Experimental and Theoretical Evaluation of a High Resolution CMOS Based Detector Under X-Ray Imaging Conditions

Fundamental imaging performance in terms of Modulation Transfer Function (MTF), Noise Power Spectrum (NPS) and Detective Quantum Efficiency (DQE) was investigated for a high resolution CMOS based imaging sensor. The device consists of a 33.91 mg/cm2 Gd2O2S:Tb scintillator screen, placed in direct contact with a CMOS photodiode array. The CMOS photodiode array, featuring 1200×1600 pixels with a pixel pitch of 22.5 μm, was used as an optical photon detector. In addition to the conventional frequency dependent parameters characterizing image quality, image information content was assessed through the application of information capacity (IC). The MTF was measured using the slanted-edge method to avoid aliasing while the Normalized NPS (NNPS) was determined by two-dimensional (2D) Fourier transforming of uniformly exposed images. Both measurements were performed under the representative radiation quality (RQA) settings, RQA-5 (70 kVp digital-radiography) and RQA-M2 (28 kVp digital-mammography) recommended by the International Electrotechnical Commission Reports 62220-1 and 62220-1-2 respectively. The DQE was assessed from the measured MTF, NPS and the direct entrance surface air-Kerma (ESAK) obtained from X-ray spectra measurement with a portable cadmium telluride (CdTe) detector. The ESAK values ranged between 11-87 μGy for RQA-5 and 6-40 μGy for RQA-M2. Additionally the output electrons per X-ray photon of the detector and its signal transfer characteristics were assessed via an analytical model, within the framework of the linear cascaded systems (LCS) theory. It was found that the detector response function was linear for the exposure ranges under investigation. Additionally our results showed that for the same RQA quality the output electrons per X-ray photon, as well as the measured and analytically predicted MTF, were not significantly affected by ESAK. MTF and DQE where found better compared to previously published data for other CCD and CMOS sensors, while the NNPS appeared to be comparable in the frequency range under investigation (0-10 cycles/mm).

[1]  Gyuseong Cho,et al.  Development and evaluation of a digital radiographic system based on CMOS image sensor , 2000, 2000 IEEE Nuclear Science Symposium. Conference Record (Cat. No.00CH37149).

[2]  J. R Greening,et al.  Fundamentals of Radiation Dosimetry , 1981 .

[3]  J M Boone,et al.  Determination of the presampled MTF in computed tomography. , 2001, Medical physics.

[4]  Pierre Magnan,et al.  Fast MTF measurement of CMOS imagers at the chip level using ISO 12233 slanted-edge methodology , 2004, SPIE Remote Sensing.

[5]  R. Speller,et al.  Modeling the imaging performance of prototype organic x-ray imagers. , 2007, Medical physics.

[6]  Qihua Zhao,et al.  Performance of a high fill factor, indirect detection prototype flat-panel imager for mammography. , 2007, Medical physics.

[7]  R. F. Wagner,et al.  Application of information theory to the assessment of computed tomography. , 1979, Medical physics.

[8]  James T. Dobbins Image Quality Metrics for Digital Systems , 2000 .

[9]  Satoshi Miyajima,et al.  CdZnTe detector in diagnostic x-ray spectroscopy. , 2002, Medical physics.

[10]  Steven K. Onishi,et al.  High-resolution imager for digital mammography: physical characterization of a prototype sensor , 2005, Physics in medicine and biology.

[11]  M. Danielsson,et al.  Photon-counting spectral computed tomography using silicon strip detectors: a feasibility study , 2010, Physics in medicine and biology.

[12]  R. K. Swank,et al.  Calculation of modulation transfer functions of x-ray fluorescent screens. , 1973, Applied optics.

[13]  Bimal P. Mathur,et al.  Analytical charge collection and MTF model for photodiode-based CMOS imagers , 2002 .

[14]  Joaquim Salvi,et al.  Review of CMOS image sensors , 2006, Microelectron. J..

[15]  I. Kandarakis,et al.  Modelling the imaging performance and low contrast detectability in digital mammography , 2009 .

[16]  Ioannis S Kandarakis,et al.  Modeling granular phosphor screens by Monte Carlo methods. , 2006, Medical physics.

[17]  E. Samei,et al.  A method for measuring the presampled MTF of digital radiographic systems using an edge test device. , 1998, Medical physics.

[18]  Qihua Zhao,et al.  System performance of a prototype flat-panel imager operated under mammographic conditions. , 2003, Medical physics.

