131 F(ab')2 fragments of monoclonal antibody in humans. , 1984, Medical physics.
[10] J G Kereiakes,et al. In vivo quantitation of lesion radioactivity using external counting methods. , 1976, Medical physics.
[11] J A Siegel,et al. The buildup factor: effect of scatter on absolute volume determination. , 1985, Journal of nuclear medicine : official publication, Society of Nuclear Medicine.
[12] M Bardiès,et al. Correction of count losses due to deadtime on a DST-XLi (SmVi-GE) camera during dosimetric studies in patients injected with iodine-131. , 2002, Physics in medicine and biology.
[13] George Sgouros,et al. Three-dimensional imaging-based radiobiological dosimetry. , 2008, Seminars in nuclear medicine.
[14] Manuel Bardiès,et al. Impact of scatter and attenuation corrections for iodine-131 two-dimensional quantitative imaging in patients. , 2003, Cancer biotherapy & radiopharmaceuticals.
[15] Yong Du,et al. Comparison of Residence Time Estimation Methods for Radioimmunotherapy Dosimetry and Treatment Planning—Monte Carlo Simulation Studies , 2008, IEEE Transactions on Medical Imaging.
[16] K F Koral,et al. Dead time of an anger camera in dual-energy-window-acquisition mode. , 1993, Medical physics.
[17] E Bombardieri,et al. A practical dead time correction method in planar activity quantification for dosimetry during radionuclide therapy. , 2009, The quarterly journal of nuclear medicine and molecular imaging : official publication of the Italian Association of Nuclear Medicine (AIMN) [and] the International Association of Radiopharmacology (IAR), [and] Section of the Society of....
[18] M R Castellani,et al. Individualized dosimetry in the management of metastatic differentiated thyroid cancer. , 2009, The quarterly journal of nuclear medicine and molecular imaging : official publication of the Italian Association of Nuclear Medicine (AIMN) [and] the International Association of Radiopharmacology (IAR), [and] Section of the Society of....
[19] M Bardiès,et al. Bispecific antibody and iodine-131-labeled bivalent hapten dosimetry in patients with medullary thyroid or small-cell lung cancer. , 1996, Journal of nuclear medicine : official publication, Society of Nuclear Medicine.
[20] R Adams,et al. Testing the count rate performance of the scintillation camera by exponential attenuation: decaying source; multiple filters. , 1988, Medical physics.
[21] D. Podoloff,et al. Improved conjugate view quantitation of I-131 by subtraction of scatter and septal penetration events with a triple energy window method. , 1995, Medical physics.
[22] B. Wessels,et al. MIRD pamphlet no. 16: Techniques for quantitative radiopharmaceutical biodistribution data acquisition and analysis for use in human radiation dose estimates. , 1999, Journal of nuclear medicine : official publication, Society of Nuclear Medicine.
[23] G T Chen,et al. An overview of imaging techniques and physical aspects of treatment planning in radioimmunotherapy. , 1993, Medical physics.
[24] Michael Ljungberg,et al. Evaluation of accuracy in activity calculations for the conjugate view method from Monte Carlo simulated scintillation camera images using experimental data in an anthropomorphic phantom. , 2005, Journal of nuclear medicine : official publication, Society of Nuclear Medicine.
[25] Michael Ljungberg,et al. The LundADose method for planar image activity quantification and absorbed-dose assessment in radionuclide therapy. , 2005, Cancer biotherapy & radiopharmaceuticals.
[26] Yuni Dewaraja,et al. Update on hybrid conjugate-view SPECT tumor dosimetry and response in 131I-tositumomab therapy of previously untreated lymphoma patients. , 2003, Journal of nuclear medicine : official publication, Society of Nuclear Medicine.
[27] J A Sorenson,et al. Methods of correcting Anger camera deadtime losses. , 1976, Journal of nuclear medicine : official publication, Society of Nuclear Medicine.