A gamma camera count rate saturation correction method for whole-body planar imaging

Whole-body (WB) planar imaging has long been one of the staple methods of dosimetry, and its quantification has been formalized by the MIRD Committee in pamphlet no 16. One of the issues not specifically addressed in the formalism occurs when the count rates reaching the detector are sufficiently high to result in camera count saturation. Camera dead-time effects have been extensively studied, but all of the developed correction methods assume static acquisitions. However, during WB planar (sweep) imaging, a variable amount of imaged activity exists in the detector's field of view as a function of time and therefore the camera saturation is time dependent. A new time-dependent algorithm was developed to correct for dead-time effects during WB planar acquisitions that accounts for relative motion between detector heads and imaged object. Static camera dead-time parameters were acquired by imaging decaying activity in a phantom and obtaining a saturation curve. Using these parameters, an iterative algorithm akin to Newton's method was developed, which takes into account the variable count rate seen by the detector as a function of time. The algorithm was tested on simulated data as well as on a whole-body scan of high activity Samarium-153 in an ellipsoid phantom. A complete set of parameters from unsaturated phantom data necessary for count rate to activity conversion was also obtained, including build-up and attenuation coefficients, in order to convert corrected count rate values to activity. The algorithm proved successful in accounting for motion- and time-dependent saturation effects in both the simulated and measured data and converged to any desired degree of precision. The clearance half-life calculated from the ellipsoid phantom data was calculated to be 45.1 h after dead-time correction and 51.4 h with no correction; the physical decay half-life of Samarium-153 is 46.3 h. Accurate WB planar dosimetry of high activities relies on successfully compensating for camera saturation which takes into account the variable activity in the field of view, i.e. time-dependent dead-time effects. The algorithm presented here accomplishes this task.

[1]  Manuel Bardiès,et al.  Comparison of four scatter correction methods for patient whole‐body imaging during therapeutic trials with iodine‐131 , 2002, Cancer.

[2]  J. Eary,et al.  Preliminary validation of the opposing view method for quantitative gamma camera imaging. , 1989, Medical physics.

[3]  D. Loeb,et al.  Dose-finding study of 153Sm-EDTMP in patients with poor prognosis osteosarcoma , 2008 .

[4]  I Buvat,et al.  Comparative assessment of nine scatter correction methods based on spectral analysis using Monte Carlo simulations. , 1995, Journal of nuclear medicine : official publication, Society of Nuclear Medicine.

[5]  R. Jaszczak,et al.  Improved SPECT quantification using compensation for scattered photons. , 1984, Journal of nuclear medicine : official publication, Society of Nuclear Medicine.

[6]  Michael Ljungberg,et al.  An activity quantification method based on registration of CT and whole-body scintillation camera images, with application to 131I. , 2002, Journal of nuclear medicine : official publication, Society of Nuclear Medicine.

[7]  J A Sorenson Deadtime characteristics of Anger cameras. , 1975, Journal of nuclear medicine : official publication, Society of Nuclear Medicine.

[8]  Michael Lassmann,et al.  The impact of PET and SPECT on dosimetry for targeted radionuclide therapy. , 2006, Zeitschrift fur medizinische Physik.

[9]  C. Mulhern,et al.  External imaging techniques for quantitation of distribution of I-131 F(ab')2 fragments of monoclonal antibody in humans. , 1984, Medical physics.

[10]  J G Kereiakes,et al.  In vivo quantitation of lesion radioactivity using external counting methods. , 1976, Medical physics.

[11]  J A Siegel,et al.  The buildup factor: effect of scatter on absolute volume determination. , 1985, Journal of nuclear medicine : official publication, Society of Nuclear Medicine.

[12]  M Bardiès,et al.  Correction of count losses due to deadtime on a DST-XLi (SmVi-GE) camera during dosimetric studies in patients injected with iodine-131. , 2002, Physics in medicine and biology.

[13]  George Sgouros,et al.  Three-dimensional imaging-based radiobiological dosimetry. , 2008, Seminars in nuclear medicine.

[14]  Manuel Bardiès,et al.  Impact of scatter and attenuation corrections for iodine-131 two-dimensional quantitative imaging in patients. , 2003, Cancer biotherapy & radiopharmaceuticals.

[15]  Yong Du,et al.  Comparison of Residence Time Estimation Methods for Radioimmunotherapy Dosimetry and Treatment Planning—Monte Carlo Simulation Studies , 2008, IEEE Transactions on Medical Imaging.

[16]  K F Koral,et al.  Dead time of an anger camera in dual-energy-window-acquisition mode. , 1993, Medical physics.

[17]  E Bombardieri,et al.  A practical dead time correction method in planar activity quantification for dosimetry during radionuclide therapy. , 2009, The quarterly journal of nuclear medicine and molecular imaging : official publication of the Italian Association of Nuclear Medicine (AIMN) [and] the International Association of Radiopharmacology (IAR), [and] Section of the Society of....

[18]  M R Castellani,et al.  Individualized dosimetry in the management of metastatic differentiated thyroid cancer. , 2009, The quarterly journal of nuclear medicine and molecular imaging : official publication of the Italian Association of Nuclear Medicine (AIMN) [and] the International Association of Radiopharmacology (IAR), [and] Section of the Society of....

[19]  M Bardiès,et al.  Bispecific antibody and iodine-131-labeled bivalent hapten dosimetry in patients with medullary thyroid or small-cell lung cancer. , 1996, Journal of nuclear medicine : official publication, Society of Nuclear Medicine.

[20]  R Adams,et al.  Testing the count rate performance of the scintillation camera by exponential attenuation: decaying source; multiple filters. , 1988, Medical physics.

[21]  D. Podoloff,et al.  Improved conjugate view quantitation of I-131 by subtraction of scatter and septal penetration events with a triple energy window method. , 1995, Medical physics.

[22]  B. Wessels,et al.  MIRD pamphlet no. 16: Techniques for quantitative radiopharmaceutical biodistribution data acquisition and analysis for use in human radiation dose estimates. , 1999, Journal of nuclear medicine : official publication, Society of Nuclear Medicine.

[23]  G T Chen,et al.  An overview of imaging techniques and physical aspects of treatment planning in radioimmunotherapy. , 1993, Medical physics.

[24]  Michael Ljungberg,et al.  Evaluation of accuracy in activity calculations for the conjugate view method from Monte Carlo simulated scintillation camera images using experimental data in an anthropomorphic phantom. , 2005, Journal of nuclear medicine : official publication, Society of Nuclear Medicine.

[25]  Michael Ljungberg,et al.  The LundADose method for planar image activity quantification and absorbed-dose assessment in radionuclide therapy. , 2005, Cancer biotherapy & radiopharmaceuticals.

[26]  Yuni Dewaraja,et al.  Update on hybrid conjugate-view SPECT tumor dosimetry and response in 131I-tositumomab therapy of previously untreated lymphoma patients. , 2003, Journal of nuclear medicine : official publication, Society of Nuclear Medicine.

[27]  J A Sorenson,et al.  Methods of correcting Anger camera deadtime losses. , 1976, Journal of nuclear medicine : official publication, Society of Nuclear Medicine.