[19]  N W Marshall,et al.  A comparison between objective and subjective image quality measurements for a full field digital mammography system , 2006, Physics in medicine and biology.

[20]  I. Kandarakis,et al.  Evaluating x-ray detectors for radiographic applications: A comparison of ZnSCdS:Ag with and screens , 1997 .

[21]  Lena Costaridou,et al.  Modeling quantum noise of phosphors used in medical X-ray imaging detectors , 1999 .

[22]  E. Heijne,et al.  X-ray imaging using single photon processing with semiconductor pixel detectors , 2003 .

[23]  Ian A. Cunningham Linear-systems modeling of parallel cascaded stochastic processes: the NPS of radiographic screens with reabsorption of characteristic x-radiation , 1998, Medical Imaging.

[24]  Gyuseong Cho,et al.  The effect of direct x-ray on CMOS APS imager for industrial application , 2003, 2003 IEEE Nuclear Science Symposium. Conference Record (IEEE Cat. No.03CH37515).

[25]  Gyuseong Cho,et al.  Evaluation of CMOS APS imager for digital radiography and mammography , 2003, 2003 IEEE Nuclear Science Symposium. Conference Record (IEEE Cat. No.03CH37515).

[26]  N W Marshall,et al.  Early experience in the use of quantitative image quality measurements for the quality assurance of full field digital mammography x-ray systems , 2007, Physics in medicine and biology.

[27]  G. Cho,et al.  Development of a lens-coupled CMOS detector for an X-ray inspection system , 2005 .

[28]  I. Kandarakis,et al.  Experimental investigation of the optical signal, gain, signal-to-noise ratio and information content characteristics of X-ray phosphor screens , 2001 .

[29]  W. Rugh Linear System Theory , 1992 .

[30]  Sang Joon Kim,et al.  A Mathematical Theory of Communication , 2006 .

[31]  F. Verdun,et al.  A comparison of the performance of modern screen-film and digital mammography systems. , 2005 .

[32]  Francesco Fauci,et al.  X-ray spectroscopy and dosimetry with a portable CdTe device , 2007 .

[33]  J Yorkston,et al.  Empirical and theoretical investigation of the noise performance of indirect detection, active matrix flat-panel imagers (AMFPIs) for diagnostic radiology. , 1997, Medical physics.

[34]  Martin Spahn,et al.  Flat detectors and their clinical applications , 2005, European Radiology.

[35]  D W Rickey,et al.  Physical characterization and performance comparison of active- and passive-pixel CMOS detectors for mammography. , 2009, Physics in medicine and biology.

[36]  Min Kook Cho,et al.  Measurements of X-ray Imaging Performance of Granular Phosphors With Direct-Coupled CMOS Sensors , 2008, IEEE Transactions on Nuclear Science.

[37]  G. Entine,et al.  Structured CsI(Tl) scintillators for X-ray imaging applications , 1997, 1997 IEEE Nuclear Science Symposium Conference Record.

[38]  I. Kandarakis,et al.  The Monte Carlo evaluation of noise and resolution properties of granular phosphor screens , 2009, Physics in medicine and biology.

[39]  Christoph Hoeschen,et al.  Measurement of the detective quantum efficiency (DQE) of digital X-ray detectors according to the novel standard IEC 62220-1. , 2005, Radiation protection dosimetry.

[40]  Andrew D. A. Maidment,et al.  Digital Mammography , 2019, Advances in Clinical Radiology.

[41]  M J Yaffe,et al.  Model of the spatial-frequency-dependent detective quantum efficiency of phosphor screens. , 1990, Medical physics.

[42]  M. Nikl Scintillation detectors for x-rays , 2006 .

[43]  John A. Rowlands,et al.  Flat Panel Detectors for Digital Radiography , 2000 .

[44]  Ehsan Samei,et al.  Determination of the detective quantum efficiency of a digital x-ray detector: comparison of three evaluations using a common image data set. , 2004, Medical physics.

[45]  L Ramello,et al.  Contrast cancellation technique applied to digital x-ray imaging using silicon strip detectors. , 2005, Medical physics.

[46]  V. Kogan,et al.  Medipix 2 in X-ray diffraction , 2007 .

[47]  C E Ravin,et al.  Principles of digital radiography with large-area, electronically readable detectors: a review of the basics. , 1999, Radiology.

[48]  I. Cunningham Applied Linear-Systems Theory , 2000 .